Study of a constrained finite element elbow prosthesis: the influence of the implant placement

被引:1
|
作者
Nalbone, Lorenzo [1 ]
Monac, Francesco [1 ]
Nalbone, Luca [2 ]
Ingrassia, Tommaso [3 ]
Ricotta, Vito [3 ]
Nigrelli, Vincenzo [3 ]
Ferruzza, Massimo [1 ]
Tarallo, Luigi [4 ]
Porcellini, Giuseppe [4 ]
Camarda, Lawrence [1 ]
机构
[1] Univ Palermo, Dept Orthoped & Traumatol DICHIRONS, Via Vespro, I-90100 Palermo, Italy
[2] Univ Messina, Dept Vet Sci, Messina, Italy
[3] Univ Palermo, Dept Engn, Palermo, Italy
[4] Univ Modena, Dept Orthoped & Traumatol, Modena, Italy
关键词
Totel elbow arthroplasty; Biomechanics; Elbow replacement; Prosthetic posizioning; Elbow finite element; ARTHROPLASTY; REPLACEMENT;
D O I
10.1186/s10195-023-00690-x
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
BackgroundThe functional results of total elbow arthroplasty (TEA) are controversial and the medium- to long-term revision rates are relatively high. The aim of the present study was to analyze the stresses of TEA in its classic configuration, identify the areas of greatest stress in the prosthesis-bone-cement interface, and evaluate the most wearing working conditions.Materials and methodsBy means of a reverse engineering process and using a 3D laser scanner, CAD (computer-aided drafting) models of a constrained elbow prosthesis were acquired. These CAD models were developed and their elastic properties, resistance, and stresses were studied through finite element analysis (finite element method-FEM). The obtained 3D elbow-prosthesis model was then evaluated in cyclic flexion-extension movements (> 10 million cycles). We highlighted the configuration of the angle at which the highest stresses and the areas most at risk of implant mobilization develop. Finally, we performed a quantitative study of the stress state after varying the positioning of the stem of the ulnar component in the sagittal plane by +/- 3 degrees.ResultsThe greatest von Mises stress state in the bone component for the 90 degrees working configuration was 3.1635 MPa, which occurred in the most proximal portion of the humeral blade and in the proximal middle third of the shaft. At the ulnar level, peaks of 4.1763 MPa were recorded at the proximal coronoid/metaepiphysis level. The minimum elastic resistance and therefore the greatest stress states were recorded in the bone region at the apex of the ulnar stem (0.001967 MPa). The results of the analysis for the working configurations at 0 degrees and 145 degrees showed significant reductions in the stress states for both prosthetic components; similarly, varying the positioning of the ulnar component at 90 degrees (- 3 degrees in the sagittal plane, 0 degrees in the frontal plane) resulted in better working conditions with a greater resulting developed force and a lower stress peak in the ulnar cement.ConclusionThe areas of greatest stress occur in specific regions of the ulnar and humeral components at the bone-cement-prosthesis interface. The heaviest configuration in terms of stresses was when the elbow was flexed at 90 degrees. Variations in the positioning in the sagittal plane can mechanically affect the movement, possibly resulting in longer survival of the implant.Level of evidence: 5
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Study of a constrained finite element elbow prosthesis: the influence of the implant placement
    Lorenzo Nalbone
    Francesco Monac
    Luca Nalbone
    Tommaso Ingrassia
    Vito Ricotta
    Vincenzo Nigrelli
    Massimo Ferruzza
    Luigi Tarallo
    Giuseppe Porcellini
    Lawrence Camarda
    Journal of Orthopaedics and Traumatology, 24
  • [2] Finite Element Analysis of the Influence of the Offset Placement of an Implant-Supported Prosthesis on Bone Stress Distribution
    Anitua, E.
    Orive, G.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2009, 89B (02) : 275 - 281
  • [3] Effects of cementless fixation of implant prosthesis: A finite element study
    Lee, Hyeonjong
    Park, Soyeon
    Kwon, Kung-Rock
    Noh, Gunwoo
    JOURNAL OF ADVANCED PROSTHODONTICS, 2019, 11 (06): : 341 - 349
  • [4] Stress Distribution in Implant Retained Finger Prosthesis: A Finite Element Study
    Amornvit, Pokpong
    Rokaya, Dinesh
    Keawcharoen, Konrawee
    Thongpuisawasdi, Nimit
    JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH, 2013, 7 (12) : 2851 - 2854
  • [5] Modeling and finite element analysis of a new revision implant for the elbow
    Herren, DB
    Ploeg, H
    Hertig, D
    Klabunde, R
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (420) : 292 - 297
  • [6] Biomechanical assessment of unconstrained elbow prosthesis after total elbow replacement: A finite element analysis
    Heidari, Milad
    Kadir, Mohammed Rafiq Abdul
    Fallahiarezoodar, Ali
    Harun, Muhamad Noor
    Alizadeh, Mina
    Kashani, Jamal
    APPLIED MECHANICS AND INDUSTRIAL TECHNOLOGIES, 2012, 234 : 7 - +
  • [7] Evaluation of Novel Prosthesis Implant Adapter by Finite Element Analysis
    Mohammed, Marwa J.
    Hamandi, Sadiq J.
    Arbilei, Marwan N.
    JOURNAL OF BIOMIMETICS BIOMATERIALS AND BIOMEDICAL ENGINEERING, 2023, 61 : 109 - 120
  • [8] Finite-Element Analysis of Stress on Dental Implant Prosthesis
    Rubo, Jose Henrique
    Capello Souza, Edson Antonio
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2010, 12 (02) : 105 - 113
  • [9] Motion Simulation And Finite Element Analysis Of Knee Prosthesis With Implant
    Fang, Zhibin
    Zhang, Shaobin
    Cheng, Jiamei
    Li, Shaoming
    JOURNAL OF APPLIED SCIENCE AND ENGINEERING, 2025, 28 (04): : 667 - 679
  • [10] Biomechanical Performance of Different Implant Spacings and Placement Angles in Partial Fixed Denture Prosthesis Restorations: A Finite Element Analysis
    Zhang, Jianguo
    Hou, Hu
    Chen, Peng
    Song, Liang
    Hu, Fengling
    Yu, Youcheng
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2024, 44 (05) : 685 - 695