Comparison of two portable clinical analyzers to one stationary analyzer for the determination of blood gas partial pressures and blood electrolyte concentrations in horses

被引:14
|
作者
Kirsch, Katharina [1 ]
Detilleux, Johann [2 ]
Serteyn, Didier [1 ]
Sandersen, Charlotte [1 ]
机构
[1] Univ Liege, Fac Vet Med, Dept Compan Anim & Equids, Equine Clin, Liege, Belgium
[2] Univ Liege, Fac Vet Med, Dept Quantitat Genet, Liege, Belgium
来源
PLOS ONE | 2019年 / 14卷 / 02期
关键词
ACID-BASE; STATISTICAL-METHODS; HEMOGLOBIN; AGREEMENT; INTERFERENCE; POTASSIUM; PRECISION;
D O I
10.1371/journal.pone.0211104
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Portable blood gas analyzers are used to facilitate diagnosis and treatment of disorders related to disturbances of acid-base and electrolyte balance in the ambulatory care of equine patients. The aim of this study was to determine whether 2 portable analyzers produce results in agreement with a stationary analyzer. Blood samples from 23 horses hospitalized for various medical reasons were included in this prospective study. Blood gas analysis and electrolyte concentrations measured by the portable analyzers VetStat and epoc were compared to those produced by the cobas b 123 analyzer via concordance analysis, Passing-Bablok regression and Bland-Altman analysis. Limits of agreement indicated relevant bias between the VetStat and cobas b 123 for partial pressure of oxygen (pO(2); 27.5-33.8 mmHg), sodium ([Na+]; 4.3-21.6 mmol/L) and chloride concentration ([CI-]; 0.3-7.9 mmol/L) and between the epoc and cobas b 123 for pH (0.070-0.022), partial pressure of carbon dioxide (pCO(2); 3.6-7.3 mmHg), pO(2) (36.2-32.7 mmHg) and [Na+] (0.38.1 mmol/L). The VetStat analyzer yielded results that were in agreement with the cobas b 123 analyzer for determination of pH, pCO(2), bicarbonate ([HCO3-]) and potassium concentration [K+], while the epoc analyzer achieved acceptable agreement for [HCO3] and [K+]. The VetStat analyzer may be useful in performing blood gas analysis in equine samples but analysis of [Na+], [Cl-] and pO(2) should be interpreted with caution. The epoc delivered reliable results for [HCO3] and [K+], while results for pH, pCO(2), pO(2) and [Na+] should be interpreted with caution.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Evaluation of a portable clinical analyzer for the determination of blood gas partial pressures, electrolyte concentrations, and hematocrit in venous blood samples collected from cattle, horses, and sheep
    Peiro, Juliana R.
    Borges, Alexandre S.
    Goncalves, Roberto C.
    Mendes, Luiz Claudio N.
    [J]. AMERICAN JOURNAL OF VETERINARY RESEARCH, 2010, 71 (05) : 515 - 521
  • [2] Use of a handheld device for analysis of blood electrolyte concentrations and blood gas partial pressures in dogs and horses
    Looney, AL
    Ludders, J
    Erb, HN
    Gleed, R
    Moon, P
    [J]. JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 1998, 213 (04) : 526 - 530
  • [3] Comparison of blood gas, electrolyte and metabolite results measured with two different blood gas analyzers and a core laboratory analyzer
    Uyanik, Metin
    Sertoglu, Erdim
    Kayadibi, Huseyin
    Tapan, Serkan
    Serdar, Muhittin A.
    Bilgi, Cumhur
    Kurt, Ismail
    [J]. SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2015, 75 (02): : 97 - 105
  • [4] Portable blood gas and electrolyte analyzer evaluated in a multiinstitutional study
    Lindemans, J
    Hoefkens, P
    van Kessel, AL
    Bonnay, M
    Külpmann, WR
    van Suijlen, JDE
    [J]. CLINICAL CHEMISTRY, 1999, 45 (01) : 111 - 117
  • [5] Evaluation of two benchtop blood gas analyzers for measurement of electrolyte concentrations in venous blood samples from dogs
    Schmitz, Katrina L.
    Jeffery, Unity
    Heinz, Justin A.
    Rutter, Christine R.
    [J]. AMERICAN JOURNAL OF VETERINARY RESEARCH, 2021, 82 (02) : 105 - 109
  • [6] Comparison of the bovine blood gas parameters produced with three types of portable blood gas analyzers
    Ro, Younghye
    Choi, Woojae
    Hong, Leegon
    Kim, Eunkyung
    Choe, Eunhui
    Kim, Danil
    [J]. JOURNAL OF VETERINARY SCIENCE, 2022, 23 (04)
  • [7] Use of the portable blood gas analyzer OPTI® CCA in measurement of arterial blood samples in horses
    Roscher, K.
    Failing, K.
    Moritz, A.
    [J]. TIERAERZTLICHE PRAXIS AUSGABE GROSSTIERE NUTZTIERE, 2007, 35 (02): : 126 - 131
  • [8] Laboratory evaluation of the SenDx100 portable blood gas and electrolyte analyzer.
    Nichols, JH
    Engelstad, ME
    Li, D
    Silberman, M
    Sokoll, LJ
    [J]. CLINICAL CHEMISTRY, 1997, 43 : 270 - 270
  • [9] Determination of lactate concentrations in blood plasma and peritoneal fluid in horses with colic by an Accusport analyzer
    Delesalle, Catherine
    Dewulf, Jeroen
    Lefebvre, Romain A.
    Schuurkes, Jan A. J.
    Proot, Joachim
    Lefere, Laurence
    Deprez, Piet
    [J]. JOURNAL OF VETERINARY INTERNAL MEDICINE, 2007, 21 (02): : 293 - 301
  • [10] Comparison of a human portable blood glucose meter, veterinary portable blood glucose meter, and automated chemistry analyzer for measurement of blood glucose concentrations in dogs
    Johnson, Beth M.
    Fry, Michael M.
    Flatland, Bente
    Kirk, Claudia A.
    [J]. JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2009, 235 (11): : 1309 - 1313