Near-infrared spectroscopic measurement of urea in dialysate samples collected during hemodialysis treatments

被引:18
|
作者
Eddy, CV
Flanigan, M
Arnold, MA [1 ]
机构
[1] Univ Iowa, Opt Sci & Technol, Dept Chem, Iowa City, IA 52242 USA
[2] Univ Iowa, Dept Internal Med, Iowa City, IA 52242 USA
关键词
near-infrared spectroscopy; hemodialysis; urea; partial least-squares regression; PLS; noninvasive chemical sensing;
D O I
10.1366/000370203769699081
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Single-beam spectra were collected over the combination region of the near-infrared spectrum for 80 samples collected from 15 people over a two-week period. Partial least-squares (PLS) regression was used to generate an optimized calibration model for urea. PLS calibration models accurately measure urea in the spent dialysate matrix. Prediction errors are on the order of 0.15 mM, which is sufficient for the clinical assessment of the dialysis process. In addition, the feasibility of a global calibration model is demonstrated by generating a calibration model from samples and spectra obtained from 12 people to predict the level of urea in samples collected from 3 different people. In this case, the standard error of prediction is 0.09 mM. Spectra were modified in order to systematically examine the impact of resolution and noise. Little impact is observed by altering the spectral resolution from 4 to 32 cm(-1). Spectral noise, however, plays an important role in the accuracy of these calibration models. Increasing the magnitude of the spectral noise increases the prediction errors and increases the width of the spectral range necessary for extracting the analytical information. The utility of the method is demonstrated by analyzing dialysate samples collected during actual dialysis treatments. In addition, the necessary resolution and spectral quality necessary for reliable on-line urea monitoring is identified. These findings indicate that a dedicated, on-line urea spectrometer must posses a resolution of 16 cm(-1) coupled with a sample thickness of 1.5 mm and spectral noise levels on the order of 25 micro-absorbance units when measured as the root-mean-square (RMS) noise of 100% lines.
引用
收藏
页码:1230 / 1235
页数:6
相关论文
共 50 条
  • [1] Near-infrared spectroscopy for measuring urea in hemodialysis fluids
    Eddy, CV
    Arnold, MA
    [J]. CLINICAL CHEMISTRY, 2001, 47 (07) : 1279 - 1286
  • [2] Online measurement of urea concentration in spent dialysate during hemodialysis
    Olesberg, JT
    Arnold, MA
    Flanigan, MJ
    [J]. CLINICAL CHEMISTRY, 2004, 50 (01) : 175 - 181
  • [3] Online measurement of urea concentration in spent dialysate during hemodialysis
    Olesberg, JT
    Armitage, B
    Arnold, MA
    Flanigan, MJ
    [J]. OPTICAL DIAGNOSTICS AND SENSING OF BIOLOGICAL FLUIDS AND GLUCOSE AND CHOLESTEROL MONITORING II, 2002, 4624 : 95 - 105
  • [4] Noninvasive urea monitoring during hemodialysis by transcutaneous near-infrared spectroscopy.
    Rohrscheib, M
    Zager, P
    Robinson, R
    Rowe, R
    [J]. JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2002, 13 : 414A - 414A
  • [5] Use of near-infrared spectroscopic measurement at the buttocks during abdominal aortic surgery
    Fukui, D
    Urayama, H
    Tanaka, K
    Kawasaki, S
    [J]. CIRCULATION JOURNAL, 2002, 66 (12) : 1128 - 1131
  • [6] On-line near-infrared spectrometer to monitor urea removal in real time during hemodialysis
    Cho, David S.
    Olesberg, Jonathon T.
    Flanigan, Michael J.
    Arnold, Mark A.
    [J]. APPLIED SPECTROSCOPY, 2008, 62 (08) : 866 - 872
  • [7] NEAR-INFRARED SPECTROSCOPIC MEASUREMENT OF GLUCOSE IN A PROTEIN MATRIX
    MARQUARDT, LA
    ARNOLD, MA
    SMALL, GW
    [J]. ANALYTICAL CHEMISTRY, 1993, 65 (22) : 3271 - 3278
  • [8] Effects of temperature on near-infrared spectroscopic measurement of glucose
    Jung, BJ
    McShane, M
    Rastegar, S
    Cote, GL
    [J]. BIOMEDICAL SENSING AND IMAGING TECHNOLOGIES, PROCEEDINGS OF, 1998, 3253 : 41 - 46
  • [9] Effects of temperature on near-infrared spectroscopic measurement of water
    Ding, D
    Luo, HY
    Zhang, HY
    Tang, YG
    Shen, XG
    [J]. INFRARED COMPONENTS AND THEIR APPLICATIONS, 2005, 5640 : 376 - 380
  • [10] Near-infrared spectroscopic measurement of lactate in human plasma
    Lafrance, D
    Lands, LC
    Hornby, L
    Burns, DH
    [J]. APPLIED SPECTROSCOPY, 2000, 54 (02) : 300 - 304