An 11-year retrospective study of totally implanted central venous access ports: Complications and patient satisfaction

被引:94
|
作者
Ignatov, A.
Hoffman, O.
Smith, B.
Fahlke, J. [2 ]
Peters, B. [3 ]
Bischoff, J.
Costa, S. -D. [1 ]
机构
[1] Otto Von Guericke Univ, Womens Clin, Dept Obstet & Gynecol, D-39108 Magdeburg, Germany
[2] Otto VonGuericke Univ Magdegurg, Dept Gen Visceral & Vasc Surg, D-39016 Magdeburg, Germany
[3] Otto VonGuericke Univ Magdegurg, Dept Biometr, D-39016 Magdeburg, Germany
来源
EJSO | 2009年 / 35卷 / 03期
关键词
Totally implanted access ports; TIAP; Central venous access ports; Catheters; Venous access devices; ATRIAL CATHETER; CHEMOTHERAPY; EXPERIENCE; DEVICES; SYSTEM; COSTS;
D O I
10.1016/j.ejso.2008.01.020
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aims: We wanted to assess the factors that predict complications and patient satisfaction of totally implanted central venous access ports (TIAP). Methods: We reviewed 550 patients with breast or gynaecological malignancies who had initial port placement for chemotherapy between 1995 and 2006. We retrospectively assessed all TIAP complications, port duration and follow-up care until the TIAPs were removed (or the last known recorded documentation) or until the death of the patient. TIAP-related patient satisfaction was also assessed via a questionnaire-based survey of 356 patients. Results: 561 TIAPs were placed in 550 cancer patients (11 patients received 2 TIAPs during the study period); the median time of port duration was 22.5 months. There were 104 complications in this group. Of these, 81 occurred during chemotherapy treatment that lasted a median time of 182 days. Removal secondary to complication was observed in 48 cases. TIAPs placed on the left chest side, through the subclavian vein or with the catheter tip localized in the peripheral part of superior vena cava demonstrated the highest incidence of complications. Patients with a BMI >28.75 had an increased risk for developing complications. Our follow-up questionnaire revealed a 93% patient satisfaction rate with the TIAP. Conclusions: Patients with left-sided ports, catheter tips lying in the upper part of the superior vena Cava and implantation via the subclavian vein are at a higher risk for TIAP-associated complications. Being excessively overweight was assessed as another risk factor for developing complications. TIAPs are highly accepted and further recommended by patients. (C) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:241 / 246
页数:6
相关论文
共 50 条
  • [31] Removal of totally implanted venous access ports for suspected infection in the intensive care unit: a multicenter observational study
    Lecronier, Marie
    Valade, Sandrine
    Bige, Naike
    de Prost, Nicolas
    Roux, Damien
    Lebeaux, David
    Maury, Eric
    Azoulay, Elie
    Demoule, Alexandre
    Dres, Martin
    ANNALS OF INTENSIVE CARE, 2018, 8
  • [32] Explantation of totally implantable venous access ports of the forearm: reasons for removal and observed complications
    Goltz, Jan P.
    Kickuth, Ralph
    Scholl, Anne
    Machann, Wolfram
    Ritter, Christian O.
    Hahn, Dietbert
    Wittenberg, Guenther
    JOURNAL OF VASCULAR ACCESS, 2011, 12 (01): : 45 - 51
  • [33] Risk factors for complications in cancer patients with totally implantable access ports: A retrospective study and review of the literature
    Bademler, Suleyman
    Ucuncu, Muhammed
    Yildirim, Ilknur
    Karanlik, Hasan
    JOURNAL OF INTERNATIONAL MEDICAL RESEARCH, 2019, 47 (02) : 702 - 709
  • [34] Totally implantable central venous access ports in patients with cystic fibrosis: a multicenter prospective cohort study
    Dal Molin, Alberto
    Di Massimo, Dino Stefano
    Braggion, Cesare
    Bisogni, Sofia
    Rizzi, Elena
    D'Orazio, Ciro
    Di Toppa, Maria Vittoria
    Alghisi, Federico
    Cristadoro, Simona
    Carnovale, Vincenzo
    Festa, Gabriella
    Rampini, Serena
    Colombo, Carla
    Oneta, Anna
    Furnari, Maria Lucia
    Calamia, Maria Antonietta
    Zunino, Maria Luisa
    Tuccio, Giuseppe
    Spadea, Vincenzo
    Messore, Barbara
    Grosso, Bianca
    Festini, Filippo
    JOURNAL OF VASCULAR ACCESS, 2012, 13 (03): : 290 - 295
  • [35] Usefulness of Totally Implantable Central Venous Access Devices in Elderly Patients: A Retrospective Study
    Imaoka, Yuki
    Kuranishi, Fumito
    Ogawa, Yoshiteru
    ANNALS OF NUTRITION AND METABOLISM, 2018, 72 (02) : 112 - 116
  • [36] Totally implantable venous access ports and associated complications in sub-Saharan Africa: a single-centre retrospective analysis
    Salawu, Kahmil
    Arowojolu, Oreoluwa
    Afolaranmi, Oluwasegun
    Jimoh, Mutiu
    Nworgu, Chukwumere
    Falase, Bode
    ECANCERMEDICALSCIENCE, 2022, 16
  • [37] Pyoderma gangrenosum after totally implanted central venous access device insertion
    Inan, Ihsan
    Myers, Patrick O.
    Braun, Rolf
    Hagen, Monica E.
    Morel, Philippe
    WORLD JOURNAL OF SURGICAL ONCOLOGY, 2008, 6 (1)
  • [38] Pyoderma gangrenosum after totally implanted central venous access device insertion
    Ihsan Inan
    Patrick O Myers
    Rolf Braun
    Monica E Hagen
    Philippe Morel
    World Journal of Surgical Oncology, 6
  • [39] Totally implantable central venous access ports for M/XDR-TB treatment
    Skrahin, Aliaksandr
    Pechinski, Dmitri
    Makovski, Nikolai
    Dobysh, Kirill
    Sinha, Animesh
    Kmyz, Mikhail
    Hlushanina, Hanna
    Skrahina, Alena
    EUROPEAN RESPIRATORY JOURNAL, 2017, 50
  • [40] Morbidity after Insertion of Totally Implantable Venous Access Ports in Oncological Patients: Results of a Retrospective Clinical Study
    Aspalter, Manuela
    Lechner, Michael
    Linni, Klaus
    Hitzl, Wolfgang
    Hoelzenbein, Thomas
    Oefner, Dietmar
    Emmanuel, Klaus
    AMERICAN SURGEON, 2014, 80 (02) : 204 - 207