Quantitative Ultrasound Imaging for Monitoring In Situ High-Intensity Focused Ultrasound Exposure

被引:17
|
作者
Ghoshal, Goutam [1 ,2 ]
Kemmerer, Jeremy P. [2 ]
Karunakaran, Chandra [2 ]
Abuhabsah, Rami [2 ]
Miller, Rita J. [2 ]
Sarwate, Sandhya [2 ]
Oelze, Michael L. [2 ]
机构
[1] Acoust MedSyst Inc, Savoy, IL USA
[2] Univ Illinois, Dept Elect & Comp Engn, Urbana, IL 61801 USA
关键词
quantitative ultrasound imaging; HIFU; effective scatterer diameter; effective scatterer concentration; parametric imaging; NONINVASIVE TEMPERATURE ESTIMATION; TISSUE TEMPERATURE; PROPAGATION SPEED; VIVO; ATTENUATION; BACKSCATTER; DEPENDENCE; FEASIBILITY; THERAPY; PHANTOM;
D O I
10.1177/0161734614524179
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Quantitative ultrasound (QUS) imaging is hypothesized to map temperature elevations induced in tissue with high spatial and temporal resolution. To test this hypothesis, QUS techniques were examined to monitor high-intensity focused ultrasound (HIFU) exposure of tissue. In situ experiments were conducted on mammary adenocarcinoma tumors grown in rats and lesions were formed using a HIFU system. A thermocouple was inserted into the tumor to provide estimates of temperature at one location. Backscattered time-domain waveforms from the tissue during exposure were recorded using a clinical ultrasonic imaging system. Backscatter coefficients were estimated using a reference phantom technique. Two parameters were estimated from the backscatter coefficient (effective scatterer diameter (ESD) and effective acoustic concentration (EAC). The changes in the average parameters in the regions corresponding to the HIFU focus over time were correlated to the temperature readings from the thermocouple. The changes in the EAC parameter were consistently correlated to temperature during both heating and cooling of the tumors. The changes in the ESD did not have a consistent trend with temperature. The mean ESD and EAC before exposure were 120 +/- 16 mm and 32 +/- 3 db/cm(3), respectively, and changed to 144 +/- 9 mm and 51 +/- 7 db/cm(3), respectively, just before the last HIFU pulse was delivered to the tissue. After the tissue cooled down to 37 degrees C, the mean ESD and EAC were 126 +/- 8 mm and 35 +/- 4 db/cm(3), respectively. Peak temperature in the range of 50-60 degrees C was recorded by a thermocouple placed just behind the tumor. These results suggest that QUS techniques have the potential to be used for non-invasive monitoring of HIFU exposure.
引用
收藏
页码:239 / 255
页数:17
相关论文
共 50 条
  • [21] An early history of high-intensity focused ultrasound
    O'Brien, William D., Jr.
    Dunn, Floyd
    PHYSICS TODAY, 2015, 68 (10) : 40 - 45
  • [22] MR imaging of prostate after treatment with high-intensity focused ultrasound
    Kirkham, Alexander P. S.
    Emberton, Mark
    Hoh, Ivan M.
    Illing, Rowland O.
    Freeman, A. Alex
    Allen, Clare
    RADIOLOGY, 2008, 246 (03) : 833 - 844
  • [23] A Review of High-Intensity Focused Ultrasound in Urology
    Cranston, David
    Leslie, Tom
    ter Haar, Gail
    CANCERS, 2021, 13 (22)
  • [24] High-Intensity Focused Ultrasound in Functional Neurosurgery
    Tyurnikov V.M.
    Gushcha A.O.
    Human Physiology, 2017, 43 (8) : 916 - 921
  • [25] High-intensity focused ultrasound for prostate cancer
    Napoli, Alessandro
    Alfieri, Giulia
    Scipione, Roberto
    Leonardi, Andrea
    Fierro, Davide
    Panebianco, Valeria
    De Nunzio, Cosimo
    Leonardo, Costantino
    Catalano, Carlo
    EXPERT REVIEW OF MEDICAL DEVICES, 2020, 17 (05) : 427 - 433
  • [26] High-intensity Focused Ultrasound: Ready for Primetime
    Rove, Kyle O.
    Sullivan, Kathryn F.
    Crawford, E. David
    UROLOGIC CLINICS OF NORTH AMERICA, 2010, 37 (01) : 27 - +
  • [27] Real-time visualization of high-intensity focused ultrasound treatment using ultrasound imaging
    Vaezy, S
    Shi, XG
    Martin, RW
    Chi, E
    Nelson, PI
    Bailey, MR
    Crum, LA
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (01): : 33 - 42
  • [28] Is high-intensity focused ultrasound therapy option?
    Weiss, Johannes
    ZEITSCHRIFT FUR GASTROENTEROLOGIE, 2014, 52 (05):
  • [29] Clinical applications of high-intensity focused ultrasound
    She, W. H.
    Cheung, T. T.
    Jenkins, Caroline R.
    Irwin, Michael G.
    HONG KONG MEDICAL JOURNAL, 2016, 22 (04) : 382 - 392
  • [30] TREATMENT OF GLAUCOMA WITH HIGH-INTENSITY FOCUSED ULTRASOUND
    BURGESS, SEP
    SILVERMAN, RH
    COLEMAN, DJ
    YABLONSKI, ME
    LIZZI, FL
    DRILLER, J
    ROSADO, A
    DENNIS, PH
    OPHTHALMOLOGY, 1986, 93 (06) : 831 - 838