Detection and Measurement of Stones With Ultrasound Strain Elastography A Phantom Study

被引:3
|
作者
Li, Qian [1 ]
Chen, Lei [1 ]
Halpern, Elkan F. [2 ]
Samir, Anthony E. [3 ]
机构
[1] Peking Univ, Hlth Sci Ctr, Hosp 1, Dept Ultrasound, Beijing 100871, Peoples R China
[2] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Inst Technol Assessment, Boston, MA USA
[3] Harvard Univ, Massachusetts Gen Hosp, Dept Radiol, Sch Med, 55 Fruit St, Boston, MA 02114 USA
关键词
ultrasonography; strain elastography; stone; horizontal dimension; vertical dimension; TIME TISSUE ELASTOGRAPHY; SOLID BREAST MASSES; DIAGNOSIS; BENIGN; SONOGRAPHY; MANAGEMENT; FIBROSIS;
D O I
10.1097/RUQ.0000000000000204
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The sonoelastographic appearances of stones in a phantom were evaluated in this study. Ten stones were embedded into a tissue-mimicking meat phantom. The stone axial (vertical) and transverse (horizontal) dimensions measured by an electronic digital caliper, gray-scale ultrasound, and strain elastography (SE) were compared in 5 groups with stones embedded at different depths. In this study, physically measured axial and transverse stone dimensions were 1.17 to 6.86 and 1.30 to 11.15 mm, respectively. Strain elastography showed a characteristic 3-layer pattern associated with stones, comprising a superficial transition region, a hard region, and a deep transition region. As SE data were available in group 5, only data of groups 1 to 4 were analyzed. Compared with physical measurements, measurement mean errors of SE horizontal and SE vertical dimensions ranged from -0.20 to 0.42 mm and from -1.28 to -0.05 mm, respectively, in the 4 groups. Paired t testing demonstrated a significant horizontal dimension measurement error difference between B mode and SE method in group 4 (0.44 vs -0.20 mm, P < 0.05; F = 1.18, P > 0.05), but not in the other groups. Strain elastography horizontal dimension measurement error was not statistically correlated with stone size in the 4 groups. Strain elastography vertical dimension measurement error significantly correlated with stone size only in group 4 (P < 0.05). Preliminary results indicate that stone horizontal and vertical dimensions can be measured using SE in a soft tissue phantom, including when shadowing precludes measurement of vertical dimension on conventional 2-dimensional ultrasound. These results provide substantial motivation to further investigate SE as a modality to image stones in clinical practice.
引用
收藏
页码:272 / 278
页数:7
相关论文
共 50 条
  • [31] Reply to: "Strain ultrasound elastography for liver diseases''
    Van Beers, Bernard E.
    JOURNAL OF HEPATOLOGY, 2015, 63 (02) : 535 - 535
  • [32] A LUNG PHANTOM MODEL TO STUDY PULMONARY EDEMA USING LUNG ULTRASOUND SURFACE WAVE ELASTOGRAPHY
    Zhou, Jinling
    Zhang, Xiaoming
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2018, 44 (11): : 2400 - 2405
  • [33] Ultrasound elastography for carpal tunnel pressure measurement: A cadaveric validation study
    Kubo, Kazutoshi
    Zhou, Boran
    Cheng, Yu-Shiuan
    Yang, Tai-Hua
    Qiang, Bo
    An, Kai-Nan
    Moran, Steven L.
    Amadio, Peter C.
    Zhang, Xiaoming
    Zhao, Chunfeng
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2018, 36 (01) : 477 - 483
  • [34] A two-step optical flow method for strain estimation in elastography: Simulation and phantom study
    Pan, Xiaochang
    Gao, Jing
    Tao, Shengzhen
    Liu, Ke
    Bai, Jing
    Luo, Jianwen
    ULTRASONICS, 2014, 54 (04) : 990 - 996
  • [35] Ultrasound detection of the skull-brain interface: A phantom study
    de Souza-Daw, Tony
    Lewis, Philip
    Junor, Paul
    Stewart, Robert
    Maller, Jerome
    Manasseh, Richard
    2012 INTERNATIONAL CONFERENCE ON CONTROL, AUTOMATION AND INFORMATION SCIENCES (ICCAIS), 2012, : 181 - 186
  • [36] In Vivo Measurement of Rotator Cuff Tendon Strain With Ultrasound Elastography An Investigation Using a Porcine Model
    Hatta, Taku
    Yamamoto, Nobuyuki
    Sano, Hirotaka
    Itoi, Eiji
    JOURNAL OF ULTRASOUND IN MEDICINE, 2014, 33 (09) : 1641 - 1646
  • [37] Strain elastography with Ultrasound Computer Tomography: a simulation study based on biomechanical models
    Hopp, T.
    Ruiter, N., V
    MEDICAL IMAGING 2021: ULTRASONIC IMAGING AND TOMOGRAPHY, 2021, 11602
  • [38] 3D ultrasound elastography for early detection of lesions. Evaluation on a pressure ulcer mimicking phantom
    Deprez, Jean-Francois
    Cloutier, Guy
    Schmitt, Cedric
    Gehin, Claudine
    Dittmar, Andre
    Basset, Olivier
    Brusseau, Elisabeth
    2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, : 79 - +
  • [39] Effect of Depth on Ultrasound Point Shear Wave Elastography in an Elasticity Phantom
    Almutairi, Fahad F.
    Abdeen, Rawan
    Alyami, Jaber
    Sultan, Salahaden R.
    APPLIED SCIENCES-BASEL, 2022, 12 (13):
  • [40] Evaluation of ultrasound point shear wave elastography reliability in an elasticity phantom
    Sultan, Salahaden R.
    Alghamdi, Amer
    Abdeen, Rawan
    Almutairi, Fahad
    ULTRASONOGRAPHY, 2022, 41 (02) : 291 - 297