Optimal electrocardiographically triggered phase for reducing motion artifact at electron-beam CT in the coronary artery

被引:19
|
作者
He, S
Dai, RP
Chen, Y
Bai, H
机构
[1] Chinese Acad Med Sci, Peking Union Med Coll, Cardiovasc Inst, Dept Radiol, Beijing 100037, Peoples R China
[2] Chinese Acad Med Sci, Fuwai Hosp, Peking Union Med Coll, Beijing 100037, Peoples R China
关键词
computed tomography (CT); artifact; electron beam; coronary vessels; CT;
D O I
10.1016/S1076-6332(03)80743-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. The authors performed this study to (a) investigate coronary movement with electron-beam computed tomography (CT) and (b) find the optimal electrocardiographic (ECG) triggering phase for eliminating motion artifact. Materials and Methods. One hundred fifty-one patients without arrhythmia were examined with electron-beam CT. First, movie scans were obtained to create displacement and velocity graphs of coronary artery movement. Then, a volume scan with an exposure time of 100 msec was obtained with various ECG trigger settings. Results. Movement patterns of coronary arteries varied with heart rate. Optimal triggering phase was before atrial systole (near 71% of the R-R interval) when heart rate was slower than 68 beats per minute and at ventricular end systole when heart rate was fast. Rate of severe motion artifacts decreased from 43% to 0% when triggering was altered from 80% of the R-R interval to the individual optimal value. Experimental values of the optimal phase at different heart rates were derived, and severe motion artifact was only 3.0% with these values. Conclusion, ECG triggering set according to the heart rate enables a great reduction in motion artifacts at electron-beam CT with a 100-msec exposure time. The results may have implications for magnetic resonance imaging of the coronary arte rv.
引用
收藏
页码:48 / 56
页数:9
相关论文
共 50 条
  • [1] Pericardial sinuses and recesses: Findings at electrocardiographically triggered electron-beam CT
    Groell, R
    Schaffler, GJ
    Rienmueller, R
    [J]. RADIOLOGY, 1999, 212 (01) : 69 - 73
  • [2] CORONARY-ARTERY STENOSES - 3-DIMENSIONAL IMAGING WITH ELECTROCARDIOGRAPHICALLY TRIGGERED, CONTRAST AGENT-ENHANCED, ELECTRON-BEAM CT
    MOSHAGE, WEL
    ACHENBACH, S
    SEESE, B
    BACHMANN, K
    KIRCHGEORG, M
    [J]. RADIOLOGY, 1995, 196 (03) : 707 - 714
  • [3] Coronary artery motion during the cardiac cycle and optimal ECG triggering for reducing motion artifact on coronary artery imaging
    Lu, B
    Mao, SS
    Zhuang, N
    Hamid, B
    Liu, SCK
    Budoff, MJ
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (02) : 447A - 447A
  • [4] Assessment of coronary artery stent patency by electron-beam CT
    Friedrich D. Knollmann
    Jörg Möller
    Axel Gebert
    Christian Bethge
    Roland Felix
    [J]. European Radiology, 2004, 14 : 1341 - 1347
  • [5] Assessment of coronary artery stent patency by electron-beam CT
    Knollmann, FD
    Möller, J
    Gebert, A
    Bethge, C
    Felix, R
    [J]. EUROPEAN RADIOLOGY, 2004, 14 (08) : 1341 - 1347
  • [6] Coronary artery anomalies with a shunt: Evaluation with electron-beam CT
    Yoshimura, N
    Hamada, S
    Takamiya, M
    Kuribayashi, S
    Kimura, K
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1998, 22 (05) : 682 - 686
  • [7] Detection and quantification of coronary artery calcification with electron-beam and conventional CT
    Becker, CR
    Knez, A
    Jakobs, TF
    Aydemir, S
    Becker, A
    Schoepf, UJ
    Bruening, R
    Haberl, R
    Reiser, MF
    [J]. EUROPEAN RADIOLOGY, 1999, 9 (04) : 620 - 624
  • [8] Spiral versus electron-beam CT for coronary artery calcium scoring
    Goldin, JG
    Yoon, HC
    Greaser, LE
    Heinze, SB
    McNitt-Gray, MM
    Brown, MS
    Sayre, JW
    Emerick, AM
    Aberle, DR
    [J]. RADIOLOGY, 2001, 221 (01) : 213 - 221
  • [9] Detection and quantification of coronary artery calcification with electron-beam and conventional CT
    C. R. Becker
    A. Knez
    T. F. Jakobs
    S. Aydemir
    A. Becker
    U. J. Schoepf
    R. Bruening
    R. Haberl
    M. F. Reiser
    [J]. European Radiology, 1999, 9 : 620 - 624
  • [10] Newer applications of electron-beam CT in imaging of coronary artery disease
    Stanford, W
    [J]. ACADEMIC RADIOLOGY, 1998, 5 (12) : 857 - 857