Performance Evaluation of microPET: A High-Resolution Lutetium Oxyorthosilicate PET Scanner for Animal Imaging

被引:1
|
作者
Chatziioannou, Anion F.
Cherry, Simon R.
Shao, Yiping
Silverman, Robert W.
Meadors, Ken
Farquhar, Thomas H.
Pedarsani, Marjan
Phelps, Michael E.
机构
[1] Crump Inst Biol Imaging, Dept Mol & Med Pharmacol, Los Angeles, CA USA
[2] Univ Calif Los Angeles, Sch Med, Lab Struct Biol & Mol Med, US DOE, Los Angeles, CA USA
关键词
PET; instrumentation; performance evaluation; animal imaging; COMPUTED-TOMOGRAPHY; SPATIAL-RESOLUTION; 3D PET; DESIGN; SYSTEM; MODEL; QUANTITATION; SENSITIVITY; CRYSTALS; DETECTOR;
D O I
10.2967/jnumed.120.251280a
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new dedicated PET scanner, microPET, was designed and developed at the University of California, Los Angeles, for imaging small laboratory animals. The goal was to provide a compact system with superior spatial resolution at a fraction of the cost of a clinical PET scanner. Methods: The system uses fiberoptic readout of individually cut lutetium oxyorthosilicate (LSO) crystals to achieve high spatial resolution. Each microPET detector consists of an 8 x 8 array of 2 x 2 x 10-mm LSO scintillation crystals that are coupled to a 64-channel photomultiplier tube by optical fibers. The tomograph consists of 30 detectors in a continuous ring with a 17.2-cm diameter and fields of view (FOVs) of 11.25 cm in the transaxial direction and 1.8 cm in the axial direction. The system has eight crystal rings and no interplane septa. It operates exclusively in the three-dimensional mode and has an electronically controlled bed that is capable of wobbling with a radius of 300 pm. We describe the performance of the tomograph in terms of its spatial, energy and timing resolution, as well as its sensitivity and counting-rate performance. We also illustrate its overall imaging performance with phantom and animal studies that demonstrate the potential applications of this device to biomedical research. Results: Images reconstructed with three-dimensional filtered backprojection show a spatial resolution of 1.8 mm at the center of the FOV (CFOV), which remains <2.5 mm for the central 5 cm of the transaxial FOV. The resulting volumetric resolution of the system is <8 mu L. The absolute system sensitivity measured with a 0.74 MBq (20 mu Ci) Ge-68 point source at the CFOV is 5.62 Hz/kBq. The maximum noise equivalent counting rate obtained with a 6.4-cm diameter cylinder spanning the central 56% of the FOV is 10 kcps, whereas the scatter fraction is 37% at the CFOV for an energy window of 250-650 keV and the same diameter cylinder. Conclusion: This is the first PET scanner to use the new scintillator LSO and uses a novel detector design to achieve high volumetric spatial resolution. The combination of imaging characteristics of this prototype system (resolution, sensitivity, counting-rate performance and scatter fraction) opens up new possibilities in the study of animal models with PET.
引用
收藏
页码:155S / 165S
页数:11
相关论文
共 50 条
  • [41] MicroPET II: An ultra-high resolution small animal PET system
    Tai, YC
    Chatziioannou, AF
    Silverman, RW
    Meadors, K
    Siegel, S
    Newport, DF
    Yang, Y
    Stickel, J
    Cherry, SR
    [J]. 2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 1848 - 1852
  • [42] Metric performance of a high-resolution laser scanner
    Gordon, S
    Lichti, D
    Stewart, M
    Tsakiri, M
    [J]. VIDEOMETRICS AND OPTICAL METHODS FOR 3D SHAPE MEASUREMENT, 2001, 4309 : 174 - 184
  • [43] High-Resolution Adaptive PET Imaging
    Zhou, Jian
    Qi, Jinyi
    [J]. INFORMATION PROCESSING IN MEDICAL IMAGING, PROCEEDINGS, 2009, 5636 : 26 - 37
  • [44] Preliminary performance evaluation of a high resolution small animal PET scanner with monolithic crystals and depth-of-interaction encoding
    Balcerzyk, Marcin
    Kontaxakis, George
    Delgado, Mercedes
    Garcia, Luis
    Benlloch, Jose M.
    Pozo, Miguel A.
    [J]. 8TH IEEE INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOENGINEERING, VOLS 1 AND 2, 2008, : 944 - +
  • [45] First results from a high-resolution small animal PET insert for PET/MRI imaging
    Andrew Goertzen
    Greg Stortz
    Jonathan Thiessen
    Daryl Bishop
    Muhammad Salman Khan
    Piotr Kozlowski
    Fabrice Retiere
    Graham Schellenberg
    Vesna Sossi
    Christopher Thompson
    [J]. EJNMMI Physics, 2 (Suppl 1)
  • [46] Performance Evaluation of SensL SiPM Arrays for High-Resolution PET
    Thiessen, Jonathan D.
    Jackson, Carl
    O'Neill, Kevin
    Bishop, Daryl
    Kozlowski, Piotr
    Retiere, Fabrice
    Shams, Ehsan
    Stortz, Greg
    Thompson, Christopher J.
    Goertzen, Andrew L.
    [J]. 2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [47] Performance Evaluation of a High-resolution TOF Clinical PET/CT
    Xu, Baixuan
    Liu, Changbin
    Dong, Yun
    Tang, Renming
    Liu, Yachao
    Yang, Hui
    Chen, Mu
    Li, Can
    Shen, Zhihui
    Dong, Yanliang
    Bi, Xiao
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2016, 57
  • [48] Comparison of 3-D maximum a posteriori and filtered backprojection algorithms for high-resolution animal imaging with microPET
    Chatziioannou, A
    Qi, J
    Moore, A
    Annala, A
    Nguyen, K
    Leahy, R
    Cherry, SR
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (05) : 507 - 512
  • [49] Evaluation of transmission methodology for the microPET Focus 220 animal scanner
    Lehnert, Wencke
    Meikle, Steven R.
    Siegel, Stefan
    Bailey, David
    Banati, Richard
    Rosenfeld, Anatoly B.
    [J]. 2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2519 - 2523
  • [50] LabPET II Pixelated APD-Based PET Scanner for High-Resolution Preclinical Imaging
    Lecomte, Roger
    Arpin, Louis
    Beaudoin, Jean-Francois
    Bergeron, Melanie
    Bouchard, Jonathan
    Bouziri, Haithem
    Cadorette, Jules
    Gaudin, Emilie
    Jurgensen, Nadia
    Koua, Konin Calliste
    Trepanier, Pierre-Yves Lauzier
    Leroux, Jean-Daniel
    Loignon-Houle, Francis
    Njejimana, Larissa
    Paille, Maxime
    Paulin, Caroline
    Pepin, Catherine
    Pratte, Jean-Francois
    Samson, Arnaud
    Thibaudeau, Christian
    Viscogliosi, Nicolas
    Fontaine, Rejean
    [J]. MEDICAL PHYSICS, 2016, 43 (08) : 4935 - 4935