10 Volt Automated Direct Comparison of Two Cryocooled Programmable Josephson Voltage Standards

被引:0
|
作者
Rufenacht, Alain [1 ]
Tang, Yi-hua [2 ]
Fox, Anna E. [1 ]
Dresselhaus, Paul D. [1 ]
Burroughs, Charles J. [1 ]
Schwall, Robert E. [1 ]
Benz, Samuel P. [1 ]
机构
[1] NIST, Boulder, CO 80305 USA
[2] NIST, Gaithersburg, MD 20899 USA
关键词
Josephson arrays; Precision measurements; Uncertainty; Voltage measurement;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We have performed direct dc comparisons between two cryocooled 10 V programmable Josephson voltage standards utilizing an automated synchronization scheme for the voltage reversals, which enables the use of a high sensitivity null detector on its 3 mu V range by preventing any overload condition. No switches or manual operations are necessary to protect the null detector. Comparing the two systems under various test conditions provides robust verification of ideal system performance, and enables verification of the key components of the uncertainty budget for both the measurement methods and system operations.
引用
收藏
页数:2
相关论文
共 50 条
  • [21] Direct Comparison between the NIST 10 V Conventional Josephson Voltage Standard and 2.5 V Programmable Josephson Voltage Standard
    Jaiswal, Shiv Kumar
    MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2011, 26 (04): : 339 - 348
  • [22] Direct comparison between the NIST 10 V Compact Josephson Voltage Standard and the 2.5 V Programmable Josephson Voltage Standard
    Jaiswal, Shiv Kumar
    Tang, Y.
    MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA, 2009, 24 (04): : 207 - 214
  • [23] 10 VSINIS Josephson junction series arrays for programmable voltage standards
    Kohlmann, J
    Schulze, H
    Behr, R
    Müller, F
    Niemeyer, J
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2001, 50 (02) : 192 - 194
  • [24] Direct comparison between the NIST 10 V Compact Josephson Voltage Standard and the 2.5 V Programmable Josephson Voltage Standard
    Shiv Kumar Jaiswal
    Y. Tang
    MAPAN, 2009, 24 : 207 - 214
  • [25] Direct comparison between the NIST 10 V Conventional Josephson Voltage Standard and 2.5 V Programmable Josephson Voltage Standard
    Shiv Kumar Jaiswal
    MAPAN, 2011, 26 : 339 - 348
  • [26] Direct Comparison of DC Josephson Voltage Standard and Programmable Josephson Voltage Standard of TUBITAK UME
    Turhan, Saliha
    Kanatoglu, Naylan
    Orhan, Recep
    2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [27] Direct comparison between a programmable and a conventional Josephson voltage standard at the level of 10V
    Djordjevic, S.
    Seron, O.
    Solve, S.
    Chayramy, R.
    METROLOGIA, 2008, 45 (04) : 429 - 435
  • [28] 10 Volt Programmable Josephson Voltage Standard Circuits Using NbSi-Barrier Junctions
    Dresselhaus, Paul D.
    Elsbury, Michael M.
    Olaya, David
    Burroughs, Charles J.
    Benz, Samuel P.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2011, 21 (03) : 693 - 696
  • [29] A Josephson Impedance Bridge Based on Programmable Josephson Voltage Standards
    Hagen, Thomas
    Palafox, Luis
    Behr, Ralf
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2017, 66 (06) : 1539 - 1545
  • [30] Cryogen-Free Operation of 10 V Programmable Josephson Voltage Standards
    Howe, L.
    Burroughs, C. J.
    Dresselhaus, P. D.
    Benz, S. P.
    Schwall, R. E.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2013, 23 (03)