Accuracy of an infrared tympanic thermometer

被引:35
|
作者
Amoateng-Adjepong, Y [1 ]
Del Mundo, J [1 ]
Manthous, CA [1 ]
机构
[1] Yale Univ, Sch Med, Bridgeport Hosp, Dept Pulm & Crit Care, New Haven, CT USA
关键词
fever; infection; nursing; temperature; thermometer; thermometry;
D O I
10.1378/chest.115.4.1002
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: The use of infrared thermometry to measure temperatures in hospitalized patients is increasing. Although infrared thermometers have been proven to be accurate when they are used by well-trained personnel, no previous studies have examine their accuracy during routine hospital use, Objective: To determine the accuracy and observer variability of temperatures measured with an infrared tympanic thermometer (TT). Design: Prospective, observational study. Setting: ICUs of a 300-bed teaching community hospital. Patients: Fifty-one critically ill patients. Measurements: The mean of three tympanic temperatures measured with the infrared TT (tempTTs) was compared to temperatures simultaneously measured with the thermistor of right heart catheters and rectal mercury thermometers for the following three groups of observers who had been certified in the use of the infrared TT: a single critical care nurse (CCN)/educator (Ed); CCNs, and floor nurses (FNs)/clinical care practitioners (CCPs). Results: Two rounds of measurements were given to 51 patients by 153 observers. Temperatures of tbe pulmonary artery (PA) measured with the thermistor of light heart: catheters (tempPAs) ranged from 96.5 to 102.6 degrees F, with a mean ( +/- SD) of 99.3 +/- 1.1 degrees F. The intraobserver variabilities (correlation coefficients) of the tempTTs ranged from 0.90 for those measured by FNs/CCPs, to 0.92 for those measured by CCNs, to 0.98 for those measured by the CCN/Ed. Accuracy, arbitrarily defined as within a deviation of +/- 0.5 degrees F of the tempPA, was 100% for the rectal mercury thermometer and 98.0% for the infrared TT when used by the CCN/Ed. The accuracy of die infrared TT was 80% when measured by CCNs and 61% when measured by FNs/CCPs. Differences between tempPAs and tempTTs measured by the CCN/Ed ranged front 0 to 0.7 degrees F, with a mean of 0.2 degrees F. Similarly, differences between tempPAs and tempTTs measured by CCNs ranged from 0 to 2.4 degrees F, with a mean difference of 0.3 degrees F. However, differences between tempPAs and tempTTs measured by FNs/CCPs ranged from 0 to 3.0 degrees F, with a mean of 0.6 degrees F (greater differences than those obtained by the CCNs; p < 0.01). The accuracy of rectal mercury thermometry tvas 100%. If a temperature greater than or equal to 101.0 degrees F had been considered as the threshold at which a fever is present, and if the mean of three measurements had been used to designate temperature, workups that were either inappropriately performed or omitted would have resulted from 2% of tempTTs measured by the CCN/Ed, 1% of those measured by CCNs, and 4% of those measured by FNs/CCPs. Conclusion: When used properly, both tympanic and rectal thermometry are very accurate. However, the infrared TT produced measurements that were both less accurate and less reproducible when used by nurses who routinely used it in clinical practice.
引用
收藏
页码:1002 / 1005
页数:4
相关论文
共 50 条
  • [1] Accuracy of tympanic temperature measurement using an infrared tympanic membrane thermometer
    Gasim G.I.
    Musa I.R.
    Abdien M.T.
    Adam I.
    [J]. BMC Research Notes, 6 (1)
  • [2] TYMPANIC THERMOMETER ACCURACY
    NIERMAN, DM
    [J]. ANNALS OF EMERGENCY MEDICINE, 1992, 21 (10) : 1300 - 1300
  • [3] Clinical Accuracy of Tympanic Thermometer and Noncontact Infrared Skin Thermometer in Pediatric Practice An Alternative for Axillary Digital Thermometer
    Apa, Hursit
    Gozmen, Salih
    Bayram, Nuri
    Catikoglu, Asli
    Devrim, Fatma
    Karaarslan, Utku
    Gunay, Ilker
    Unal, Nurettin
    Devrim, Ilker
    [J]. PEDIATRIC EMERGENCY CARE, 2013, 29 (09) : 992 - 997
  • [4] INFRARED TYMPANIC THERMOMETER - EVALUATION OF A NEW CLINICAL THERMOMETER
    SHINOZAKI, T
    DEANE, R
    PERKINS, FM
    [J]. CRITICAL CARE MEDICINE, 1988, 16 (02) : 148 - 150
  • [5] Comparison of Infrared Tympanic Thermometer with Non-Contact Infrared Thermometer
    Bayhan, Cihangul
    Ozsurekci, Yasemin
    Tekcam, Necmiye
    Guloglu, Ayse
    Ehliz, Gulten
    Ceyhan, Mehmet
    Kara, Ates
    [J]. JOURNAL OF PEDIATRIC INFECTION, 2014, 8 (02): : 52 - 55
  • [6] Comparison of Infrared Tympanic Thermometer with Non-Contact Infrared Thermometer
    Ilce, Arzu
    [J]. JOURNAL OF PEDIATRIC INFECTION, 2014, 8 (04): : 200 - 200
  • [7] Clinical accuracy of non-contact forehead infrared thermometer and infrared tympanic thermometer in postoperative adult patients: A comparative study.
    Kameda, Norihiro
    [J]. JOURNAL OF PERIOPERATIVE PRACTICE, 2022, 32 (06) : 142 - 148
  • [8] Accuracy of parents in measuring body temperature with a tympanic thermometer
    Robinson J.L.
    Jou H.
    Spady D.W.
    [J]. BMC Family Practice, 6 (1)
  • [9] THE GENIUS INFRARED TYMPANIC THERMOMETER - AN EVALUATION FOR CLINICAL USE
    EDGE, G
    MORGAN, M
    [J]. ANAESTHESIA, 1993, 48 (07) : 604 - 607
  • [10] Intraoperative body temperature control: esophageal thermometer versus infrared tympanic thermometer
    Poveda, Vanessa de Brito
    Nascimento, Ariane de Souza
    [J]. REVISTA DA ESCOLA DE ENFERMAGEM DA USP, 2016, 50 (06) : 945 - 950