Reduced cement volume does not affect screw stability in augmented pedicle screws

被引:19
|
作者
Weiser, Lukas [1 ]
Sellenschloh, Kay [2 ]
Pueschel, Klaus [3 ]
Morlock, Michael M. [2 ]
Viezens, Lennart [1 ]
Lehmann, Wolfgang [1 ]
Huber, Gerd [2 ]
机构
[1] Univ Med Ctr Gottingen, Dept Trauma Surg Orthopaed Surg & Plast Surg, Robert Koch Str 40, D-37099 Gottingen, Germany
[2] TUHH Hamburg Univ Technol, Inst Biomech, Hamburg, Germany
[3] Univ Med Ctr Hamburg Eppendorf, Inst Forens Med, Hamburg, Germany
关键词
Pedicle screw; Cement augmentation; Cement volume; CAPSI; Bone mineral density; BONE-MINERAL DENSITY; OSTEOPOROTIC SPINE; POLYMETHYLMETHACRYLATE AUGMENTATION; BIOMECHANICAL CHARACTERISTICS; FIXATION; INSTRUMENTATION; STRENGTH; LEAKAGE; FRACTURES; DISEASES;
D O I
10.1007/s00586-020-06376-w
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Purpose Cement augmentation of pedicle screws is able to improve screw anchorage in osteoporotic vertebrae but is associated with a high complication rate. The goal of this study was to evaluate the impact of different cement volumes on pedicle screw fatigue strength. Methods Twenty-five human vertebral bodies (T12-L4) were collected from donors between 73 and 97 years of age. Bone density (BMD) was determined by quantitative computed tomography. Vertebral bodies were instrumented by conventional pedicle screws, and unilateral cement augmentation was performed. Thirteen vertebrae were augmented with a volume of 1 ml and twelve with a volume of 3 ml bone cement. A fatigue test was performed using a cranial-caudal sinusoidal, cyclic load (0.5 Hz) with increasing compression force (100 N + 0.1 N/cycles). Results The load to failure was 183.8 N for the non-augmented screws and was increased significantly to 268.1 N (p < 0.001) by cement augmentation. Augmentation with 1 ml bone cement increased the fatigue load by 41% while augmentation with 3 ml increased the failure load by 51% compared to the non-augmented screws, but there was no significant difference in fatigue loads between the specimens with screws augmented with 1 ml and screws augmented with 3 ml of bone cement (p = 0.504). Conclusion Cement augmentation significantly increases pedicle screw stability. The benefit of augmentation on screw anchorage was not significantly affected by reducing the applied volume of cement from 3 ml to 1 ml. Considering the high risk of cement leakage during augmentation, we recommend the usage of a reduced volume of 1 ml bone cement for each pedicle screw. Graphic Abstracts These slides can be retrieved under Electronic Supplementary Material . [GRAPHICS] .
引用
收藏
页码:1297 / 1303
页数:7
相关论文
共 50 条
  • [1] Reduced cement volume does not affect screw stability in augmented pedicle screws
    Lukas Weiser
    Kay Sellenschloh
    Klaus Püschel
    Michael M. Morlock
    Lennart Viezens
    Wolfgang Lehmann
    Gerd Huber
    European Spine Journal, 2020, 29 : 1297 - 1303
  • [2] Increased stability due to symmetric cement volume in augmented pedicle screws? A biomechanical study
    Weber, Maximilian
    Lenz, Maximilian
    Egenolf, Philipp
    Prescher, Andreas
    Walter, Sebastian
    Heck, Vincent J.
    Eysel, Peer
    Scheyerer, Max Joseph
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2023, 112 (01)
  • [3] Cement leakage in pedicle screw augmentation: a prospective analysis of 98 patients and 474 augmented pedicle screws
    Mueller, Jan U.
    Baldauf, Joerg
    Marx, Sascha
    Kirsch, Michael
    Schroeder, Henry W. S.
    Pillich, Dirk T.
    JOURNAL OF NEUROSURGERY-SPINE, 2016, 25 (01) : 103 - 109
  • [4] The cement leakage in cement-augmented pedicle screw instrumentation in degenerative lumbosacral diseases: a retrospective analysis of 202 cases and 950 augmented pedicle screws
    Guo, Hui-zhi
    Tang, Yong-chao
    Guo, Dan-qing
    Zhang, Shun-cong
    Li, Yong-xian
    Mo, Guo-ye
    Luo, Pei-jie
    Zhou, Ten-peng
    Ma, Yan-huai
    Liang, De
    Jiang, Xiao-bing
    EUROPEAN SPINE JOURNAL, 2019, 28 (07) : 1661 - 1669
  • [5] The cement leakage in cement-augmented pedicle screw instrumentation in degenerative lumbosacral diseases: a retrospective analysis of 202 cases and 950 augmented pedicle screws
    Hui-zhi Guo
    Yong-chao Tang
    Dan-qing Guo
    Shun-cong Zhang
    Yong-xian Li
    Guo-ye Mo
    Pei-jie Luo
    Ten-peng Zhou
    Yan-huai Ma
    De Liang
    Xiao-bing Jiang
    European Spine Journal, 2019, 28 : 1661 - 1669
  • [6] Screw inserting in different phase of cement affect the pull-out strength of cement augmented screws
    Iamthanaporn, Khanin
    Aonsong, Wachirakorn
    Tuntarattanapong, Pakjai
    Yuenyongviwat, Varah
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2022, 108 (08)
  • [7] Biomechanical study of expandable pedicle screw fixation in severe osteoporotic bone comparing with conventional and cement-augmented pedicle screws
    Chen, Yi-Long
    Chen, Wen-Chuan
    Chou, Chi-Wei
    Chen, Jou-Wen
    Chang, Chia-Ming
    Lai, Yu-Shu
    Cheng, Cheng-Kung
    Wang, Shih-Tien
    MEDICAL ENGINEERING & PHYSICS, 2014, 36 (11) : 1416 - 1420
  • [8] The Potential Impact of Basivertebral Foramen Morphology and Pedicle Screw Placement on Epidural Cement Leakage With Cement-Augmented Fenestrated Pedicle Screw Fixation: A Multicenter Retrospective Study of 282 Patients and 1404 Augmented Screws
    Yu, Weibo
    Liang, De
    Yao, Zhensong
    Zhang, Haiyan
    Zhong, Yuanming
    Tang, Yongchao
    Jiang, Xiaobing
    Cai, Daozhang
    NEUROSURGERY, 2023, 93 (01) : 66 - 74
  • [9] Cement Augmented Pedicle Screw Instrumentation in Pediatric Spine Surgery
    Mraja, Hamisi Mwarindano
    Gok, Halil
    Daadour, Inas Mohamed Fawzy
    Ulusoy, Onur Levent
    Sanli, Tunay
    Karadereler, Selhan
    Enercan, Meric
    Hamzaoglu, Azmi
    WORLD NEUROSURGERY, 2023, 174 : E126 - E130
  • [10] Primary stability of pedicle screws depends on the screw positioning and alignment
    Costa, Francesco
    Villa, Tomaso
    Anasetti, Federica
    Tomei, Massimo
    Ortolina, Alessandro
    Cardia, Andrea
    La Barbera, Luigi
    Fornari, Maurizio
    Galbusera, Fabio
    SPINE JOURNAL, 2013, 13 (12): : 1934 - 1939