Influence of Implant Dimensions in the Resorbed and Bone Augmented Mandible: A Finite Element Study

被引:4
|
作者
Agop Forna, Doriana [1 ]
Forna, Norina Consuela [2 ]
Butnaru Moldoveanu, Sinziana Anca [3 ]
机构
[1] Univ Med & Pharm Grigore T Popa, Dept Dentoalveolar & Maxillofacial Surg, Fac Dent Med, Iasi, Romania
[2] Univ Med & Pharm Grigore T Popa, Dept Implantol, Removable Dentures, Dentures Technol,Fac Dent Med, Iasi, Romania
[3] Univ Med & Pharm Grigore T Popa, Fac Med Bioengn, Dept Biomed Sci, 9-16 M Kogalniceanu St, Iasi 700454, Romania
关键词
Bone augmentation; bone resorption; finite element analysis; implant dimensions; peri-implant stress; STRESS-DISTRIBUTION; DENTAL IMPLANTS; FOLLOW-UP; DIAMETER; LENGTH; PARAMETERS; PROSTHESES;
D O I
10.4103/ccd.ccd_366_19
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aims: The scope of this study was to analyze the influence of clinically feasible implant diameter and length on the stress transmitted to the peri-implant bone in the case of a resorbed and bone augmented mandible through finite element analysis. Settings and Design: The study was carried out in silico. Subjects and Methods: Resorbed and bone-augmented 3D models were derived from in vivo cone-beam computed tomography scans of the same patient. Corresponding implant systems were modeled with the diameter ranging from 3.3 to 6 mm and length ranging from 5 to 13 mm, and masticatory loads were applied on the abutment surface. Statistical Analysis Used: None. Results: In the bone augmented ridge, maximum stress values in the peri-implant region drastically decreased only when using implants of a diameter of 5 mm and 6 mm. Implants up to 4 mm in diameter led to comparable stress values with the ones obtained in the resorbed ridge, when using the larger implants. The increase of length reduced stress in the resorbed mandible, whereas in the bone augmented model, it led to small variations only in implants up to 4 mm in diameter. Conclusions: It was concluded that bone augmentation provides the optimal framework for clinicians to use larger implants, which, in turn, reduces stress in the peri-implant region. Diameter and length play an equally important role in decreasing stress. Implant dimensions should be carefully considered with ridge geometry.
引用
收藏
页码:336 / 341
页数:6
相关论文
共 50 条
  • [21] Finite Element Analysis of Dental Implant Neck Effects on Primary Stability and Osseointegration in a Type IV Bone Mandible
    Huang, You-Min
    Chou, I-Chiang
    Jiang, Cho-Pei
    Wu, Yi-Syun
    Lee, Shyh-Yuan
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2014, 24 (01) : 1407 - 1415
  • [22] Influence of Implant/Abutment Connection on Stress Distribution to Implant-Surrounding Bone: A Finite Element Analysis
    Hanaoka, Marcia
    Alexandre Gehrke, Sergio
    Mardegan, Fabio
    Gennari, Cesar Roberto
    Taschieri, Silvio
    Del Fabbro, Massimo
    Corbella, Stefano
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2014, 23 (07): : 565 - 571
  • [23] Answer to the letter addressed to Editor concerning "Biomechanical study of mandible bone supporting a four-implant retained bridge. Finite element analysis of the influence of bone anisotropy and foodstuff position"
    Bonnet, A. S.
    Postaire, M.
    Lipinski, P.
    MEDICAL ENGINEERING & PHYSICS, 2010, 32 (04): : 415 - 416
  • [24] Finite Element Analysis (FEA) for Influence of Variation in Dental Implant Dimensions (Length and Diameter) on Peri-implant Bone Stress/Strain Distribution: A Systematic Review
    Sayed, Arshad Jamal
    Mohsin, Syed Fareed
    Garg, Kamlesh Kumar
    Agwan, Muhammad Atif Saleem
    Tareen, Sabahat Ullah Khan
    Alruthea, Mohammed Sulaiman
    Mohan, Sasankoti
    PAKISTAN JOURNAL OF MEDICAL SCIENCES, 2025, 41 (01) : 318 - 330
  • [25] Experimental and finite element study of a human mandible
    Vollmer, D
    Meyer, U
    Joos, U
    Vègh, A
    Piffkò, J
    JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY, 2000, 28 (02) : 91 - 96
  • [26] Experimental and finite element study of a human mandible
    Vollmer, D
    Meyer, U
    Piffko, J
    Bourauel, C
    Joos, U
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING - 3, 2001, : 655 - 660
  • [27] Comparing the Distribution of Strains With the Distribution of Bone Tissue in a Human Mandible: A Finite Element Study
    Groening, Flora
    Fagan, Michael
    O'Higgins, Paul
    ANATOMICAL RECORD-ADVANCES IN INTEGRATIVE ANATOMY AND EVOLUTIONARY BIOLOGY, 2013, 296 (01): : 9 - 18
  • [28] Restoration of resorbed mandible using resorbable membrane and bone graft material with simultaneous implant placement: Report of two cases
    Park, Jun-Beom
    QUINTESSENCE INTERNATIONAL, 2010, 41 (01): : 21 - 25
  • [29] Study of a constrained finite element elbow prosthesis: the influence of the implant placement
    Nalbone, Lorenzo
    Monac, Francesco
    Nalbone, Luca
    Ingrassia, Tommaso
    Ricotta, Vito
    Nigrelli, Vincenzo
    Ferruzza, Massimo
    Tarallo, Luigi
    Porcellini, Giuseppe
    Camarda, Lawrence
    JOURNAL OF ORTHOPAEDICS AND TRAUMATOLOGY, 2023, 24 (01)
  • [30] Study of a constrained finite element elbow prosthesis: the influence of the implant placement
    Lorenzo Nalbone
    Francesco Monac
    Luca Nalbone
    Tommaso Ingrassia
    Vito Ricotta
    Vincenzo Nigrelli
    Massimo Ferruzza
    Luigi Tarallo
    Giuseppe Porcellini
    Lawrence Camarda
    Journal of Orthopaedics and Traumatology, 24