Biomechanical analysis of plate systems for proximal humerus fractures: a systematic literature review

被引:41
|
作者
Jabran, Ali [1 ]
Peach, Chris [1 ,2 ]
Ren, Lei [1 ]
机构
[1] Univ Manchester, Sch Mech Aerosp & Civil Engn, Manchester, Lancs, England
[2] Univ Hosp South Manchester, Dept Shoulder & Elbow Surg, Manchester, Lancs, England
来源
基金
英国工程与自然科学研究理事会;
关键词
Proximal humerus fractures; Locking plates; Open reduction internal fixation; Biomechanical testing; SURGICAL NECK FRACTURES; INTRAMEDULLARY FIBULAR ALLOGRAFT; CALCIUM-PHOSPHATE CEMENT; HUMAN SHOULDER SPECIMENS; MEDIAL CORTICAL SUPPORT; FINITE-ELEMENT-ANALYSIS; ANGULAR STABLE PLATE; ROTATOR CUFF SUTURES; IN-VITRO ASSESSMENT; LOCKING PLATE;
D O I
10.1186/s12938-018-0479-3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Proximal humerus fractures are the third most common in the human body but their management remains controversial. Open reduction and internal fixation with plates is one of the leading modes of operative treatment for these fractures. The development of technologies and techniques for these plates, during the recent decades, promise a bright future for their clinical use. A comprehensive review of in vitro biomechanical studies is needed for the comparison of plates' mechanical performance and the testing methodologies. This will not only guide clinicians with plate selection but also with the design of future in vitro biomechanical studies. This review was aimed to systematically categorise and review the in vitro biomechanical studies of these plates based on their protocols and discuss their results. The technologies and techniques investigated in these studies were categorised and compared to reach a census where possible. Methods and results: Web of Science and Scopus database search yielded 62 studies. Out of these, 51 performed axial loading, torsion, bending and/or combined bending and axial loading while 11 simulated complex glenohumeral movements by using tendons. Loading conditions and set-up, failure criteria and performance parameters, as well as results for each study, were reviewed. Only two studies tested four-part fracture model while the rest investigated two-and three-part fractures. In ten studies, synthetic humeri were tested instead of cadaveric ones. In addition to load-displacement data, three-dimensional motion analysis systems, digital image correlation and acoustic emission testing have been used for measurement. Conclusions: Overall, PHILOS was the most tested plate and locking plates demonstrated better mechanical performance than non-locking ones. Conflicting results have been published for their comparison with non-locking blade plates and polyaxial locking screws. Augmentation with cement [calcium phosphate or poly(methyl methacrylate)] or allografts (fibular and femoral head) was found to improve bone-plate constructs' mechanical performance. Controversy still lies over the use of rigid and semi-rigid implants and the insertion of inferomedial screws for calcar region support. This review will guide the design of in vitro and in silico biomechanical tests and also supplement the study of clinical literature.
引用
收藏
页数:30
相关论文
共 50 条
  • [1] Biomechanical analysis of plate systems for proximal humerus fractures: a systematic literature review
    Ali Jabran
    Chris Peach
    Lei Ren
    BioMedical Engineering OnLine, 17
  • [2] Biomechanical analysis of plate osteosynthesis systems for proximal humerus fractures
    Lever, James P.
    Aksenov, Sergei A.
    Zdero, Rad
    Ahn, Henry
    Mckee, Michael D.
    Schemitsch, Emil H.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2008, 22 (01) : 23 - 29
  • [3] A systematic review of locking plate fixation of proximal humerus fractures
    Sproul, Robert C.
    Iyengar, Jaicharan J.
    Devcic, Zlatko
    Feeley, Brian T.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2011, 42 (04): : 408 - 413
  • [4] Hybrid blade and locking plate fixation for proximal humerus fractures: a comparative biomechanical analysis
    Jabran, Ali
    Peach, Chris
    Zou, Zhenmin
    Ren, Lei
    BIOMEDICAL ENGINEERING ONLINE, 2018, 17
  • [5] Hybrid blade and locking plate fixation for proximal humerus fractures: a comparative biomechanical analysis
    Ali Jabran
    Chris Peach
    Zhenmin Zou
    Lei Ren
    BioMedical Engineering OnLine, 17
  • [6] Locked plate fixation and intramedullary nailing for proximal humerus fractures: A biomechanical evaluation
    Hessmann, MH
    Hansen, WSM
    Krummenauer, F
    Pol, TF
    Rommens, M
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2005, 58 (06): : 1194 - 1201
  • [7] Proximal humerus fractures in the pediatric population: a systematic review
    Pahlavan, Sohrab
    Baldwin, Keith D.
    Pandya, Nirav K.
    Namdari, Surena
    Hosalkar, Harish
    JOURNAL OF CHILDRENS ORTHOPAEDICS, 2011, 5 (03) : 187 - 194
  • [8] Nonoperative Treatment of Proximal Humerus Fractures: A Systematic Review
    Iyengar, Jaicharan J.
    Devcic, Zlatko
    Sproul, Robert C.
    Feeley, Brian T.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2011, 25 (10) : 612 - 617
  • [9] Surgical treatment of proximal humerus fractures: a systematic review and meta-analysis
    Erik Hohmann
    Natalie Keough
    Vaida Glatt
    Kevin Tetsworth
    European Journal of Orthopaedic Surgery & Traumatology, 2023, 33 : 2215 - 2242
  • [10] Surgical treatment of proximal humerus fractures: a systematic review and meta-analysis
    Hohmann, Erik
    Keough, Natalie
    Glatt, Vaida
    Tetsworth, Kevin
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2023, 33 (06): : 2215 - 2242