A Critique of Resonance Frequency Analysis and a Novel Method for Quantifying Dental Implant Stability in Vitro

被引:12
|
作者
Khouja, Naseeba
Tai, Wei Che
Shen, I. Y.
Sorensen, John A.
机构
[1] Biomimetics Biomaterials Biophotonics Biomechanics and Technology Laboratory, Department of Restorative Dentistry, School of Dentistry, University of Washington, 1959 NE Pacific St., Seattle, WA
[2] Department of Mechanical Engineering, University of Washington, Seattle, WA
关键词
angular stiffness; dental implant; experimental modal analysis; finite element analysis; natural frequency; stability; STRESS-DISTRIBUTION; POSTERIOR MAXILLA; ARTIFICIAL BONE; DENSITY; OSSEOINTEGRATION; FAILURE;
D O I
10.11607/jomi.7173
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: This study assessed the ability of resonance frequency measurements to differentiate the stability of implants with different lengths and diameters, and in different densities of bone. Another objective was to identify an alternative parameter capable of quantifying dental implant stability, thus facilitating greater sensitivity for efficacious detection of compromised or failing implants. Materials and Methods: Implants of two different diameters (4 and 5 mm) and six different lengths were individually placed in synthetic bone blocks of three different densities (15, 40/20, and 40 pounds per cubic foot) in combination with two different abutments (short and tall) to evaluate their stability. Resonance frequency measurements were obtained via Osstell ISQ and experimental modal analysis. The resonance frequency measurements were further confirmed via finite element analysis using commercially available software ANSYS. Results: Resonance frequencies measured via Osstell ISQ and experimental modal analysis did not change with respect to the length of the implants. The finite element analysis also confirmed the measured results. Finite element analysis simulations further indicated that angular stiffness at the neck of the implant (ie, the base of the abutment) varied considerably with respect to the implant length and diameter. Moreover, the calculated angular stiffness was independent of the type of abutment used. Conclusion: The results obtained from resonance frequency analyses did not accurately represent dental implant stability. Changes to implant length and diameter did not affect resonance frequencies. In contrast, angular stiffness at the neck of the implant represented a superior index for quantifying dental implant stability. It not only successfully differentiated stability of implants of both varying lengths and diameters, but also produced quantitative data that were independent of the type of abutments used.t
引用
收藏
页码:595 / 603
页数:9
相关论文
共 50 条
  • [1] Torsional resonance frequency analysis: a novel method for assessment of dental implant stability
    Tang, Yu-Long
    Li, Bing
    Jin, Wei
    Li, De-Hua
    CLINICAL ORAL IMPLANTS RESEARCH, 2015, 26 (06) : 615 - 622
  • [2] Design of transducers for resonance frequency measurement to assess the dental implant stability in vitro
    Sunil, Elthuri
    Chakraborty, Abhijit
    Ghosh, Ratna
    Goswami, Bhaswati
    IECON 2007: 33RD ANNUAL CONFERENCE OF THE IEEE INDUSTRIAL ELECTRONICS SOCIETY, VOLS 1-3, CONFERENCE PROCEEDINGS, 2007, : 2158 - +
  • [3] Resonance frequency analysis of dental implant stability during the healing period
    Boronat Lopez, Araceli
    Balaguer Martinez, Jose
    Lamas Pelayo, Joana
    Carrillo Garcia, Celia
    Penarrocha Diago, Miguel
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2008, 13 (04): : E244 - E247
  • [4] Measurement of dental implant stability by resonance frequency analysis: A review of the literature
    Pilar Quesada-Garcia, Ma
    Prados-Sanchez, Estrella
    Victoria Olmedo-Gaya, Ma
    Munoz-Soto, Esther
    Paloma Gonzalez-Rodriguez, Ma
    Vallecillo-Capilla, Manuel
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2009, 14 (10): : E538 - E546
  • [5] A simple mathematical model for the resonance frequency analysis of dental implant stability: Implant clamping quotient
    Argatov, Ivan
    Iantchenko, Alexei
    MECHANICS RESEARCH COMMUNICATIONS, 2019, 95 : 67 - 70
  • [6] A newly designed resonance frequency analysis device for dental implant stability detection
    Chang, Wei-Jen
    Lee, Sheng-Yang
    Wu, Chen-Che
    Lin, Che-Tong
    Abiko, Yoshimitsu
    Yamamichi, Nobuyuki
    Huang, Haw-Ming
    DENTAL MATERIALS JOURNAL, 2007, 26 (05) : 665 - 671
  • [7] Relationship between dental implant stability determined by resonance frequency analysis measurements and peri-implant vertical defects:: an in vitro study
    Tozum, T. F.
    Turkyilmaz, I.
    McGlumphy, E. A.
    JOURNAL OF ORAL REHABILITATION, 2008, 35 (10) : 739 - 744
  • [8] RESONANCE FREQUENCY DETECTION OF DENTAL IMPLANT STABILITY WITH VARIOUS BONE DEFECTS: IN-VITRO EXPERIMENTS
    Zhuang, H-B
    Chen, Ch-S
    Lee, S-Y
    Pan, M-Ch
    PROCEEDINGS OF THE ASME INTERNATIONAL DESIGN ENGINEERING TECHNICAL CONFERENCES AND COMPUTERS AND INFORMATION IN ENGINEERING CONFERENCE, VOL 1, PTS A AND B, 2010, : 765 - 772
  • [9] Is It Possible to Monitor Implant Stability on a Prosthetic Abutment? An In Vitro Resonance Frequency Analysis
    Lopez-Jarana, Paula
    Maria Diaz-Castro, Carmen
    Falcao, Artur
    Rios-Carrasco, Blanca
    Fernandez-Palacin, Ana
    Rios-Santos, Jose-Vicente
    Herrero-Climent, Mariano
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2020, 17 (11) : 1 - 14
  • [10] Clinical study on the primary stability of two dental implant systems with resonance frequency analysis
    Annette Rabel
    Steffen Gerhard Köhler
    Andrea Maria Schmidt-Westhausen
    Clinical Oral Investigations, 2007, 11 : 257 - 265