Technical Note: Nuclear imaging with an x-ray flat panel detector: A proof-of-concept study

被引:2
|
作者
Dietze, Martijn M. A. [1 ,2 ,3 ]
Koppert, Wilco J. C. [1 ,2 ]
van Rooij, Rob [1 ,2 ]
de Jong, Hugo W. A. M. [1 ,2 ]
机构
[1] Univ Utrecht, Radiol & Nucl Med, POB 85500, Utrecht 3508 GA, Netherlands
[2] Univ Med Ctr Utrecht, POB 85500, Utrecht 3508 GA, Netherlands
[3] Univ Utrecht, Image Sci Inst, POB 85500, Utrecht 3508 GA, Netherlands
基金
欧洲研究理事会;
关键词
CBCT; collimator; flat panel; Gamma camera; SPECT; GAMMA-CAMERA ARCHITECTURE; OPTIMIZATION;
D O I
10.1002/mp.14191
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Interventional procedures involving radionuclides (e.g., radioembolization) would benefit from single-photon emission computed tomography (SPECT) performed in the intervention room because the activity distribution could be immediately visualized. We believe it might be possible to perform SPECT with the C-arm cone beam computed tomography (CBCT) scanner present in the intervention room by equipping the x-ray flat panel detector with a collimator. The purpose of this study is to demonstrate the approach and to investigate the achievable SPECT reconstruction quality. Methods A proof-of-concept experiment was performed to evaluate the possibility of nuclear imaging with an x-ray flat panel detector. The experiment was digitally replicated to study the accuracy of the simulations. Three flat panel configurations (with standard hardware and reconstruction methodology, with sophisticated reconstruction methodology, and with expected future hardware) and a conventional gamma camera were evaluated. The Jaszczak and the NEMA IQ phantom (filled with Tc-99m) were simulated and assessed on resolution and contrast-to-noise ratio (CNR). Results The proof-of-concept experiment demonstrated that nuclear images could be obtained from the flat panel detector. The simulation of the same configuration demonstrated that simulations could accurately predict the flat panel detector response. The CNR of the 37 mm sphere in the NEMA IQ phantom was 22.8 +/- 1.2 for the gamma camera reconstructions, while it was 11.3 +/- 0.7 for the standard flat panel detector. With sophisticated reconstruction methodology, the CNR improved to 13.5 +/- 1.4. The CNR can be expected to advance to 18.1 +/- 1.3 for future flat panel detectors. Conclusions The x-ray flat panel detector of a CBCT scanner might be used to perform nuclear imaging. The SPECT reconstruction quality will be lower than that achieved by a conventional gamma camera. The flat panel detector approach could, however, be useful in providing a cost-effective alternative to the purchase of a mobile SPECT scanner for enabling interventional scanning.
引用
收藏
页码:3363 / 3368
页数:6
相关论文
共 50 条
  • [31] Physical Characteristics of a New Flat-Panel Digital X-Ray Detector
    Cho, H.
    Kim, H.
    Park, H.
    Lee, C.
    Yu, A.
    Kim, D.
    Lee, Y.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [32] A Method for Measuring X-Ray Beam Profiles Using a Flat Panel Detector
    Sun, M.
    Star-Lack, J.
    Virshup, G.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [33] Measurement of the MTF of a flat panel detector in variable resolution x-ray detection
    Zhang, S.
    DiBianca, F.
    MEDICAL PHYSICS, 2007, 34 (06) : 2620 - 2620
  • [34] Proof-of-concept Study of Smart Panel for Space Structures
    Mizutani, T.
    STRUCTURAL HEALTH MONITORING 2013, VOLS 1 AND 2, 2013, : 437 - 444
  • [35] High resolution flat-panel X-ray detector for digital radiography
    Park, JK
    Kang, SS
    Cha, BY
    Kim, JY
    Kim, SY
    Choi, HK
    Nam, SH
    Healthcom 2005: 7th International Workshop on Enterprise Networking and Computing in Healthcare Industry, Proceedings, 2005, : 359 - 363
  • [36] Evaluation of a new radiographic system based on a flat panel X-ray detector
    Maack, I
    GuntherKohfahl, S
    Schroeder, H
    Spiess, K
    Neitzel, U
    RADIOLOGY, 1997, 205 : 109 - 109
  • [37] Simulation Study of Plasma Display Panel with GMD Structure for X-Ray Imaging Detector
    Lee, Hakjae
    Min, Eungi
    Lee, Kisung
    Eom, Sangheum
    Park, Hanho
    Kang, Jungwon
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 3223 - 3227
  • [38] X-ray Multi-energy Imaging Method Using Multi-layer Flat Panel Detector
    Wang Zongpeng
    Lan Chongzhou
    Wen Minru
    Ye Chao
    ACTA PHOTONICA SINICA, 2021, 50 (09) : 215 - 223
  • [39] Proof of concept of an innovative detector for nuclear imaging applications
    Pirrone, G.
    NUOVO CIMENTO C-COLLOQUIA AND COMMUNICATIONS IN PHYSICS, 2014, 37 (04): : 175 - 182
  • [40] Technical assessment of 2D and 3D imaging performance of an IGZO-based flat-panel X-ray detector
    Sheth, Niral Milan
    Uneri, Ali
    Helm, Patrick A.
    Zbijewski, Wojciech
    Siewerdsen, Jeffrey H.
    MEDICAL PHYSICS, 2022, 49 (05) : 3053 - 3066