Investigation of imaging performance of amorphous selenium flat-panel detectors for digital mammography

被引:13
|
作者
Zhao, W [1 ]
Ji, WG [1 ]
Rowlands, JA [1 ]
Debrie, A [1 ]
机构
[1] SUNY Stony Brook, Hlth Sci Ctr L 4, Dept Radiol, Stony Brook, NY 11794 USA
关键词
flat-panel detectors; digital mammography; MTF; NPS; DQE; selenium;
D O I
10.1117/12.430876
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Our work is to investigate and understand the factors affecting the imaging performance of amorphous selenium (a-Se) flat-panel detectors for digital mammography. Both theoretical and experimental methods were developed to investigate the spatial frequency dependent detective quantum efficiency [DQE(f)] of a-Se flat-panel detectors for digital mammography. Since the k-edge of a-Se is 12.66 keV and within the energy range of a mammographic spectrum, a cascaded linear system model was developed which takes into account the effect of k-fluorescence on the modulation transfer function (MTF), noise power spectrum (NPS) and DQE(f) of the detector. This model was used to understand the performance of a prototype detector with 85 mum. pixel size. The presampling MTF, NPS and DQE(f) of the prototype were measured, and compared to the theoretical calculation by the model. The calculation showed that k-fluorescence reduces the MTF by 15% at the Nyquist frequency (f(NY)) of the prototype detector, and the NPS at f(NY) was reduced to 82% of that at zero spatial frequency. The measurement of presampling MTF of the prototype detector revealed an additional source of blurring, which was attributed to the blocking layer at the interface between a-Se and the active matrix. This introduced high frequency drop in both presampling MTF and NPS, and reduced aliasing in the NPS. As a result, the DQE(f) of the prototype detector at f(NY) approaches 50% of that at zero spatial frequency.
引用
下载
收藏
页码:536 / 546
页数:11
相关论文
共 50 条
  • [41] Pneumothorax detection: Comparison of flat-panel amorphous silicon and computed radiography detectors
    Kampschulte, A
    Hansmann, J
    Hahmann, M
    Leutloff, UC
    Fink, C
    Kauffmann, GW
    RADIOLOGY, 2000, 217 : 124 - 124
  • [42] Selenium-based amorphous silicon flat-panel digital X-ray imager for protein crystallography
    Sultana, Afrin
    Karim, K. S.
    Rowlands, J. A.
    CANADIAN JOURNAL OF ELECTRICAL AND COMPUTER ENGINEERING-REVUE CANADIENNE DE GENIE ELECTRIQUE ET INFORMATIQUE, 2008, 33 (3-4): : 139 - 143
  • [43] An evaluation of noise in radiotracer emission imaging using flat-panel detectors
    Bowsher, J.
    Yin, F.
    Chawla, A.
    Greer, K.
    Song, H.
    Samei, E.
    Willett, C.
    MEDICAL PHYSICS, 2007, 34 (06) : 2636 - 2636
  • [44] Investigation of the effect of anode/filter materials on the dose and image quality of a digital mammography system based on an amorphous selenium flat panel detector
    Baldelli, P.
    Phelan, N.
    Egan, G.
    BRITISH JOURNAL OF RADIOLOGY, 2010, 83 (988): : 290 - 295
  • [45] Medical imaging with mercuric iodide direct digital radiography flat-panel X-ray detectors
    Schieber, M
    Hermon, H
    Zentai, G
    Partain, L
    Street, RA
    Gilboa, H
    RADIOLOGY, 2001, 221 : 464 - 464
  • [46] Medical imaging with mercuric iodide direct digital radiography flat-panel X-ray detectors
    Gilboa, H
    Zuck, A
    Dagan, O
    Vilensky, A
    Breen, BN
    Taieb, A
    Reisman, B
    Hermon, H
    Zentai, G
    Partain, L
    Street, R
    Ready, S
    X-RAY AND GAMMA-RAY DETECTORS AND APPLICATIONS IV, 2002, 4784 : 315 - 325
  • [47] Investigation of optimal kVp settings for CT mammography using a flat-panel imager
    Glick, SJ
    Vedantham, S
    Karellas, A
    MEDICAL IMAGING 2002: PHYSICS OF MEDICAL IMAGING, 2002, 4682 : 392 - 402
  • [48] A theoretical investigation of advanced design active matrix flat-panel imagers for mammography
    Jee, KW
    Antonuk, LE
    El-Mohri, Y
    Cionca, C
    Maolinbay, M
    Nassif, S
    Zhao, QH
    MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING, 2000, 3977 : 250 - 256
  • [49] Digital Flat Panel Detectors in Medical Imaging Systems
    Avakyan, A. K.
    Zavestovskaya, I. N.
    Lobzhanidze, T. K.
    Polikhov, S. A.
    Smirnov, V. P.
    BULLETIN OF THE LEBEDEV PHYSICS INSTITUTE, 2018, 45 (11) : 356 - 359
  • [50] Digital Flat Panel Detectors in Medical Imaging Systems
    A. K. Avakyan
    I. N. Zavestovskaya
    T. K. Lobzhanidze
    S. A. Polikhov
    V. P. Smirnov
    Bulletin of the Lebedev Physics Institute, 2018, 45 : 356 - 359