Development land reliability of a measurement device for flexion force of the first metatarsophalangeal joint

被引:2
|
作者
Swanenburg, J
Stappaerts, KH
Tirez, B
Uebelhart, D
Aufdemkampe, G
机构
[1] Univ Zurich Hosp, Dept Rheumatol, CH-8091 Zurich, Switzerland
[2] Univ Zurich Hosp, Inst Med Phys, CH-8091 Zurich, Switzerland
[3] Catholic Univ Louvain, Dept Rehabil Sci, B-3000 Louvain, Belgium
[4] Univ Leuven, Dept Phys, Louvain, Belgium
[5] Univ Hosp Leuven, Dept Phys Med & Rehabil, Louvain, Belgium
[6] Polytech Utrecht, Dept Phys Therapy, Utrecht, Netherlands
关键词
hallux; aged; women; strength;
D O I
10.1123/jab.19.3.271
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to present a method for repeated measurement of flexion force of the hallux in the metatarsophalangeal joint. The reliability of this measurement device was also examined. This device is suitable for situations where weight-bearing is contraindicated or when it is not possible for patients to bear load on their toes, such as hallux valgus patients. Since most such patients are female, the participants in this study were 24 healthy female volunteers. Age, weight, height, and leg dominance were determined for each. Muscle strength was measured using a device with a built-in MicroFET dynamometer. The result for the left hallux was ICC(3,1).89 (95 % CI.77-95). The result for the right hallux was ICC(3,1).94 (95% CI.87-97). In the Bland and Altman plots, the reliability again appeared to be sufficient. The Pearson product-moment correlations gave poor results for the association between body weight, height, age, and mean force of the four trails. The test results indicate good reliability of the measurement device as used in this study. The advantage of this testing device is that it makes it easier to standardize measurements as opposed to the MicroFET used as a hand-held dynamometer. Also, patients can be tested in a nonload situation, which makes it possible to test hallux valgus at any time, and therefore it is possible to monitor variations in progression (or regression).
引用
收藏
页码:271 / 278
页数:8
相关论文
共 50 条
  • [31] Development of a diagnostic rule for identifying radiographic osteoarthritis in people with first metatarsophalangeal joint pain
    Zammit, G. V.
    Munteanu, S. E.
    Menz, H. B.
    OSTEOARTHRITIS AND CARTILAGE, 2011, 19 (08) : 939 - 945
  • [32] A Development of Ankle Joint Range of Motion Measurement System Using Electromagnetic Tracking Device and Force Sensor
    Sakashita, Ryo
    Nagamune, Kouki
    Kuroda, Ryosuke
    2018 WORLD AUTOMATION CONGRESS (WAC), 2018, : 10 - 13
  • [33] Validity and Intra Rater Reliability of a New Device for Tongue Force Measurement
    Diaz-Saez, Marta Carlota
    Beltran-Alacreu, Hector
    Gil-Castillo, Javier
    Navarro-Fernandez, Gonzalo
    Cebrian Carretero, Jose Luis
    Gil-Martinez, Alfonso
    INTERNATIONAL JOURNAL OF INTERACTIVE MULTIMEDIA AND ARTIFICIAL INTELLIGENCE, 2023, 8 (02): : 60 - 68
  • [34] Development and Validation of a Wearable Plantar Force Measurement Device
    McMillan, Lachlan B.
    Von Treuheim, Theodor Di Pauli
    Murphy, Anna
    Zengin, Ayse
    Ebeling, Peter R.
    Scott, David
    IEEE SENSORS JOURNAL, 2019, 19 (11) : 4008 - 4016
  • [35] Result of IOFIX (Intra Osseous FIXation) device for first metatarsophalangeal joint arthrodesis: A single surgeon's series
    Singhal, Rohit
    Kwaees, Tariq
    Mohamed, Mohamed
    Argyropoulos, Miltiadis
    Amarasinghe, Pradeep
    Toh, E. M.
    FOOT AND ANKLE SURGERY, 2018, 24 (05) : 466 - 470
  • [36] Rate of Development of Hallucal Interphalangeal Degenerative Joint Disease After First Metatarsophalangeal Joint Arthrodesis: A Retrospective Radiographic Analysis
    Shah, Nisha N.
    Richardson, Marcus P.
    Chu, Anson K.
    Hyer, Christopher F.
    FOOT & ANKLE SPECIALIST, 2019, 12 (04) : 357 - 362
  • [37] Validity and reliability of a novel subtalar joint axis of rotation locator measurement device
    BH Kim
    SC Lee
    HD Lee
    SY Lee
    Journal of Foot and Ankle Research, 7 (Suppl 1)
  • [38] Development of a Gravity Compensation Device for Rotary Joint Using Magnetic Force
    Zhu, Weizheng
    Shan, Leimeng
    Lee, Kyung-Min
    TRANSACTIONS OF THE KOREAN SOCIETY OF MECHANICAL ENGINEERS A, 2023, 47 (07) : 559 - 564
  • [39] Development, Validity, and Reliability of a Novel Walking Speed Measurement Device: the GaitBox
    Jarvis, Leighanne M.
    Peterson, Matthew J.
    Caves, Kevin M.
    GAIT & POSTURE, 2021, 84 : 52 - 57
  • [40] Heel Height as an Etiology of Hallux Abductus Valgus Development: An electromagnetic Static and Dynamic First Metatarsophalangeal Joint Study
    Sanchez-Gomez, Ruben
    Becerro de Bengoa-Vallejo, Ricardo
    Elena Losa-Iglesias, Marta
    Calvo-Lobo, Cesar
    Romero-Morales, Carlos
    Maria Martinez-Jimenez, Eva
    Palomo-Lopez, Patricia
    Lopez-Lopez, Daniel
    SENSORS, 2019, 19 (06)