APD-based PET detector for simultaneous PET/MR imaging

被引:63
|
作者
Grazioso, Ronald [1 ]
Zhang, Nan
Corbeil, James
Schmand, Matthias
Ladebeck, Ralf
Vester, Markus
Schnur, Guenter
Renz, Wolfgang
Fischer, Hubertus
机构
[1] Siemens Mol Imaging, Knoxville, TN USA
[2] Seimens AG Med Solut MR, Erlangen, Germany
关键词
avalanche photodiode; positron emission tomography; magnetic resonance;
D O I
10.1016/j.nima.2006.08.121
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Two, APD-based, PET modules have been evaluated for use in combined PET/MR imaging. Each module consists of 4 independent, optically isolated detectors. Each detector consists of an 8 x 8 array of 2 x 2 x 20 mm LSO crystals read out by a 2 x 2 array of 5 x 5 mm Hamamatsu S8664-55 APDs. The average crystal energy resolution and time resolution (against a plastic scintillator on a PMT) of the detectors was 17% and 1.8 ns, respectively. The modules were positioned in the tunnel of a 1.5 T Siemens Symphony MR scanner. The presence of the PET modules decreased the M R signal-to-noise ratio by about 15% but no image interference was observed. The gradient and RE pulse sequences of the MR produced adverse effects on the PET event signals. These high-frequency pulses did not affect the true PET events but did increase the dead time of the PET system. Simultaneous, artifact-free, images were acquired with the PET and MR system using a small Derenzo phantom. These results show that APD-based PET detectors can be used for a high-resolution and cost-effective integrated PET/MR system. (c) 2006 Elsevier B.V. All rights reserved.
引用
下载
收藏
页码:301 / 305
页数:5
相关论文
共 50 条
  • [21] Multiparametric imaging with simultaneous MR/PET
    Gatidis, S.
    Schmidt, H.
    Claussen, C. D.
    Schwenzer, N. F.
    RADIOLOGE, 2013, 53 (08): : 669 - 675
  • [22] Initial results of a truly pixelated APD-based PET scanner for high-resolution preclinical imaging
    Gaudin, Emilie
    Thibaudeau, Christian
    Arpin, Louis
    Leroux, Jean-Daniel
    Beaudoin, Jean-Franois
    Cadorette, Jules
    Njejimana, Larissa
    Samson, Arnaud
    Bouchard, Jonathan
    Koua, Konin
    Fontaine, jean
    Lecomte, Roger
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [23] Performance of APD-based monolithic-crystal detectors for small animal PET
    Maas, Marnix C.
    van der Laan, D. J.
    Schaart, Dennis R.
    Bruyndonckx, Peter
    Lemaitre, Cedric
    van Eijk, Carel W. E.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2017 - 2021
  • [24] SimPET: a Preclinical PET Insert for Simultaneous PET/MR Imaging
    Son, Jeong-Whan
    Kim, Kyeong Yun
    Park, Ji Yong
    Kim, Kyuwan
    Lee, Yun-Sang
    Ko, Guen Bae
    Lee, Jae Sung
    MOLECULAR IMAGING AND BIOLOGY, 2020, 22 (05) : 1208 - 1217
  • [25] Performance Characterization of a Dual-Threshold Time-over-Threshold APD-based Detector Front-End Module for PET Imaging
    Gaudin, Emilie
    Arpin, Louis
    Bouchard, Jonathan
    Paille, Maxime
    Bouziri, Haithem
    Bergeron, Melanie
    Pepin, Catherine M.
    Cadorette, Jules
    Fontaine, Rejean
    Lecomte, Roger
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [26] SimPET: a Preclinical PET Insert for Simultaneous PET/MR Imaging
    Jeong-Whan Son
    Kyeong Yun Kim
    Ji Yong Park
    Kyuwan Kim
    Yun-Sang Lee
    Guen Bae Ko
    Jae Sung Lee
    Molecular Imaging and Biology, 2020, 22 : 1208 - 1217
  • [27] Simultaneous PET/MR imaging with a MR-compatible SiPM PET scanner
    Yoon, Hyun Suk
    Ko, Guen Bae
    Kwon, Sun Il
    Lee, Chan Mi
    Ito, Mikiko
    Song, In Chan
    Lee, Dong Soo
    Hong, Seong Jong
    Lee, Jae Sung
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [28] Evaluation of dedicated PET/MR breast coil in integrated PET/MR system for simultaneous PET/MR breast imaging
    Cho, Ihn Ho
    Kong, Eun-jung
    Chun, KyunghAh
    Jeong, Ju Hye
    Kwon, SangDon
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (03)
  • [29] Development of an APD-Based PET Module and Preliminary Resolution Performance of an Experimental Prototype Gantry
    Kataoka, Jun
    Matsuda, Hidenori
    Nishikido, Fumihiko
    Koizumi, Makoto
    Ikeda, Hirokazu
    Yoshino, Masao
    Miura, Takamasa
    Tanaka, Satoshi
    Ishikawa, Yoshitaka
    Kawabata, Nobuyuki
    Shimizu, Keiji
    Matsunaga, Yusuke
    Kishimoto, Shunji
    Kubo, Hidetoshi
    Yanagida, Yoshio
    Nakamori, Takeshi
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (05) : 2448 - 2454
  • [30] Evaluation of a silicon photomultiplier PET insert for simultaneous PET and MR imaging
    Ko, Guen Bae
    Kim, Kyeong Yun
    Yoon, Hyun Suk
    Lee, Min Sun
    Son, Jeong-Whan
    Im, Hyung-Jun
    Lee, Jae Sung
    MEDICAL PHYSICS, 2016, 43 (01) : 72 - 83