Validation of the A&D UA-651 Plus/UA-651SL Plus automated sphygmomanometer according to the ISO 81060-2, 2018 and ISO 81060-2 Amendment 1, 2020, which resulted in the currently pending Amendment to Amendment 1

被引:5
|
作者
Alpert, Bruce S. [1 ]
机构
[1] Univ Tennessee, Dept Pediat, Hlth Sci Ctr, Memphis, TN USA
来源
JOURNAL OF CLINICAL HYPERTENSION | 2022年 / 24卷 / 04期
关键词
2020 ISO amendment; ISO validation; overlapping cuffs;
D O I
10.1111/jch.14464
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
In 2020 the ANSI/AAMI/ISO published Amendment 1 to the 81060-2 Standard. The purpose was to try to address issues relating to the distribution of subjects with respect to limb size ranges. It also required a new plot showing the limb circumferences on the x-axis and the device minus reference errors on the y-axis (Alpert plot). The study reported here showed inadequacies of the Amendment if significantly overlapping cuffs were to be tested. The innovative approach led to the testing of 135 subjects. Requirements for blood pressure (BP), arm circumference, and gender were fulfilled. The standard same arm sequential protocol was used. Criterion 1 calculations gave a mean +/- standard deviation (SD) device minus manual BP values of 0.22 +/- 7.90 mmHg for systolic BP and -0.68 +/- 7.36 mmHg for diastolic BP. The maximum allowed mean value for either measurement is 5.0 mmHg. The SD values for Criterion 2 were 6.03 mmHg for systolic BP (maximum allowed 6.95) and 6.47 mmHg for diastolic BP (maximum allowed 6.90). All results passed the Standard requirements. This study demonstrated inadequacies of Amendment 1 and led to the development of an Amendment 2, still in the process of finalization. The new Amendment corrects the issues with significantly overlapping cuffs and "wide-range" cuffs. The A&D UA-651 Plus/UA-651SL Plus BP monitors and the five cuffs not only passed the requirements of 81060-2:2018 and Amendment 1 but also a revised and more rigorous protocol with more subjects involved. These devices and cuffs can be used with confidence.
引用
收藏
页码:513 / 518
页数:6
相关论文
共 22 条
  • [1] Validation of the A&D BP UA-1200BLE device for home blood pressure measurement according to the ISO 81060-2:2013 standard
    Kario, Kazuomi
    Saito, Kimiyo
    Sato, Keiko
    Hamasaki, Haruna
    Suwa, Hiromi
    Okura, Ayako
    Hoshide, Satoshi
    BLOOD PRESSURE MONITORING, 2018, 23 (06) : 312 - 314
  • [2] Validation of the A&D UM-212BLE monitor according to ISO 81060-2, 2018: a device with clinically important programmability
    Alpert, Bruce S.
    BLOOD PRESSURE MONITORING, 2023, 28 (02) : 113 - 115
  • [3] Validation testing of five home blood pressure monitoring devices for the upper arm according to the ISO 81060-2:2018/AMD 1:2020 protocol
    Yaw Amofa Peprah
    Ji Young Lee
    Stephen D. Persell
    Journal of Human Hypertension, 2023, 37 : 134 - 140
  • [4] Validation of the Uscom BP plus automated oscillometric blood pressure monitor for professional office use in children and adolescents according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018)
    Glenning, Jonathan P.
    Sheeran, Freya
    Cuthbert, Jodie
    Harris, William
    Quinlan, Catherine
    Mynard, Jonathan P.
    HYPERTENSION RESEARCH, 2025, 48 (03) : 1125 - 1134
  • [5] Validation testing of five home blood pressure monitoring devices for the upper arm according to the ISO 81060-2:2018/AMD 1:2020 protocol
    Peprah, Yaw Amofa
    Lee, Ji Young
    Persell, Stephen D. D.
    JOURNAL OF HUMAN HYPERTENSION, 2023, 37 (02) : 134 - 140
  • [6] Validation of the ANDON KD-7920 wrist blood pressure monitor in general population according to the ISO 81060-2:2018+AMD1:2020 protocol
    Li, Duan
    Li, Yiming
    Li, Jiaheng
    Gu, Yuan
    Shang, Fujun
    BLOOD PRESSURE MONITORING, 2025, 30 (02)
  • [7] Validation of TMB-2285-BT wrist blood pressure monitor in general population according to the ISO 81060-2:2018+Amd.1:2020 Protocol
    Xie, Zijian
    Peng, Bin
    Hu, Jia
    Wang, Xinda
    Du, Xiaoqin
    Li, Chaoya
    Liang, Jiahui
    BLOOD PRESSURE MONITORING, 2025, 30 (01) : 44 - 47
  • [8] Validation of the JAMR upper-arm pressure monitor B23 according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018/AMD 1:2020)
    Li, Jing
    Han, Li-Hong
    Feng, Ru
    Wu, Li-Hao
    Cai, Jie
    Zeng, Tian-Ying
    Liu, Hui
    Wang, Lei
    Li, Ju-Yun
    Zhan, Zhong-Qun
    BLOOD PRESSURE MONITORING, 2025, 30 (01) : 40 - 43
  • [9] Validation of the DBP-1333b upper-arm blood pressure monitor according to the AAMI/ESH/ISO universal standard (ISO 81060-2:2018+Amd.1:2020)
    Hou, Guocun
    Wu, Yulu
    Wang, Jiaying
    Zhi, Jia
    BLOOD PRESSURE MONITORING, 2024, 29 (03) : 149 - 155
  • [10] Validation of the U60EH Wrist Electronic Blood Pressure Monitor in general population according to the ISO 81060-2:2018/AMD 1:2020 Protocol
    Mu, Sicong
    Zhu, Min
    Ma, Liping
    Fang, Lixiu
    BLOOD PRESSURE MONITORING, 2023, 28 (05) : 280 - 283