Device and technique for extracorporeal blood volume sequestration during hemodialysis

被引:1
|
作者
Wimmer, Juergen
Bachler, Isolde
Haditsch, Bernd
Stadlbauer, Vanessa
Holzer, Herwig
Schneditz, Daniel
机构
[1] Graz Med Univ, Inst Physiol, Ctr Physiol Med, A-8010 Graz, Austria
[2] Graz Med Univ, Div Neurol, Dept Internal Med, A-8010 Graz, Austria
[3] Graz Med Univ, Div Gastroenterol, Dept Internal Med, A-8010 Graz, Austria
关键词
D O I
10.1097/01.mat.0000235812.55055.5b
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We developed a technique for a controlled and reversible volume perturbation of the cardiovascular system during hemodialysis. The capacitance of the extracorporeal circulation was modified by an expansion bag, using separate filling and draining lines connected to postpump and prepump sections of the arterial line segment, respectively. The connection to this bag was manually operated by three-way valves. The volume sequestered into the blood bag was continuously measured by a weighing scale. Filling and draining of the expansion bag was measured in eight patients in two subsequent routine dialysis treatments. Four volume shifts were done in each treatment. Filling and draining rates and times depended on extracorporeal blood flow as well as arterial and venous line pressures. Bag inflow and outflow volumes were 215.8 +/- 28.5 ml/min and 221.6 +/- 37.0 ml/min, respectively. The total volume transiently sequestered in the bag was 479.8 +/- 61.3 ml. The duration of the whole test was 4.5 +/- 0.8 minutes. During filling, residual dialyzer blood flow was transiently reduced to 58.5 +/- 21.6 ml/min and the filtration fraction reached 30 +/- 13%. There were no adverse events such as clotting in the blood bag or in the dialyzer. The method provides rapid, reversible, and safe volume shifts between the patient and the extracorporeal circulation with the potential to elicit a hemodynamic response for characterizing the patient during each dialysis treatment.
引用
收藏
页码:662 / 669
页数:8
相关论文
共 50 条
  • [41] Dynamic oxygenator blood volume during prolonged extracorporeal life support
    Hendrix, Rik H. J.
    Kurniawati, Eva R.
    Schins, Sanne F. C.
    Maessen, Jos G.
    Weerwind, Patrick W.
    PLOS ONE, 2022, 17 (02):
  • [42] FLUID VOLUME MONITORING WITH ELECTRICAL-IMPEDANCE TECHNIQUE DURING HEMODIALYSIS
    TEDNER, B
    LINS, LE
    ARTIFICIAL ORGANS, 1984, 8 (01) : 66 - 71
  • [43] EXTRACORPOREAL PRESSURE INCREASES DURING HEMODIALYSIS
    STEWART, WK
    FLEMING, LW
    HILL, A
    BIOMEDICAL ENGINEERING, 1973, 8 (06) : 257 - 259
  • [44] Effects of relative blood volume and ultrafiltration volume on pulse wave characteristics during hemodialysis
    Krenn, Simon
    Hecking, Manfred
    Mussnig, Sebastian
    Wassertheurer, Siegfried
    Mayer, Christopher
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2024, 39 : I2754 - I2756
  • [45] Effects of relative blood volume and ultrafiltration volume on pulse wave characteristics during hemodialysis
    Krenn, Simon
    Hecking, Manfred
    Mussnig, Sebastian
    Wassertheurer, Siegfried
    Mayer, Christopher
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2024, 39
  • [46] Influence of ultrafiltration volume on blood volume changes during hemodialysis as observed in day-of-the-week analysis of hemodialysis sessions
    Dasselaar, Judith J.
    de Jong, Paul E.
    Huisman, Roel M.
    Franssen, Casper F. M.
    ASAIO JOURNAL, 2007, 53 (04) : 479 - 484
  • [47] A BLOOD PROTEIN MONITOR FOR THE CONTINUOUS MEASUREMENT OF BLOOD-VOLUME CHANGES DURING HEMODIALYSIS
    SCHNEDITZ, D
    POGGLITSCH, H
    HORINA, J
    BINSWANGER, U
    KIDNEY INTERNATIONAL, 1990, 38 (02) : 342 - 346
  • [48] Design and Construct an Optical Device to Determine Relative Blood Volume in Patients Undergoing Hemodialysis
    Dormanesh, Banafshe
    Tofangchiha, Shahnaz
    Abouei, Vahid
    Sharifian, Hani
    IRANIAN RED CRESCENT MEDICAL JOURNAL, 2014, 16 (04)
  • [49] OPTIMIZING ULTRAFILTRATION RATE PROFILES FOR THE ESTIMATION OF BLOOD VOLUME DURING HEMODIALYSIS
    Hollot, C. V.
    Horowitz, J.
    Shrestha, R. P.
    Germain, M. G.
    Chait, Y.
    PROCEEDINGS OF THE ASME 8TH ANNUAL DYNAMIC SYSTEMS AND CONTROL CONFERENCE, 2015, VOL 1, 2016,
  • [50] BLOOD-VOLUME CHANGES DURING HEMODIALYSIS DETECTED BY CONDUCTIVITY MEASUREMENTS
    DEVRIES, PMJM
    KOUW, PM
    MEIJER, JH
    VISSER, V
    OE, PL
    SCHNEIDER, H
    DONKER, AJM
    CURRENT THERAPY IN NEPHROLOGY, 1989, 24 : 217 - 219