Quantitative Ultrasound Imaging for Monitoring In Situ High-Intensity Focused Ultrasound Exposure

被引:17
|
作者
Ghoshal, Goutam [1 ,2 ]
Kemmerer, Jeremy P. [2 ]
Karunakaran, Chandra [2 ]
Abuhabsah, Rami [2 ]
Miller, Rita J. [2 ]
Sarwate, Sandhya [2 ]
Oelze, Michael L. [2 ]
机构
[1] Acoust MedSyst Inc, Savoy, IL USA
[2] Univ Illinois, Dept Elect & Comp Engn, Urbana, IL 61801 USA
关键词
quantitative ultrasound imaging; HIFU; effective scatterer diameter; effective scatterer concentration; parametric imaging; NONINVASIVE TEMPERATURE ESTIMATION; TISSUE TEMPERATURE; PROPAGATION SPEED; VIVO; ATTENUATION; BACKSCATTER; DEPENDENCE; FEASIBILITY; THERAPY; PHANTOM;
D O I
10.1177/0161734614524179
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Quantitative ultrasound (QUS) imaging is hypothesized to map temperature elevations induced in tissue with high spatial and temporal resolution. To test this hypothesis, QUS techniques were examined to monitor high-intensity focused ultrasound (HIFU) exposure of tissue. In situ experiments were conducted on mammary adenocarcinoma tumors grown in rats and lesions were formed using a HIFU system. A thermocouple was inserted into the tumor to provide estimates of temperature at one location. Backscattered time-domain waveforms from the tissue during exposure were recorded using a clinical ultrasonic imaging system. Backscatter coefficients were estimated using a reference phantom technique. Two parameters were estimated from the backscatter coefficient (effective scatterer diameter (ESD) and effective acoustic concentration (EAC). The changes in the average parameters in the regions corresponding to the HIFU focus over time were correlated to the temperature readings from the thermocouple. The changes in the EAC parameter were consistently correlated to temperature during both heating and cooling of the tumors. The changes in the ESD did not have a consistent trend with temperature. The mean ESD and EAC before exposure were 120 +/- 16 mm and 32 +/- 3 db/cm(3), respectively, and changed to 144 +/- 9 mm and 51 +/- 7 db/cm(3), respectively, just before the last HIFU pulse was delivered to the tissue. After the tissue cooled down to 37 degrees C, the mean ESD and EAC were 126 +/- 8 mm and 35 +/- 4 db/cm(3), respectively. Peak temperature in the range of 50-60 degrees C was recorded by a thermocouple placed just behind the tumor. These results suggest that QUS techniques have the potential to be used for non-invasive monitoring of HIFU exposure.
引用
收藏
页码:239 / 255
页数:17
相关论文
共 50 条
  • [31] Enhancement of ultrasound contrast agent in High-Intensity focused ultrasound ablation
    Wen Luo
    Xiaodong Zhou
    Xue Tian
    Xialong Ren
    Minjuan Zheng
    Kejun Gu
    Guangbin He
    Advances in Therapy, 2006, 23 : 861 - 868
  • [32] Enhancement of ultrasound contrast agent in high-intensity focused ultrasound ablation
    Luo, Wen
    Zhou, Xiaodong
    Tian, Xue
    Ren, Xialong
    Zheng, Minjuan
    Gu, Kejun
    He, Guangbin
    ADVANCES IN THERAPY, 2006, 23 (06) : 861 - 868
  • [33] A method to synchronize high-intensity, focused ultrasound with an arbitrary ultrasound imager
    Owen, NR
    Bailey, MR
    Hossack, J
    Crum, LA
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2006, 53 (03) : 645 - 650
  • [34] Study on cavitation behavior during high-intensity focused ultrasound exposure by using optical and ultrasonic imaging
    Taguchi, Kei
    Takagi, Ryo
    Yasuda, Jun
    Yoshizawa, Shin
    Umemura, Shin-ichiro
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2016, 55 (07)
  • [35] A Laboratory Facility for the Controlled Exposure of Blood Vessels to High-Intensity Focused Ultrasound
    Berkovich A.E.
    Bursian A.A.
    Senchik K.U.
    Petrishchev N.N.
    Tsibin A.U.
    Yukina G.U.
    Biomedical Engineering, 2016, 50 (02) : 96 - 99
  • [36] Lysis of Chlamydomonas reinhardtii by high-intensity focused ultrasound as a function of exposure time
    Bigelow, Timothy A.
    Xu, Jin
    Stessman, Dan J.
    Yao, Linxing
    Spalding, Martin H.
    Wang, Tong
    ULTRASONICS SONOCHEMISTRY, 2014, 21 (03) : 1258 - 1264
  • [37] Golay-Encoded Ultrasound Monitoring of Simultaneous High-Intensity Focused Ultrasound Treatment: A Phantom Study
    Shen, Che-Chou
    Lin, Ri-Cheng
    Wu, Nien-Hung
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2022, 69 (04) : 1370 - 1381
  • [38] Enhanced Lesion-to-Bubble Ratio on Ultrasonic Nakagami Imaging for Monitoring of High-Intensity Focused Ultrasound
    Zhang, Siyuan
    Li, Chong
    Zhou, Fanyu
    Wan, Mingxi
    Wang, Supin
    JOURNAL OF ULTRASOUND IN MEDICINE, 2014, 33 (06) : 959 - 970
  • [39] Spatiotemporal Monitoring of High-Intensity Focused Ultrasound Therapy with Passive Acoustic Mapping
    Jensen, Carl R.
    Ritchie, Robert W.
    Gyoengy, Miklos
    Collin, James R. T.
    Leslie, Tom
    Coussios, Constantin-C.
    RADIOLOGY, 2012, 262 (01) : 252 - 261
  • [40] Real-Time Photoacoustic Thermometry Combined With Clinical Ultrasound Imaging and High-Intensity Focused Ultrasound
    Kim, Jeesu
    Choi, Wonseok
    Park, Eun-Yeong
    Kang, Youngnam
    Lee, Ki Jong
    Kim, Hyung Ham
    Kim, Won Jong
    Kim, Chulhong
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2019, 66 (12) : 3330 - 3338