Influence of the orientation of the Osstell® transducer during measurement of dental implant stability using resonance frequency analysis: A numerical approach

被引:20
|
作者
Capek, Lukas [1 ]
Simunek, Antonin [2 ]
Slezak, Radovan [2 ]
Dzan, Ladislav [3 ]
机构
[1] Tech Univ Liberec, Dept Mech Engn, Liberec 46117 1, Czech Republic
[2] Charles Univ Prague, Fac Med Hradec Kralove, Dept Dent, Hradec Kralove 50001, Czech Republic
[3] Reg Hosp Liberec, Dept Maxillofacial Surg, Liberec 46001, Czech Republic
关键词
Resonance frequency analysis; Dental implants; Stability; Finite element analysis; Resonance frequency; Eigenfrequency; DAMPING CAPACITY ASSESSMENT; IN-VITRO; INTERFACE; BEHAVIOR; SURGERY; SITES; MODEL; TIBIA;
D O I
10.1016/j.medengphy.2009.02.003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Stability of dental implants is measured by means of the Osstell (R) device using a method of resonance frequency analysis. The aim of the study was to evaluate the role of direction-dependence of the Osstell (R) transducer. For this purpose, a set of parametrical finite element analyses has been used. When rotating the transducer around the vertical axis the crucial phenomenon is the change in behavior of the system, which occurs in positions between 30 degrees and 80 degrees to the long axis of the alveolar crest. It seems from the presented results that, when measuring approximately perpendicularly to this long axis, the deviation from the ideal perpendicular position must not exceed 30 degrees. In this case, the first resonance frequency is recorded. When measuring in the approximately parallel position to the long axis of the alveolar crest, the second resonance frequency is recorded. The deviation from the ideal parallel position must not exceed 10 degrees. These values are not comparable with the previous ones. (C) 2009 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:764 / 769
页数:6
相关论文
共 50 条
  • [21] A newly designed resonance frequency analysis device for dental implant stability detection
    Chang, Wei-Jen
    Lee, Sheng-Yang
    Wu, Chen-Che
    Lin, Che-Tong
    Abiko, Yoshimitsu
    Yamamichi, Nobuyuki
    Huang, Haw-Ming
    DENTAL MATERIALS JOURNAL, 2007, 26 (05) : 665 - 671
  • [22] An examination of immediately loaded dental implant stability in the diabetic patient using resonance frequency analysis (RFA)
    Balshi, Stephen F.
    Wolfinger, Glenn J.
    Balshi, Thomas J.
    QUINTESSENCE INTERNATIONAL, 2007, 38 (04): : 271 - 279
  • [23] Influence of Implant Dimensions and Position on Implant Stability: A Prospective Clinical Study in Maxilla Using Resonance Frequency Analysis
    Nappo, Antonio
    Rengo, Carlo
    Pantaleo, Giuseppe
    Spagnuolo, Gianrico
    Ferrari, Marco
    APPLIED SCIENCES-BASEL, 2019, 9 (05):
  • [24] A numerical approach to resonance frequency analysis for the investigation of oral implant osseointegration
    Natali, A. N.
    Pavan, P. G.
    Schileo, E.
    Williams, K. R.
    JOURNAL OF ORAL REHABILITATION, 2006, 33 (09) : 674 - 681
  • [25] Measurement of dental implant stability by resonance frequency analysis and damping capacity assessment: Comparison of both techniques in a clinical trial
    Zix, Juergen
    Hug, Stefan
    Kessler-Liechti, Gerda
    Mericske-Stern, Regina
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2008, 23 (03) : 525 - 530
  • [26] Clinical study on the primary stability of two dental implant systems with resonance frequency analysis
    Annette Rabel
    Steffen Gerhard Köhler
    Andrea Maria Schmidt-Westhausen
    Clinical Oral Investigations, 2007, 11 : 257 - 265
  • [27] Assessment of dental implant stability in osseodistraction-generated bone: a resonance frequency analysis
    Bilbao, Arturo
    Hernandez-De Oliveira, M.
    Varela-Centelles, Pablo I.
    Seoane, J.
    CLINICAL ORAL IMPLANTS RESEARCH, 2009, 20 (08) : 772 - 777
  • [28] Clinical study on the primary stability of two dental implant systems with resonance frequency analysis
    Rabel, Annette
    Koehler, Steffen Gerhard
    Schmidt-Westhausen, Andrea Maria
    CLINICAL ORAL INVESTIGATIONS, 2007, 11 (03) : 257 - 265
  • [29] Dental Implant Surface Topography and Stability With Resonance Frequency Analysis: An Overview and Case Report
    Dhalla, Nipun
    Palwankar, Pooja
    Gopal, Lipika
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (09)
  • [30] A Critique of Resonance Frequency Analysis and a Novel Method for Quantifying Dental Implant Stability in Vitro
    Khouja, Naseeba
    Tai, Wei Che
    Shen, I. Y.
    Sorensen, John A.
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2019, 34 (03) : 595 - 603