Comparison of the bovine blood gas parameters produced with three types of portable blood gas analyzers

被引:3
|
作者
Ro, Younghye [1 ,2 ]
Choi, Woojae [1 ]
Hong, Leegon [1 ]
Kim, Eunkyung [2 ]
Choe, Eunhui [2 ]
Kim, Danil [1 ,2 ,3 ]
机构
[1] Seoul Natl Univ, Coll Vet Med, Dept Farm Anim Med, 1 Gwanak Ro, Seoul 08826, South Korea
[2] Seoul Natl Univ, Farm Anim Clin Training & Res Ctr, Inst Green Bio Sci, Pyeongchang 25354, South Korea
[3] Seoul Natl Univ, Res Inst Vet Sci, Seoul 08826, South Korea
关键词
Blood gas analysis; calcium; cattle; pH; point-of-care systems; IONIZED CALCIUM-CONCENTRATION; SUBCLINICAL HYPOCALCEMIA; VALIDATION; METABOLISM; PH;
D O I
10.4142/jvs.22050
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background: A definite diagnosis should be made in the bovine practice field, however, it was difficult to perform laboratory analysis immediately. Currently, three types of portable blood gas analyzers are available in Korea. Objectives: This study aimed to evaluate the correlations among these three analyzers. Methods: Seventy-two plasma samples from Holstein-Friesian cows were used for blood gas analysis, and three instruments (EDAN i15 Vet, VETSCAN i-STAT, and EPOC) were operated simultaneously. Moreover, plasma calcium levels were compared between these portable analyzers and blood chemistry device, which is usually used in a laboratory environment. Pearson analysis was performed to confirm the correlation of each parameter produced with the three instruments and blood chemistry analyzer. Results: As results, high correlation was observed in parameters of pH, pO(2), potassium ion, ionized calcium, and glucose (p < 0.001, r > 0.7). In addition, pCO(2) showed a moderate correlation among the three analyzers (p < 0.001, r > 0.5), and there was no correlation among all instruments for sodium ions. There was also a high correlation between ionized calcium from the three portable devices and total calcium from the biochemistry analyzer (p < 0.001, r > 0.9). Conclusions: In conclusion, there was a high correlation between results from the three different blood gas analyzers used in the bovine clinical field in Korea. Thus, a consistent diagnosis can be made even with different equipment if the operator is aware of the strengths and weaknesses of each piece of equipment and operates it properly.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Blood Gas Analyzers
    Gonzalez, Anthony L.
    Waddell, Lori S.
    [J]. TOPICS IN COMPANION ANIMAL MEDICINE, 2016, 31 (01) : 27 - 34
  • [3] Interchangeability of parameters on POCT blood gas analyzers and core laboratory analyzers
    Vanoverschelde, L.
    Dierickx, E.
    De Paepe, P.
    Decruyenaere, J.
    Stove, V.
    [J]. ACTA CLINICA BELGICA, 2015, 70 : S14 - S15
  • [4] A COMPARISON OF 4 DIFFERENT BLOOD-GAS ANALYZERS
    KOFSTAD, J
    [J]. SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1981, 41 (04): : 409 - 414
  • [5] AUTOMATIC BLOOD-GAS ANALYZERS
    TWEEDDALE, PM
    [J]. BRITISH MEDICAL JOURNAL, 1979, 1 (6162): : 547 - 548
  • [6] AUTOMATED BLOOD-GAS ANALYZERS
    BLATT, ML
    [J]. JOURNAL OF CLINICAL LABORATORY AUTOMATION, 1983, 3 (06): : 416 - 419
  • [7] Product Showcase: Blood Gas Analyzers
    Hunt, Jeanely
    [J]. AMERICAN LABORATORY, 2013, 45 (03) : 42 - 43
  • [8] Three-year monitoring and comparison of results from two identical blood gas analyzers
    Huang, Yun
    Dean, Robert
    Dubbelman, Yvonne
    Vincent, Anne
    Chan, Ying-pui Michael
    [J]. PRACTICAL LABORATORY MEDICINE, 2022, 31
  • [9] Three-year monitoring and comparison of results from two identical blood gas analyzers
    Huang, Yun
    Dean, Robert
    Dubbelman, Yvonne
    Vincent, Anne
    Chan, Ying-pui Michael
    [J]. PRACTICAL LABORATORY MEDICINE, 2022, 31
  • [10] Comparison of two portable clinical analyzers to one stationary analyzer for the determination of blood gas partial pressures and blood electrolyte concentrations in horses
    Kirsch, Katharina
    Detilleux, Johann
    Serteyn, Didier
    Sandersen, Charlotte
    [J]. PLOS ONE, 2019, 14 (02):