An All Solid State Electrochemical pH Sensor Based on Niobium Modified Electrode

被引:5
|
作者
Xu, Kebin [1 ,2 ]
Wu, Binyu [1 ,2 ]
Wan, Junliang [1 ,2 ]
Li, Ying [1 ,2 ]
Li, Min [1 ,2 ]
机构
[1] Northeastern Univ, Sch Met, Shenyang 110819, Peoples R China
[2] Liaoning Key Lab Met Sensor Mat & Technol, Shenyang 110819, Peoples R China
基金
美国国家科学基金会;
关键词
Electrodes; Hydrogen; Niobium; Sodium; Ions; Hydrogen storage; Electric potential; Electronics and physics of transduction for chemical and biological signals pH sensor; sensor testing and evaluation; FIELD-EFFECT-TRANSISTOR; TEMPERATURE; CAPACITY; HEALTH; SALIVA; MILK; ACID; UREA;
D O I
10.1109/JSEN.2021.3123633
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
pH is an important chemical index. The measurement of pH is widely used in various fields. At present, the most common pH meter on the market is the glass electrode pH meter, which has many advantages, such as high accuracy, short response time and so on. However, when industrial wastewater and environmental water samples need long-time real-time monitoring, such as using glass electrode pH meter will lead to leakage of internal liquid, fragile glass and sodium difference. Therefore, it is urgent to design and develop a new type of pH meter with all solid, simple structure and no nanodeviation. In this study, an all solid state electrochemical modified pH electrode was developed based on the adsorption and storage of hydrogen by niobium. Hydrogen in hydrochloric acid was adsorbed and stored in niobium electrode by constant potential electrolysis. The results showed that the detection linear range was pH1-pH13, the response slope was -56.4 mV dec(-1), and the response time was 5 seconds. In the actual sample detection, the detection results are consistent with the glass electrode pH meter. In addition, cyclic voltammetry and electron microscopy were used to study the response mechanism.
引用
收藏
页码:27275 / 27281
页数:7
相关论文
共 50 条
  • [1] New NASICON-based all-solid-state reference electrode towards a pH sensor for the electrochemical measurement of harsh environments
    Jiayan Miao
    Cuiyan Yu
    Jinfeng Xia
    Zhenhai Xue
    Xin Li
    Mengting Dai
    Yuerong Heng
    Danyu Jiang
    Qiang Li
    Ionics, 2022, 28 : 2437 - 2445
  • [2] New NASICON-based all-solid-state reference electrode towards a pH sensor for the electrochemical measurement of harsh environments
    Miao, Jiayan
    Yu, Cuiyan
    Xia, Jinfeng
    Xue, Zhenhai
    Li, Xin
    Dai, Mengting
    Heng, Yuerong
    Jiang, Danyu
    Li, Qiang
    IONICS, 2022, 28 (05) : 2437 - 2445
  • [3] All-solid-state pH Sensor Based on Conducting Polymer
    Natedungta, W.
    Prissanaroon-Ouajai, W.
    FUNCTIONALIZED AND SENSING MATERIALS, 2010, 93-94 : 591 - 594
  • [4] All-solid-state potassium-selective sensor based on carbon black modified thermoplastic electrode
    Ozer, Tugba
    Henry, Charles S.
    ELECTROCHIMICA ACTA, 2022, 404
  • [5] All solid-state pH electrode based on titanium nitride sensitive film
    Wang, Yan
    Yuan, Hongyan
    Lu, Xulin
    Zhou, Zaide
    Xiao, Dan
    ELECTROANALYSIS, 2006, 18 (15) : 1493 - 1498
  • [6] Solid state pH and chloride sensor with microfluidic reference electrode
    Zevenbergen, M. A. G.
    Altena, G.
    Dam, V. A. T.
    Goedbloed, M.
    Bembnowicz, P.
    McGuinness, P.
    Berney, H.
    Berduque, A.
    O'Dwyer, T.
    van Schaijk, R.
    2016 IEEE INTERNATIONAL ELECTRON DEVICES MEETING (IEDM), 2016,
  • [7] Development and Performance of an All-Solid-Stated pH Sensor Based on Modified Membranes
    Xu, Kun
    Zhang, Xiliang
    Chen, Cheng
    Geng, Miaomiao
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2018, 13 (03): : 3080 - 3090
  • [8] Potentiometric pH sensor based on electrochemical properties of carbon nanotubes modified electrode/guanine system
    Zhang Xu-Zhi
    Zhao, Chang-Zhi
    Kui, Jiao
    Li Shuang-De
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2008, 36 (02) : 197 - 200
  • [9] Effects of Surface Area on All-Solid-Stated pH Sensor Based on Antimony Electrode
    Zhang, Xiliang
    Lu, Chengxuan
    Geng, Miaomiao
    Xu, Kun
    Zong, Shengkang
    IEEE SENSORS JOURNAL, 2020, 20 (02) : 680 - 688
  • [10] Electrochemical sensor based on cerium niobium oxide nanoparticles modified electrode for sensing of environmental toxicity in water samples
    Manjula, Natesan
    Pulikkutty, Subramaniyan
    Chen, Tse-Wei
    Chen, Shen-Ming
    Fan, Chi Hsuan
    Ali, M. Ajmal
    Al-Hemaid, Fahad M.
    COLLOIDS AND SURFACES A-PHYSICOCHEMICAL AND ENGINEERING ASPECTS, 2022, 637