APD-based PET detector for simultaneous PET/MR imaging

被引:63
|
作者
Grazioso, Ronald [1 ]
Zhang, Nan
Corbeil, James
Schmand, Matthias
Ladebeck, Ralf
Vester, Markus
Schnur, Guenter
Renz, Wolfgang
Fischer, Hubertus
机构
[1] Siemens Mol Imaging, Knoxville, TN USA
[2] Seimens AG Med Solut MR, Erlangen, Germany
关键词
avalanche photodiode; positron emission tomography; magnetic resonance;
D O I
10.1016/j.nima.2006.08.121
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Two, APD-based, PET modules have been evaluated for use in combined PET/MR imaging. Each module consists of 4 independent, optically isolated detectors. Each detector consists of an 8 x 8 array of 2 x 2 x 20 mm LSO crystals read out by a 2 x 2 array of 5 x 5 mm Hamamatsu S8664-55 APDs. The average crystal energy resolution and time resolution (against a plastic scintillator on a PMT) of the detectors was 17% and 1.8 ns, respectively. The modules were positioned in the tunnel of a 1.5 T Siemens Symphony MR scanner. The presence of the PET modules decreased the M R signal-to-noise ratio by about 15% but no image interference was observed. The gradient and RE pulse sequences of the MR produced adverse effects on the PET event signals. These high-frequency pulses did not affect the true PET events but did increase the dead time of the PET system. Simultaneous, artifact-free, images were acquired with the PET and MR system using a small Derenzo phantom. These results show that APD-based PET detectors can be used for a high-resolution and cost-effective integrated PET/MR system. (c) 2006 Elsevier B.V. All rights reserved.
引用
下载
收藏
页码:301 / 305
页数:5
相关论文
共 50 条
  • [41] Design and Integration of a High Accuracy Multichannel Analog CMOS Peak Detect and Hold Circuit for APD-Based PET Imaging
    Fang, Xiaochao
    Brasse, David
    Hu-Guo, Christine
    Hu, Yann
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2012, 6 (02) : 179 - 187
  • [42] A study of artefacts in simultaneous PET and MR imaging using a prototype MR compatible PET scanner
    Slates, RB
    Farahani, K
    Shao, YP
    Marsden, PK
    Taylor, J
    Summers, PE
    Williams, S
    Beech, J
    Cherry, SR
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (08): : 2015 - 2027
  • [43] PET Performance Evaluation of a RF-Penetrable PET Insert for Simultaneous PET/MR Imaging
    Chang, Chen-Ming
    Grant, Alexander M.
    Lee, Brian J.
    Levin, Craig S.
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [44] Simultaneous PET/SPECT imaging with shared detector on an existing preclinical PET scanner
    Wu, Jing
    Chen, Si
    Liu, Yaqiang
    Wang, Shi
    Ma, Tianyu
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [45] Performance simulation of a compact PET insert for simultaneous PET/MR breast imaging
    Liang, Yicheng
    Peng, Hao
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2014, 751 : 23 - 30
  • [46] Simultaneous PET/MR imaging with a radio frequency- penetrable PET insert
    Grant, Alexander M.
    Lee, Brian J.
    Chang, Chen-Ming
    Levin, Craig S.
    MEDICAL PHYSICS, 2017, 44 (01) : 112 - 120
  • [47] Simultaneous PET/MR imaging results of a rat brain and a mouse heart with an MRI compatible PET detector in a 9.4 T microMRI
    Maramraju, Harsha
    Smith, S. David
    Stoll, Sean
    Schulz, Daniela
    Purschke, Martin
    Ravindranath, Bosky
    Junnarkar, Sachin
    Vaska, Paul
    Woody, Craig
    Schlyer, David
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [48] Design and performance of 0.18-μm CMOS charge preamplifiers for APD-based PET scanners
    Robert, S
    Pratte, JF
    DeGeronimo, G
    O'Connor, P
    Stoll, S
    Pepin, CM
    Fontaine, R
    Lecomte, R
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 1242 - 1246
  • [49] Design and performance of 0.18-μm CMOS charge preamplifiers for APD-based PET scanners
    Pratte, JF
    Robert, S
    De Geronimo, G
    O'Connor, P
    Stoll, S
    Pepin, CM
    Fontaine, R
    Lecomte, R
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (05) : 1979 - 1985
  • [50] Real time digital signal processing implementation for an APD-based PET scanner with phoswich detectors
    Fontaine, R.
    Tetrault, M. -A.
    Belanger, F.
    Viscogliosi, N.
    Himmich, R.
    Michaud, J. -B.
    Robert, S.
    Leroux, J. -D.
    Semmaoui, H.
    Berard, P.
    Cadorette, J.
    Pepin, C. M.
    Lecomte, R.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (03) : 784 - 788