Investigations of a flat-panel detector for quality assurance measurements in ion beam therapy

被引:15
|
作者
Hartmann, Bernadette [1 ,2 ]
Telsemeyer, Julia [1 ,2 ]
Huber, Lucas [1 ]
Ackermann, Benjamin [3 ]
Jaekel, Oliver [1 ,2 ,3 ]
Martisikova, Maria [1 ]
机构
[1] German Canc Res Ctr, Dept Med Phys Radiat Oncol, D-69120 Heidelberg, Germany
[2] Univ Klinikum Heidelberg, Klin Radioonkol & Strahlentherapie, D-69120 Heidelberg, Germany
[3] Univ Klinikum Heidelberg, HIT, D-69120 Heidelberg, Germany
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2012年 / 57卷 / 01期
关键词
RADIOTHERAPY; MODEL;
D O I
10.1088/0031-9155/57/1/51
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Increased accuracy in radiation delivery to a patient provided by scanning particle beams leads to high demands on quality assurance (QA). To meet the requirements, an extensive quality assurance programme has been implemented at the Heidelberg Ion Beam Therapy Center. Currently, high-resolution radiographic films are used for beam spot position measurements and homogeneity measurements for scanned fields. However, given that using this film type is time and equipment demanding, considerations have been made to replace the radiographic films in QA by another appropriate device. In this study, the suitability of the flat-panel detector RID 256 L based on amorphous silicon was investigated as an alternative method. The currently used radiographic films were taken as a reference. Investigations were carried out for proton and carbon ion beams. The detectors were irradiated simultaneously to allow for a direct comparison. The beam parameters (e. g. energy, focus, position) currently used in the daily QA procedures were applied. Evaluation of the measurements was performed using newly implemented automatic routines. The results for the flat-panel detector were compared to the standard radiographic films. Additionally, a field with intentionally decreased homogeneity was applied to test the detector's sensitivities toward possible incorrect scan parameters. For the beam position analyses, the flat-panel detector results showed good agreement with radiographic films. For both detector types, deviations between measured and planned spot distances were found to be below 1% (1 mm). In homogeneously irradiated fields, the flat-panel detector showed a better dose response homogeneity than the currently used radiographic film. Furthermore, the flat-panel detector is sensitive to field irregularities. The flat-panel detector was found to be an adequate replacement for the radiographic film in QA measurements. In addition, it saves time and equipment because no post-exposure treatment and no developer and darkroom facilities are needed.
引用
收藏
页码:51 / 68
页数:18
相关论文
共 50 条
  • [1] Characterization of a flat-panel detector for ion beam spot measurements
    Martisikova, Maria
    Hartmann, Bernadette
    Hesse, Bernd M.
    Brons, Stephan
    Ackermann, Benjamin
    Jaekel, Oliver
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (02): : 485 - 497
  • [2] Quantitative carbon ion beam radiography and tomography with a flat-panel detector
    Telsemeyer, Julia
    Jaeakel, Oliver
    Martisikova, Maria
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (23): : 7957 - 7971
  • [3] Fluence verification for patient specific quality assurance in ion beam therapy. Use of an a-Si flat panel detector
    Horn, J.
    Ackermann, B.
    Brons, S.
    Lahrmann, S.
    Lampe, C.
    Ellerbrock, M.
    Jaekel, O.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S438 - S438
  • [4] Meeting the Demands of Highly Precise Ion Beam Therapy - How a Flat-Panel Detector can Improve Conformance Tests
    Hartmann, Bernadette
    Telsemeyer, Julia
    Ackermann, Benjamin
    Brons, Stephan
    Jaekel, Oliver
    Martisikova, Maria
    [J]. 2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 4308 - 4313
  • [5] Geometric Calibration of Cone-beam CT with a Flat-panel Detector
    Wu, Dufan
    Li, Liang
    Zhang, Li
    Xing, Yuxiang
    Chen, Zhiqiang
    Xiao, Yongshun
    [J]. 2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2952 - 2955
  • [6] Planar cone-beam computed tomography with a flat-panel detector
    Kim, S. H.
    Kim, D. W.
    Youn, H.
    Kim, D.
    Kam, S.
    Jeon, H.
    Kim, H. K.
    [J]. JOURNAL OF INSTRUMENTATION, 2015, 10
  • [7] Ring artifact corrections in flat-panel detector based cone beam CT
    Abu Anas, Emran Mohammad
    Kim, Jaegon
    Lee, Soo Yeol
    Hasan, Md. Kamrul
    [J]. MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [8] Comparison of flat-panel detector with image-intensifier detector for cone-beam CT
    Baba, R
    Konno, Y
    Ueda, M
    Ikeda, S
    [J]. RADIOLOGY, 2000, 217 : 405 - 405
  • [9] Comparison of flat-panel detector and image-intensifier detector for cone-beam CT
    Baba, R
    Konno, Y
    Ueda, K
    Ikeda, S
    [J]. COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2002, 26 (03) : 153 - 158
  • [10] Indirect flat-panel detector with avalanche gain
    Zhao, W
    Hunt, DC
    Tanioka, K
    Rowlands, JA
    [J]. MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 150 - 161