Online measurement of urea concentration in spent dialysate during hemodialysis

被引:4
|
作者
Olesberg, JT [1 ]
Armitage, B [1 ]
Arnold, MA [1 ]
Flanigan, MJ [1 ]
机构
[1] Univ Iowa, Dept Chem & Opt Sci, Iowa City, IA 52242 USA
关键词
urea; hemodialysis; noninvasive; spectroscopy; online monitoring; creatinine;
D O I
10.1117/12.468312
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We describe on-line optical measurements of urea concentration during the regular hemodialysis treatment of several patients. The spectral measurements were performed in the effluent dialysate stream after the dialysis membrane using an FTIR spectrometer equipped with a flow-through cell. Spectra were recorded across the 5000-4000 cm(-1) (2.0-2.5 mum) wavelength range at 1-minute intervals. Optically determined concentrations matched concentrations obtained from standard chemical assays with a root-mean-square error of 0.29 mM for urea (0.8 mg/dl urea nitrogen), 0.03 MM for creatinine, 0.11 mM for lactate, and 0.22 mM for glucose. The observed concentration ranges were 0-11 mm for urea, 00.35 mM for creatinine, 0-0.75 mM for lactate, and 9-12.5 mM for glucose.
引用
收藏
页码:95 / 105
页数:11
相关论文
共 50 条
  • [1] Online measurement of urea concentration in spent dialysate during hemodialysis
    Olesberg, JT
    Arnold, MA
    Flanigan, MJ
    [J]. CLINICAL CHEMISTRY, 2004, 50 (01) : 175 - 181
  • [2] Online Urea Concentration Estimation from Spent Dialysate Using Optical Sensor
    Pilt, Kristjan
    Arund, Jurgen
    Adoberg, Annika
    Leis, Liisi
    Luman, Merike
    Fridolin, Ivo
    [J]. XV MEDITERRANEAN CONFERENCE ON MEDICAL AND BIOLOGICAL ENGINEERING AND COMPUTING - MEDICON 2019, 2020, 76 : 1459 - 1464
  • [3] Nanoengineered optical urea biosensor for estimating hemodialysis parameters in spent dialysate
    Swati, M.
    Hase, N. K.
    Srivastava, Rohit
    [J]. ANALYTICA CHIMICA ACTA, 2010, 676 (1-2) : 68 - 74
  • [4] A new biosensor for continuous monitoring of the spent dialysate urea level in standard hemodialysis
    Calzavara, P
    Calconi, G
    Da Rin, G
    Canini, E
    Paolini, F
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1998, 21 (03): : 147 - 150
  • [5] Comparison of hemodialysis urea clearance using spent dialysate and Kt/Vurea equations
    Khatri, Priyanka
    Davenport, Andrew
    [J]. ARTIFICIAL ORGANS, 2024,
  • [6] Removal of Urea, β2-Microglobulin, and Indoxyl Sulfate Assessed by Absorbance and Fluorescence in the Spent Dialysate During Hemodialysis
    Lauri, Kai
    Arund, Jurgen
    Holmar, Jana
    Tanner, Risto
    Kalle, Sigrid
    Luman, Merike
    Fridolin, Ivo
    [J]. ASAIO JOURNAL, 2020, 66 (06) : 698 - 705
  • [7] Near-infrared spectroscopic measurement of urea in dialysate samples collected during hemodialysis treatments
    Eddy, CV
    Flanigan, M
    Arnold, MA
    [J]. APPLIED SPECTROSCOPY, 2003, 57 (10) : 1230 - 1235
  • [8] Dose of dialysis predicted by continuous measurement of dialysate urea concentration
    Cappello, A
    Tartarini, R
    Paolini, F
    Calzavara, P
    [J]. PROCEEDINGS OF THE 18TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 18, PTS 1-5, 1997, 18 : 1813 - 1815
  • [9] DIALYSATE ALUMINUM CONCENTRATION AND ALUMINUM TRANSFER DURING HEMODIALYSIS
    GRAF, H
    STUMMVOLL, HK
    MEISINGER, V
    [J]. LANCET, 1982, 1 (8262): : 46 - 47
  • [10] OPTIMUM DIALYSATE CALCIUM CONCENTRATION DURING MAINTENANCE HEMODIALYSIS
    JOHNSON, WJ
    [J]. NEPHRON, 1976, 17 (04): : 241 - 258