Comparison of the bovine blood gas parameters produced with three types of portable blood gas analyzers

被引:3
|
作者
Ro, Younghye [1 ,2 ]
Choi, Woojae [1 ]
Hong, Leegon [1 ]
Kim, Eunkyung [2 ]
Choe, Eunhui [2 ]
Kim, Danil [1 ,2 ,3 ]
机构
[1] Seoul Natl Univ, Coll Vet Med, Dept Farm Anim Med, 1 Gwanak Ro, Seoul 08826, South Korea
[2] Seoul Natl Univ, Farm Anim Clin Training & Res Ctr, Inst Green Bio Sci, Pyeongchang 25354, South Korea
[3] Seoul Natl Univ, Res Inst Vet Sci, Seoul 08826, South Korea
关键词
Blood gas analysis; calcium; cattle; pH; point-of-care systems; IONIZED CALCIUM-CONCENTRATION; SUBCLINICAL HYPOCALCEMIA; VALIDATION; METABOLISM; PH;
D O I
10.4142/jvs.22050
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background: A definite diagnosis should be made in the bovine practice field, however, it was difficult to perform laboratory analysis immediately. Currently, three types of portable blood gas analyzers are available in Korea. Objectives: This study aimed to evaluate the correlations among these three analyzers. Methods: Seventy-two plasma samples from Holstein-Friesian cows were used for blood gas analysis, and three instruments (EDAN i15 Vet, VETSCAN i-STAT, and EPOC) were operated simultaneously. Moreover, plasma calcium levels were compared between these portable analyzers and blood chemistry device, which is usually used in a laboratory environment. Pearson analysis was performed to confirm the correlation of each parameter produced with the three instruments and blood chemistry analyzer. Results: As results, high correlation was observed in parameters of pH, pO(2), potassium ion, ionized calcium, and glucose (p < 0.001, r > 0.7). In addition, pCO(2) showed a moderate correlation among the three analyzers (p < 0.001, r > 0.5), and there was no correlation among all instruments for sodium ions. There was also a high correlation between ionized calcium from the three portable devices and total calcium from the biochemistry analyzer (p < 0.001, r > 0.9). Conclusions: In conclusion, there was a high correlation between results from the three different blood gas analyzers used in the bovine clinical field in Korea. Thus, a consistent diagnosis can be made even with different equipment if the operator is aware of the strengths and weaknesses of each piece of equipment and operates it properly.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] Comparison of bicarbonate and base excess values analyzed by four different blood gas analyzers
    Nakamaru, Katsuhito
    Hatakeyama, Noboru
    Yamada, Masana
    Yamazaki, Mitsuaki
    [J]. JOURNAL OF ANESTHESIA, 2007, 21 (03) : 429 - 432
  • [22] Comparison of bicarbonate and base excess values analyzed by four different blood gas analyzers
    Katsuhito Nakamaru
    Noboru Hatakeyama
    Masana Yamada
    Mitsuaki Yamazaki
    [J]. Journal of Anesthesia, 2007, 21
  • [23] COMPARISON OF BLOOD ACID BASE GAS PARAMETERS IN VENOUS AND ARTERIAL BLOOD OF HEALTHY BUFFALOES
    Hussain, S. A.
    Uppal, S. K.
    [J]. BUFFALO BULLETIN, 2014, 33 (03): : 328 - 331
  • [24] Comparison of cord blood lactate measurement by gas analyzer and portable electrochemical devices
    Romanov, Andrey Yurievich
    Prikhodko, Andrey Mikhailovich
    Tysyachnyi, Oleg Vladimirovich
    Baev, Oleg Radomirovich
    Yarotskaya, Ekaterina Lvovna
    Sukhikh, Gennady Tikhonovich
    [J]. JOURNAL OF PERINATAL MEDICINE, 2020, 48 (02) : 157 - 161
  • [25] A PORTABLE BLOOD-GAS ANALYZER FOR EQUINE VENOUS-BLOOD
    MITTEN, LA
    HINCHCLIFF, KW
    SAMS, R
    [J]. JOURNAL OF VETERINARY INTERNAL MEDICINE, 1995, 9 (05) : 353 - 356
  • [26] ARTERIAL BLOOD COLLECTION FOR GAS AND OTHER ANALYSES. COMPARISON OF RESULTS OBTAINED WITH THREE TYPES OF SYRINGES
    Daures, M.-F.
    Vallat, C.
    Combescure, C.
    Cristol, J. P.
    [J]. BIOTECHNOLOGY & BIOTECHNOLOGICAL EQUIPMENT, 2009, 23 (02) : 1259 - 1265
  • [27] Accuracy of a portable blood gas analyzer incorporating optodes for canine blood
    Shiroshita, Y
    Tanaka, R
    Shibazaki, A
    Yamane, Y
    [J]. JOURNAL OF VETERINARY INTERNAL MEDICINE, 1999, 13 (06) : 597 - 600
  • [28] Electrolyte measurement in goats: Comparison of 2 blood gas analyzers and evaluation of the preanalytical blood sample preparation on measurement results
    Luethy, Daniela
    Hopster, Klaus
    [J]. TIERAERZTLICHE PRAXIS AUSGABE GROSSTIERE NUTZTIERE, 2020, 48 (03): : 148 - 155
  • [29] INTERINSTRUMENT COMPARISON OF BLOOD-GAS ANALYZERS AND ASSESSMENT OF TONOMETRY USING FRESH HEPARINIZED WHOLE HUMAN-BLOOD
    VANKESSEL, AL
    EICHHORN, JH
    CLAUSEN, JL
    STONE, ME
    ROTMAN, HH
    CRAPO, RO
    [J]. CHEST, 1987, 92 (03) : 418 - 422
  • [30] Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers
    Leino, Aila
    Kurvinen, Kaisa
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 (07) : 1187 - 1191