An All Solid State Electrochemical pH Sensor Based on Niobium Modified Electrode

被引:5
|
作者
Xu, Kebin [1 ,2 ]
Wu, Binyu [1 ,2 ]
Wan, Junliang [1 ,2 ]
Li, Ying [1 ,2 ]
Li, Min [1 ,2 ]
机构
[1] Northeastern Univ, Sch Met, Shenyang 110819, Peoples R China
[2] Liaoning Key Lab Met Sensor Mat & Technol, Shenyang 110819, Peoples R China
基金
美国国家科学基金会;
关键词
Electrodes; Hydrogen; Niobium; Sodium; Ions; Hydrogen storage; Electric potential; Electronics and physics of transduction for chemical and biological signals pH sensor; sensor testing and evaluation; FIELD-EFFECT-TRANSISTOR; TEMPERATURE; CAPACITY; HEALTH; SALIVA; MILK; ACID; UREA;
D O I
10.1109/JSEN.2021.3123633
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
pH is an important chemical index. The measurement of pH is widely used in various fields. At present, the most common pH meter on the market is the glass electrode pH meter, which has many advantages, such as high accuracy, short response time and so on. However, when industrial wastewater and environmental water samples need long-time real-time monitoring, such as using glass electrode pH meter will lead to leakage of internal liquid, fragile glass and sodium difference. Therefore, it is urgent to design and develop a new type of pH meter with all solid, simple structure and no nanodeviation. In this study, an all solid state electrochemical modified pH electrode was developed based on the adsorption and storage of hydrogen by niobium. Hydrogen in hydrochloric acid was adsorbed and stored in niobium electrode by constant potential electrolysis. The results showed that the detection linear range was pH1-pH13, the response slope was -56.4 mV dec(-1), and the response time was 5 seconds. In the actual sample detection, the detection results are consistent with the glass electrode pH meter. In addition, cyclic voltammetry and electron microscopy were used to study the response mechanism.
引用
收藏
页码:27275 / 27281
页数:7
相关论文
共 50 条
  • [31] An all-solid-state potentiometric SO2 gas sensor with solid reference electrode
    Ling Wang
    R. V. Kumar
    Ionics, 2003, 9 : 357 - 364
  • [32] An all-solid-state potentiometric SO2 gas sensor with solid reference electrode
    Wang, L
    Kumar, RV
    IONICS, 2003, 9 (5-6) : 357 - 364
  • [33] A new solid-state pH sensor and its application in the construction of all solid-state urea biosensor
    Pandey, PC
    Upadhyay, S
    Singh, G
    Prakash, R
    Srivastava, RC
    Seth, PK
    ELECTROANALYSIS, 2000, 12 (07) : 517 - 521
  • [34] Electrochemical Sensor for Hydroquinone based on Polyeugenol/PVA-Modified Graphite Electrode
    Marwati, Siti
    Trisunaryanti, Wega
    Louise, Isana Supiah Yosephine
    Siswanta, Dwi
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2020, 15 (11): : 11509 - 11521
  • [35] A sensitive electrochemical DNA sensor based on reduced graphene oxide modified electrode
    Xu, Binxiang
    Hu, Yuping
    Shu, Qingxia
    Wang, Mei
    Chen, Zhiyang
    Wei, Wei
    Wen, Jinmei
    Li, Rui
    Liao, Fusheng
    Cheng, Lin
    Fan, Hao
    JOURNAL OF THE CHINESE CHEMICAL SOCIETY, 2022, 69 (05) : 822 - 830
  • [36] Hydroxylamine electrochemical sensor based on magnesium aluminate spinel nanoparticles modified electrode
    Riahi, Ramin
    Taher, Mohammad Ali
    Beitollahi, Hadi
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 2022, 104 (17) : 5172 - 5184
  • [37] An Electrochemical Sensor Based on Gold Nanodendrite/Surfactant Modified Electrode for Bisphenol A Detection
    Nguyen Thi Lien
    Le Quoc Hung
    Nguyen Tien Hoang
    Vu Thi Thu
    Dau Thi Ngoc Nga
    Pham Thi Hai Yen
    Pham Hong Phong
    Vu Thi Thu Ha
    JOURNAL OF ANALYTICAL METHODS IN CHEMISTRY, 2020, 2020
  • [38] Electrochemical sensor based on phospholipid modified glassy carbon electrode - determination of paraquat
    Tomkova, H.
    Sokolova, R.
    Opletal, T.
    Kucerova, P.
    Kucera, L.
    Souckova, J.
    Skopalova, J.
    Bartak, P.
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2018, 821 : 33 - 39
  • [39] Electrochemical sensor for hydroperoxides determination based on Prussian blue film modified electrode
    Adhoum, Nafaa
    Monser, Lotfi
    SENSORS AND ACTUATORS B-CHEMICAL, 2008, 133 (02): : 588 - 592
  • [40] An electrochemical DNA sensor based on a layers-film construction modified electrode
    Zhang, Yi
    Zeng, Guang-Ming
    Tang, Lin
    Li, Yuan-Ping
    Chen, Li-Juan
    Pang, Ya
    Li, Zhen
    Feng, Chong-Ling
    Huang, Guo-He
    ANALYST, 2011, 136 (20) : 4204 - 4210