3D Monte Carlo dosimetry of intraoperative electron radiation therapy (IOERT)

被引:11
|
作者
Alhamada, Husein [1 ,2 ]
Simon, Stephane [1 ,2 ]
Philippson, Catherine [2 ]
Vandekerkhove, Christophe [2 ]
Jourani, Younes [2 ]
Pauly, Nicolas [1 ]
VanGestel, Dirk [2 ]
Reynaert, Nick [1 ,2 ]
机构
[1] Univ Libre Bruxelles, Ecole Polytech, Nucl Metrol, Brussels, Belgium
[2] Inst Jules Bordet, Radiotherapy Dept, Brussels, Belgium
关键词
DOSE CALCULATIONS; RADIOTHERAPY; IORT; DISTRIBUTIONS; APPLICATOR; PHOTON;
D O I
10.1016/j.ejmp.2018.12.037
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This paper studies the feasibility of using Monte Carlo (MC) for treatment planning of intraoperative electron radiation therapy (IOERT) procedure to get 3D dose by using patient's CT images. Methods: The IOERT treatment planning was performed using the following successive steps: I) The Mobetron 1000 degrees machine was modelled with the EGSnrc MC codes. II) The MC model was validated with measurements of percentage depth doses and profiles for three energies (12, 9, 6) MeV. III) CT images were imported as DICOM files. IV) Contouring of the planning target volume (PTV) and the organs at risk was done by the radiation oncologist. V) The medical physicist with the radiation oncologist, had chosen the same parameters of IOERT procedures like energy, applicator (type, size) and using or not bolus. VI) Finally, dose calculation and analysis of 3D maps was carried out. Results: The tuning process of the MC model provides good results, as the maximum value of the root mean square deviation (RMSD) was less than 3% between the MC simulated PDDs and the measured PDDs. The contouring and dose analysis review were easy to conduct for the classical treatment planning system. The radiation oncologist had many tools for dose analysis such as DVH and color wash for all the slides. Summation of the 3D dose of IOERT with other radiotherapy plans is possible and helpful for total dose estimation. Archiving and documentation is as good as treatment planning system (TPS). Conclusions: The method displayed in this paper provides a step forward for IOERT Dosimetry and allows to obtain accurate dosimetry of treated volumes.
引用
收藏
页码:207 / 214
页数:8
相关论文
共 50 条
  • [1] Monte Carlo dose calculations of shielding disks with different material combinations in intraoperative electron radiation therapy (IOERT)
    Alhamada, H.
    Simon, S.
    Philippson, C.
    Vandekerkhove, C.
    Jourani, Y.
    Pauly, N.
    Van Gestel, D.
    Reynaert, N.
    CANCER RADIOTHERAPIE, 2020, 24 (02): : 128 - 134
  • [2] Dosimetry verification of IntraOperative Radiation Therapy (IORT): a Monte Carlo Study
    Alhamada, H.
    Simon, S.
    Philippson, C.
    Vandekerkhove, C.
    Jourani, Y.
    Pauly, N.
    Reynaert, N.
    Van Gestel, D.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S976 - S976
  • [3] In vivo dosimetry using Gafchromic films during pelvic intraoperative electron radiation therapy (IOERT)
    Costa, Filipa
    Sarmento, Sandra
    Gomes, Dora
    Magalhaes, Helena
    Arrais, Rosario
    Moreira, Graciete
    Cruz, Maria Fatima
    Silva, Jose Pedro
    Santos, Lucio
    Sousa, Olga
    BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1063):
  • [4] Dosimetry characteristics of degraded electron beams investigated by Monte Carlo calculations in a setup for intraoperative radiation therapy
    Björk, P
    Nilsson, P
    Knöös, T
    PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (02): : 239 - 256
  • [5] Dosimetry verification of an applicator system for intraoperative radiation therapy by Monte Carlo simulation
    Nevelsky, A.
    Borzov, E.
    Ber-Deroma, R.
    Orion, I.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S280 - S280
  • [6] 3D Monte Carlo radiation transfer modelling of photodynamic therapy
    Campbell, C. Louise
    Christison, Craig
    Brown, C. Tom A.
    Wood, Kenneth
    Valentine, Ronan M.
    Moseley, Harry
    BIOPHOTONICS SOUTH AMERICA, 2015, 9531
  • [7] Development and verification of an electron Monte Carlo engine for applications in intraoperative radiation therapy
    Rank, Luisa
    Lysakovski, Peter
    Major, Gerald
    Ferrari, Alfredo
    Tessonnier, Thomas
    Debus, Juergen
    Mairani, Andrea
    MEDICAL PHYSICS, 2024, 51 (09) : 6348 - 6364
  • [8] Intraoperative electron radiation therapy (IOERT) in the quart sequence: A pilot study
    Odantini, R
    Mazzitelli, R
    Bellia, SR
    Marchione, A
    Di Pietro, A
    Al Sayyad, S
    Malara, S
    Capua, A
    EUROPEAN JOURNAL OF CANCER, 2002, 38 : S152 - S152
  • [9] Intraoperative Electron Radiation Therapy (IOERT) in the management of locally recurrent rectal cancer
    Falk Roeder
    Joerg-Michael Goetz
    Gregor Habl
    Marc Bischof
    Robert Krempien
    Markus W Buechler
    Frank W Hensley
    Peter E Huber
    Juergen Weitz
    Juergen Debus
    BMC Cancer, 12
  • [10] Intraoperative Electron Radiation Therapy (IOERT) in the management of locally recurrent rectal cancer
    Roeder, Falk
    Goetz, Joerg-Michael
    Habl, Gregor
    Bischof, Marc
    Krempien, Robert
    Buechler, Markus W.
    Hensley, Frank W.
    Huber, Peter E.
    Weitz, Juergen
    Debus, Juergen
    BMC CANCER, 2012, 12