μ-dDIHEN: a new micro-flow liquid sample introduction system for direct injection nebulization in ICP-MS

被引:12
|
作者
Louvat, Pascale [1 ]
Tharaud, Mickael [1 ]
Buisson, Matthieu [1 ]
Rollion-Bard, Claire [1 ]
Benedetti, Marc F. [1 ]
机构
[1] Univ Paris, CNRS, Inst Phys Globe Paris, F-75005 Paris, France
关键词
HIGH-EFFICIENCY NEBULIZER; PLASMA-MASS SPECTROMETRY; ORGANIC/HYDRO-ORGANIC MATRICES; OPTICAL-EMISSION SPECTROMETRY; BIOLOGICAL SAMPLES; CHROMATOGRAPHY; ELEMENTS; BORON; PRODUCTS; SEAWATER;
D O I
10.1039/c9ja00146h
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The mu-dDIHEN allows direct nebulization of small samples into the plasma at a flow rate down to 5 mu L min(-1). It is composed of a demountable direct injection high efficiency nebulizer (d-DIHEN), a flow injection analysis (FIA) valve and a gas displacement pump (GDP) connected to a mu-mass-flow meter delivering very stable liquid flow. The system was successfully tested with sample loops of 10 and 50 mu L and for liquid flow rates between 5 and 50 mu L min(-1) on two different inductively coupled plasma mass spectrometry (ICP-MS) instruments, a sector-field-ICP-MS (SF-ICP-MS) and a multi-collector-ICP-MS (MC-ICP-MS), and for three different applications: boron isotopic ratio measurement of geological samples, multi-element analyses of natural water samples and gold nanoparticle characterization by single particle ICPMS (spICP-MS). Signal sensitivity was increasing with the liquid uptake rate of the mu-dDIHEN (and the GDP gas flow rate), but stopped increasing above 30 mu L min(-1), due to bad spray quality and poor ionization under extra-wet conditions. We successfully measured boron isotopic ratios (delta B-11) for reference material solutions with B concentrations between 10 and 200 mu g L-1; the reproducibility of the measurements depending on both the B concentration of the sample and the uptake rate: B-10 signals below 20-30 mV (with 10(11) ohms amplifiers) gave 2SD reproducibility >= 0.5 parts per thousand (n = 5). The best measurement conditions were not necessarily with the lowest uptake rate but rather at 25 mu L min(-1), allowing for higher signals. We also tested transient versus continuous measurement modes for a series of natural samples (with 10 and 50 mu L sample loops). The two modes gave fully comparable results but the transient mode allows triplicate delta B-11 measurements and is recommended for very small samples (<= 100 mu L and B content of a few ng). For multi-element analyses by SF-ICP-MS, the limits of quantification (LOQs) of the majority of the trace and ultra-trace elements were lower (up to 15 times) with the mu-dDIHEN than with a concentric nebulizer and spray chamber, due to the much lower sample volume needed. Lastly, the mu-dDIHEN is highly promising for nanoparticle (NP) characterization by spICPMS. The first measurements for gold NPs of 40 nm certified size led to a transport efficiency of 85%.
引用
收藏
页码:1553 / 1563
页数:11
相关论文
共 50 条
  • [31] Determination of Fe, Ti in Seawater Using a Flow-injection Pretreatment System with ICP-MS
    Ge, Huanying
    Chen, Ding
    Li, Shu
    Zheng, Airong
    Zhu, Zuhao
    Zheng, Xuehong
    JOURNAL OF COASTAL RESEARCH, 2019, : 77 - 82
  • [32] Improving sensitivity and detection limits in ICP-MS with a novel high-efficiency sample introduction system
    Debrah, E
    Légère, G
    ATOMIC SPECTROSCOPY, 1999, 20 (02) : 73 - 77
  • [33] Effect of the sample introduction system in ICP-MS on the formation of both oxygenated polyatomic and doubly charged ions
    Tubaro, F
    Barbangelo, F
    Toniolo, R
    Di Narda, F
    Bontempelli, G
    ANNALI DI CHIMICA, 1999, 89 (11-12) : 863 - 872
  • [34] Development and characterization of a Flow Focusing multi nebulization system for sample introduction in ICP-based spectrometric techniques
    Kovachev, Nikolay
    Almagro, Beatriz
    Angel Aguirre, Miguel
    Hidalgo, Montserrat
    Ganan-Calvo, Alfonso M.
    Canals, Antonio
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2009, 24 (09) : 1213 - 1221
  • [35] Development and characterization of a Flow Focusing multi nebulization system for sample introduction in ICP-based spectrometric techniques
    Kovachev, Nikolay
    Almagro, Beatriz
    Aguirre, Miguel Ángel
    Hidalgo, Montserrat
    Gañán-Calvo, Alfonso M.
    Canals, Antonio
    Journal of Analytical Atomic Spectrometry, 2009, 24 (09): : 1213 - 1221
  • [36] ICP-MS MULTIELEMENT ANALYSES OF SOLID AND LIQUID SPECIMENS BY LASER VAPORIZATION AND FLOW-INJECTION ANALYSIS
    VOLKMER, MK
    STAHL UND EISEN, 1993, 113 (07): : 107 - 108
  • [37] Spectroscopy tutorial - A beginner's guide to ICP-MS - Part II: The sample-introduction system
    Thomas, R
    SPECTROSCOPY, 2001, 16 (05) : 56 - +
  • [38] Comparative study of three sample preparation approaches for the fast determination of americium in urine by flow injection ICP-MS
    Epov, V. N.
    Benkhedda, K.
    Brownell, D.
    Cornett, R. J.
    Evans, R. D.
    CANADIAN JOURNAL OF ANALYTICAL SCIENCES AND SPECTROSCOPY, 2005, 50 (01): : 14 - 22
  • [39] Direct multielemental analysis of human serum by ICP-MS with on-line standard addition using flow injection
    Huang, CY
    Beauchemin, D
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2003, 18 (08) : 951 - 952
  • [40] Micro-flow injection analysis system: on-chip sample preconcentration, injection and delivery using coupled monolithic electroosmotic pumps
    Nie, Fu-Qiang
    Macka, Mirek
    Paull, Brett
    LAB ON A CHIP, 2007, 7 (11) : 1597 - 1599