Influence of the orientation of the Osstell® transducer during measurement of dental implant stability using resonance frequency analysis: A numerical approach

被引:20
|
作者
Capek, Lukas [1 ]
Simunek, Antonin [2 ]
Slezak, Radovan [2 ]
Dzan, Ladislav [3 ]
机构
[1] Tech Univ Liberec, Dept Mech Engn, Liberec 46117 1, Czech Republic
[2] Charles Univ Prague, Fac Med Hradec Kralove, Dept Dent, Hradec Kralove 50001, Czech Republic
[3] Reg Hosp Liberec, Dept Maxillofacial Surg, Liberec 46001, Czech Republic
关键词
Resonance frequency analysis; Dental implants; Stability; Finite element analysis; Resonance frequency; Eigenfrequency; DAMPING CAPACITY ASSESSMENT; IN-VITRO; INTERFACE; BEHAVIOR; SURGERY; SITES; MODEL; TIBIA;
D O I
10.1016/j.medengphy.2009.02.003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Stability of dental implants is measured by means of the Osstell (R) device using a method of resonance frequency analysis. The aim of the study was to evaluate the role of direction-dependence of the Osstell (R) transducer. For this purpose, a set of parametrical finite element analyses has been used. When rotating the transducer around the vertical axis the crucial phenomenon is the change in behavior of the system, which occurs in positions between 30 degrees and 80 degrees to the long axis of the alveolar crest. It seems from the presented results that, when measuring approximately perpendicularly to this long axis, the deviation from the ideal perpendicular position must not exceed 30 degrees. In this case, the first resonance frequency is recorded. When measuring in the approximately parallel position to the long axis of the alveolar crest, the second resonance frequency is recorded. The deviation from the ideal parallel position must not exceed 10 degrees. These values are not comparable with the previous ones. (C) 2009 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:764 / 769
页数:6
相关论文
共 50 条
  • [41] Laser Resonance Frequency Analysis: A Novel Measurement Approach to Evaluate Acetabular Cup Stability During Surgery
    Kikuchi, Shunsuke
    Mikami, Katsuhiro
    Nakashima, Daisuke
    Kitamura, Toshiyuki
    Hasegawa, Noboru
    Nishikino, Masaharu
    Kanaji, Arihiko
    Nakamura, Masaya
    Nagura, Takeo
    SENSORS, 2019, 19 (22)
  • [42] The Correlation Between Crestal Bone Resorption and Implant Stability During Healing Period Using Resonance Frequency Analysis
    Al-Juboori, Mohammed Jasim
    Bin AbdulRahaman, Shafluzan
    Hassan, Akram
    IMPLANT DENTISTRY, 2013, 22 (04) : 351 - 355
  • [43] Measurement of Primary Stability of Mini Implants Using Resonance Frequency Analysis
    Singh, Ashish Kumar
    Kannan, Sridhar
    Arora, Nitin
    Bajaj, Yatin
    Revankar, Ameet V.
    APOS TRENDS IN ORTHODONTICS, 2018, 8 (03) : 139 - 145
  • [44] Can Finger-Generated Force Be Used Reliably to Connect the Transducer for Resonance Frequency Analysis in Determining Implant Stability?
    Pelegrine, Andre Antonio
    Miyauchi Kubo, Fabiola Mayumi
    Salatti, Debora Barella
    Teixeira, Marcelo Lucchesi
    Moshaverinia, Alireza
    Moy, Peter Karyen
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2020, 35 (06) : 1141 - 1148
  • [45] Influence of implant dimensions and position on primary and secondary stability: a prospective clinical study in the mandible using resonance frequency analysis
    Rengo, C.
    Nappo, A.
    Pantaleo, G.
    Spagnuolo, G.
    Ferrari, M.
    JOURNAL OF OSSEOINTEGRATION, 2020, 12 (01) : 34 - 38
  • [46] Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis
    Meredith, N
    Alleyne, D
    Cawley, P
    CLINICAL ORAL IMPLANTS RESEARCH, 1996, 7 (03) : 261 - 267
  • [47] Evaluation of dental implant stability in bone phantoms: Comparison between a quantitative ultrasound technique and resonance frequency analysis
    Vayron, Romain
    Vu-Hieu Nguyen
    Lecuelle, Benoit
    Haiat, Guillaume
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2018, 20 (04) : 470 - 478
  • [48] Relationship between dental implant stability determined by resonance frequency analysis measurements and peri-implant vertical defects:: an in vitro study
    Tozum, T. F.
    Turkyilmaz, I.
    McGlumphy, E. A.
    JOURNAL OF ORAL REHABILITATION, 2008, 35 (10) : 739 - 744
  • [49] The influence of fixation in formalin on the measurement of stability of implants using resonance frequency analysis and Periotest M®: A study in a dog
    Doi, Kazuya
    Kajihara, Shiho
    Morita, Koji
    Makihara, Yusuke
    Okada, Shinsuke
    Akagawa, Yasumasa
    BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 2014, 52 (01): : 29 - 33
  • [50] Resonance frequency analysis and damping capacity assessment -: Part I:: an in vitro study on measurement reliability and a method of comparison in the determination of primary dental implant stability
    Lachmann, S
    Jäger, B
    Axmann, D
    Gomez-Roman, G
    Groten, M
    Weber, H
    CLINICAL ORAL IMPLANTS RESEARCH, 2006, 17 (01) : 75 - 79