A new high-performance interface for capillary electrophoresis electrospray ionization mass spectrometry

被引:0
|
作者
Severs, JC [1 ]
Harms, AC [1 ]
Smith, RD [1 ]
机构
[1] PACIFIC NW NATL LAB,ENVIRONM MOL SCI LAB,RICHLAND,WA 99352
关键词
D O I
暂无
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel capillary electrophoresis electrospray ionization mass spectrometry (CE/ESI-MS) interface has been developed based upon an alternative means of establishing electrical contact at the CE capillary terminus. In this design, the terminus of the CE separation capillary and a short ESI emitter capillary are connected and sheathed by polysulphone micro-dialysis tubing. The electrical connection, to close the CE circuit and simultaneously establish the electrospray voltage, is provided outside the membrane via an electrode in a small liquid reservoir. In contrast to the sheath flow and liquid junction interfaces, the flow rate of liquid to the ESI source consists of the CE eluent and is not increased. The use of a 'micro-spray' type emitter produces a stable electrospray at the very low flow rates typical of CE, providing a highly sensitive, efficient interface. Post-separation acidification of analytes via the dialysis junction is also demonstrated.
引用
收藏
页码:1175 / 1178
页数:4
相关论文
共 50 条
  • [41] Application of capillary electrophoresis electrospray ionization mass spectrometry in the determination of molecular diversity
    Dunayevskiy, YM
    Vouros, P
    Wintner, EA
    Shipps, GW
    Carell, T
    Rebek, J
    PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1996, 93 (12) : 6152 - 6157
  • [42] A new sheathless electrospray interface for coupling of capillary electrophoresis to ion-trap mass spectrometry
    Bendahl, L
    Hansen, SH
    Olsen, J
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2002, 16 (24) : 2333 - 2340
  • [43] Study of the electrical connection mechanism of sheathless interface for capillary electrophoresis-electrospray ionization-mass spectrometry
    Tie, Cai
    Zhang, De-Wen
    Chen, Hong-Xu
    Song, Sen-Lin
    Zhang, Xin-Xiang
    JOURNAL OF MASS SPECTROMETRY, 2012, 47 (11): : 1429 - 1434
  • [44] CAPILLARY ZONE ELECTROPHORESIS AND ISOTACHOPHORESIS MASS-SPECTROMETRY OF POLYPEPTIDES AND PROTEINS BASED UPON AN ELECTROSPRAY IONIZATION INTERFACE
    SMITH, RD
    LOO, JA
    BARINAGA, CJ
    EDMONDS, CG
    UDSETH, HR
    JOURNAL OF CHROMATOGRAPHY, 1989, 480 : 211 - 232
  • [45] CAPILLARY ELECTROPHORESIS ELECTROSPRAY-IONIZATION ION-TRAP MASS-SPECTROMETRY USING A SHEATHLESS INTERFACE
    RAMSEY, RS
    MCLUCKEY, SA
    JOURNAL OF MICROCOLUMN SEPARATIONS, 1995, 7 (05) : 461 - 469
  • [46] Capillary monoliths for the analysis of nucleic acids by high-performance liquid chromatography-electrospray ionization mass spectrometry
    Oberacher, H
    Huber, CG
    TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2002, 21 (03) : 166 - 174
  • [47] NEW DEVELOPMENTS IN MICROSCALE SEPARATIONS AND MASS-SPECTROMETRY FOR BIOMONITORING - CAPILLARY ELECTROPHORESIS AND ELECTROSPRAY-IONIZATION MASS-SPECTROMETRY
    SMITH, RD
    WAHL, JH
    LIGHTWAHL, KJ
    WINGER, BE
    JOURNAL OF TOXICOLOGY AND ENVIRONMENTAL HEALTH, 1993, 40 (2-3): : 147 - 158
  • [48] Capillary electrophoresis mass spectrometry with sheathless electrospray ionization for high sensitivity analysis of underivatized amino acids
    Jeong, Ji-Seon
    Kim, Sook-Kyung
    Park, Sang-Ryoul
    ELECTROPHORESIS, 2012, 33 (14) : 2112 - 2121
  • [49] New approach for the detection of BSH and its metabolites using capillary electrophoresis and electrospray ionization mass spectrometry
    Mauri, PL
    Basilico, F
    Pietta, PG
    Pasini, E
    Monti, D
    Sauerwein, W
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 788 (01): : 9 - 16
  • [50] A sheathless nanoflow electrospray interface for on-line capillary electrophoresis mass spectrometry
    Janini, GM
    Conrads, TP
    Wilkens, KL
    Issaq, HJ
    Veenstra, TD
    ANALYTICAL CHEMISTRY, 2003, 75 (07) : 1615 - 1619