Performance Evaluation of microPET: A High-Resolution Lutetium Oxyorthosilicate PET Scanner for Animal Imaging

被引:1
|
作者
Chatziioannou, Anion F.
Cherry, Simon R.
Shao, Yiping
Silverman, Robert W.
Meadors, Ken
Farquhar, Thomas H.
Pedarsani, Marjan
Phelps, Michael E.
机构
[1] Crump Inst Biol Imaging, Dept Mol & Med Pharmacol, Los Angeles, CA USA
[2] Univ Calif Los Angeles, Sch Med, Lab Struct Biol & Mol Med, US DOE, Los Angeles, CA USA
关键词
PET; instrumentation; performance evaluation; animal imaging; COMPUTED-TOMOGRAPHY; SPATIAL-RESOLUTION; 3D PET; DESIGN; SYSTEM; MODEL; QUANTITATION; SENSITIVITY; CRYSTALS; DETECTOR;
D O I
10.2967/jnumed.120.251280a
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new dedicated PET scanner, microPET, was designed and developed at the University of California, Los Angeles, for imaging small laboratory animals. The goal was to provide a compact system with superior spatial resolution at a fraction of the cost of a clinical PET scanner. Methods: The system uses fiberoptic readout of individually cut lutetium oxyorthosilicate (LSO) crystals to achieve high spatial resolution. Each microPET detector consists of an 8 x 8 array of 2 x 2 x 10-mm LSO scintillation crystals that are coupled to a 64-channel photomultiplier tube by optical fibers. The tomograph consists of 30 detectors in a continuous ring with a 17.2-cm diameter and fields of view (FOVs) of 11.25 cm in the transaxial direction and 1.8 cm in the axial direction. The system has eight crystal rings and no interplane septa. It operates exclusively in the three-dimensional mode and has an electronically controlled bed that is capable of wobbling with a radius of 300 pm. We describe the performance of the tomograph in terms of its spatial, energy and timing resolution, as well as its sensitivity and counting-rate performance. We also illustrate its overall imaging performance with phantom and animal studies that demonstrate the potential applications of this device to biomedical research. Results: Images reconstructed with three-dimensional filtered backprojection show a spatial resolution of 1.8 mm at the center of the FOV (CFOV), which remains <2.5 mm for the central 5 cm of the transaxial FOV. The resulting volumetric resolution of the system is <8 mu L. The absolute system sensitivity measured with a 0.74 MBq (20 mu Ci) Ge-68 point source at the CFOV is 5.62 Hz/kBq. The maximum noise equivalent counting rate obtained with a 6.4-cm diameter cylinder spanning the central 56% of the FOV is 10 kcps, whereas the scatter fraction is 37% at the CFOV for an energy window of 250-650 keV and the same diameter cylinder. Conclusion: This is the first PET scanner to use the new scintillator LSO and uses a novel detector design to achieve high volumetric spatial resolution. The combination of imaging characteristics of this prototype system (resolution, sensitivity, counting-rate performance and scatter fraction) opens up new possibilities in the study of animal models with PET.
引用
收藏
页码:155S / 165S
页数:11
相关论文
共 50 条
  • [21] Performance evaluation of the microPET P4: a PET system dedicated to animal imaging
    Tai, YC
    Chatziioannou, A
    Siegel, S
    Young, J
    Newport, D
    Goble, RN
    Nutt, RE
    Cherry, SR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (07): : 1845 - 1862
  • [22] Performance evaluation of the microPET R4 PET scanner for rodents
    Knoess, C
    Siegel, S
    Smith, A
    Newport, D
    Richerzhagen, R
    Winkeler, A
    Jacobs, A
    Goble, RN
    Graf, R
    Wienhard, K
    Heiss, WD
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (05) : 737 - 747
  • [23] Lutetium oxyorthosilicate (LSO) intrinsic activity correction and minimal detectable target activity study for SPECT imaging with a LSO-based animal PET scanner
    Yao, Rutao
    Ma, Tianyu
    Shao, Yiping
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (16): : 4399 - 4415
  • [24] Performance evaluation of a high resolution dedicated breast PET scanner
    Garcia Hernandez, Trinitat
    Vicedo Gonzalez, Aurora
    Ferrer Rebolleda, Jose
    Sanchez Jurado, Raul
    Rosello Ferrando, Joan
    Brualla Gonzalez, Luis
    Granero Cabanero, Domingo
    Cozar Santiago, Maria del Puig
    [J]. MEDICAL PHYSICS, 2016, 43 (05) : 2261 - 2272
  • [25] Performance evaluation of a high-resolution large field-of-view digital PET/CT scanner
    Stolk, Sophie
    Archibald, Zane
    Butterfield, Regan
    Covington, Matthew
    Leung, Edwin
    Sun, Xishan
    Yap, Jeffrey
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2023, 64
  • [26] Performance evaluation of the microPET R4 PET scanner for rodents
    Christof Knoess
    Stefan Siegel
    Anne Smith
    Danny Newport
    Norbert Richerzhagen
    Alexandra Winkeler
    Andreas Jacobs
    Rhonda N. Goble
    Rudolf Graf
    Klaus Wienhard
    Wolf-Dieter Heiss
    [J]. European Journal of Nuclear Medicine and Molecular Imaging, 2003, 30 : 737 - 747
  • [27] Performance evaluation of Microinsert II - A submillimeter resolution small animal PET scanner
    Ravindranath, Bosky
    Wen, Jie
    Mathews, Aswin
    Tomov, Dimitre
    Catherall, David
    Li, Ke
    Wang, Qiang
    Komarov, Sergey
    O'Sullivan, Joseph
    Tai, Yuan-chuan
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [28] Performance evaluation of Microinsert II - A submillimeter resolution small animal PET scanner
    Ravindranath, Bosky
    Wen, Jie
    Mathews, Aswin
    Tomov, Dimitre
    Catherall, David
    Li, Ke
    Wang, Qiang
    Komarov, Sergey
    O'Sullivan, Joseph
    Tai, Yuan-chuan
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [29] MicroPET II: design, development and initial performance of an improved microPET scanner for small-animal imaging
    Tai, YC
    Chatziioannou, AF
    Yang, YF
    Silverman, RW
    Meadors, K
    Siegel, S
    Newport, DF
    Stickel, JR
    Cherry, SR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (11): : 1519 - 1537
  • [30] High-resolution 3D Bayesian image reconstruction using the microPET small-animal scanner
    Qi, JY
    Leahy, RM
    Cherry, SR
    Chatziioannou, A
    Farquhar, TH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (04): : 1001 - 1013