Reduction of breathing artifacts in multifrequency magnetic resonance elastography of the abdomen

被引:30
|
作者
Shahryari, Mehrgan [1 ]
Meyer, Tom [1 ]
Warmuth, Carsten [1 ]
Herthum, Helge [1 ]
Bertalan, Gergely [1 ]
Tzschatzsch, Heiko [1 ]
Stencel, Lisa [1 ]
Lukas, Steffen [1 ]
Lilaj, Ledia [1 ]
Braun, Jurgen [2 ]
Sack, Ingolf [1 ]
机构
[1] Charite Univ Med Berlin, Dept Radiol, Charitepl 1, D-10117 Berlin, Germany
[2] Charite Univ Med Berlin, Inst Med Informat, Berlin, Germany
关键词
abdomen; breathing artifacts; image registration; multifrequency MRE; navigator; stiffness; MR ELASTOGRAPHY; HEPATIC-FIBROSIS; LIVER STIFFNESS; TOMOELASTOGRAPHY; PANCREAS;
D O I
10.1002/mrm.28558
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose With abdominal magnetic resonance elastography (MRE) often suffering from breathing artifacts, it is recommended to perform MRE during breath-hold. However, breath-hold acquisition prohibits extended multifrequency MRE examinations and yields inconsistent results when patients cannot hold their breath. The purpose of this work was to analyze free-breathing strategies in multifrequency MRE of abdominal organs. Methods Abdominal MRE with 30, 40, 50, and 60 Hz vibration frequencies and single-shot, multislice, full wave-field acquisition was performed four times in 11 healthy volunteers: once with multiple breath-holds and three times during free breathing with ungated, gated, and navigated slice adjustment. Shear wave speed maps were generated by tomoelastography inversion. Image registration was applied for correction of intrascan misregistration of image slices. Sharpness of features was quantified by the variance of the Laplacian. Results Total scan times ranged from 120 seconds for ungated free-breathing MRE to 376 seconds for breath-hold examinations. As expected, free-breathing MRE resulted in larger organ displacements (liver, 4.7 +/- 1.5 mm; kidneys, 2.4 +/- 2.2 mm; spleen, 3.1 +/- 2.4 mm; pancreas, 3.4 +/- 1.4 mm) than breath-hold MRE (liver, 0.7 +/- 0.2 mm; kidneys, 0.4 +/- 0.2 mm; spleen, 0.5 +/- 0.2 mm; pancreas, 0.7 +/- 0.5 mm). Nonetheless, breathing-related displacement did not affect mean shear wave speed, which was consistent across all protocols (liver, 1.43 +/- 0.07 m/s; kidneys, 2.35 +/- 0.21 m/s; spleen, 2.02 +/- 0.15 m/s; pancreas, 1.39 +/- 0.15 m/s). Image registration before inversion improved the quality of free-breathing examinations, yielding no differences in image sharpness to uncorrected breath-hold MRE in most organs (P > .05). Conclusion Overall, multifrequency MRE is robust to breathing when considering whole-organ values. Respiration-related blurring can readily be corrected using image registration. Consequently, ungated free-breathing MRE combined with image registration is recommended for multifrequency MRE of abdominal organs.
引用
收藏
页码:1962 / 1973
页数:12
相关论文
共 50 条
  • [1] Free-breathing motion corrected magnetic resonance elastography of the abdomen
    van Schelt, Anne-Sophie
    Wassenaar, Nienke Petronella Maria
    Runge, Jurgen Henk
    Nelissen, Jules Laurent
    van Laarhoven, Hanneke Wilma Marlies
    Stoker, Jaap
    Nederveen, Aart Johannes
    Schrauben, Eric Mathew
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2024, 14 (05) : 3447 - 3460
  • [2] Magnetic resonance elastography of abdomen
    Sudhakar Kundapur Venkatesh
    Richard L. Ehman
    Abdominal Imaging, 2015, 40 : 745 - 759
  • [3] Magnetic resonance elastography of abdomen
    Venkatesh, Sudhakar Kundapur
    Ehman, Richard L.
    ABDOMINAL IMAGING, 2015, 40 (04): : 745 - 759
  • [4] Multifrequency inversion in magnetic resonance elastography
    Papazoglou, Sebastian
    Hirsch, Sebastian
    Braun, Juergen
    Sack, Ingolf
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (08): : 2329 - 2346
  • [5] Reduction of peristaltic artifacts on magnetic resonance imaging of the abdomen: A comparative evaluation of three drugs
    MartiBonmati, L
    Graells, M
    RoncheraOms, CL
    ABDOMINAL IMAGING, 1996, 21 (04): : 309 - 313
  • [6] Multifrequency Magnetic Resonance Elastography for the Assessment of Renal Allograft Function
    Garcia, Stephan Rodrigo Marticorena
    Fischer, Thomas
    Duerr, Michael
    Gueltekin, Emin
    Braun, Juergen
    Sack, Ingolf
    Guo, Jing
    INVESTIGATIVE RADIOLOGY, 2016, 51 (09) : 591 - 595
  • [7] In vivo multifrequency magnetic resonance elastography of the human intervertebral disk
    Streitberger, Kaspar-Josche
    Diederichs, Gerd
    Guo, Jing
    Fehlner, Andreas
    Hamm, Bernd
    Braun, Juergen
    Sack, Ingolf
    MAGNETIC RESONANCE IN MEDICINE, 2015, 74 (05) : 1380 - 1387
  • [8] Optimization of a Multifrequency Magnetic Resonance Elastography Protocol for the Human Brain
    Kurt, Mehmet
    Wu, Lyndia
    Laksari, Kaveh
    Ozkaya, Efe
    Suar, Zeynep M.
    Lv, Han
    Epperson, Karla
    Epperson, Kevin
    Sawyer, Anne M.
    Camarillo, David
    Pauly, Kim Butts
    Wintermark, Max
    JOURNAL OF NEUROIMAGING, 2019, 29 (04) : 440 - 446
  • [9] Feasibility and Reproducibility of Multifrequency Magnetic Resonance Elastography in Healthy and Diseased Pancreases
    Song, Qike
    Shi, Yu
    Gao, Feng
    Meng, Yin
    Yang, Rui
    Liu, Yuanyuan
    Zhong, Shiling
    Hong, Yang
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2022, 56 (06) : 1769 - 1780
  • [10] REDUCTION OF ARTIFACTS FROM BREATHING AND PERISTALSIS IN PHASE-CONTRAST MRA OF THE CHEST AND ABDOMEN
    DUMOULIN, CL
    STEINBERG, FL
    YUCEL, EK
    DARROW, RD
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1993, 17 (02) : 328 - 332