Potentiometric biosensor for lysine analysis based on a chemically immobilized lysine oxidase membrane

被引:32
|
作者
Saurina, J
Hernández-Cassou, S
Fàbregas, E
Alegret, S
机构
[1] Univ Barcelona, Dept Analyt Chem, E-08028 Barcelona, Spain
[2] Univ Autonoma Barcelona, Dept Chem, E-08193 Barcelona, Spain
关键词
potentiometric lysine biosensor; immobilized lysine oxidase membranes; all-solid-state ammonium electrode; lysine analysis;
D O I
10.1016/S0003-2670(98)00310-9
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An enzymatic lysine electrode is developed for the potentiometric analysis of lysine (Lys). This lysine biosensor consists of a chemically immobilized lysine oxidase membrane attached to an all-solid-state ammonium electrode. In this system, the enzymatic degradation of lysine produces ammonia, which is then detected with the ammonium electrode. The influence of immobilization conditions on the potentiometric response and the lifetime of the electrode was studied. Under optimal conditions, the relationship between the potentiometric response and the log [Lys] was linear between 3x10(-5) and 10(-3) M, the sensitivity was 50 mV/decade, the repeatability 1.4%, and the response time t(95%) was 15 s. (C) 1998 Elsevier Science B.V, All rights reserved.
引用
收藏
页码:49 / 56
页数:8
相关论文
共 50 条
  • [1] Fabrication and application of an amperometric lysine biosensor based on covalently immobilized lysine oxidase nanoparticles onto Au electrode
    Nohwal, Bhawna
    Chaudhary, Reeti
    Kumar, Parveen
    Pundir, C. S.
    INTERNATIONAL JOURNAL OF BIOLOGICAL MACROMOLECULES, 2020, 146 : 907 - 915
  • [2] FLOW-INJECTION AMPEROMETRIC BIOSENSOR BASED ON IMMOBILIZED L-LYSINE-ALPHA-OXIDASE FOR L-LYSINE DETERMINATION
    SIMONIAN, AL
    BADALIAN, IE
    BEREZOV, TT
    SMIRNOVA, IP
    KHADUEV, SH
    ANALYTICAL LETTERS, 1994, 27 (15) : 2849 - 2860
  • [3] Flow injection differential potentiometric determination of lysine by using a lysine biosensor
    García-Villar, N
    Saurina, J
    Hernández-Cassou, S
    ANALYTICA CHIMICA ACTA, 2003, 477 (02) : 315 - 324
  • [4] Amperometric determination of lysine using a lysine oxidase biosensor based on rigid-conducting composites
    Saurina, J
    Hernández-Cassou, S
    Alegret, S
    Fàbregas, E
    BIOSENSORS & BIOELECTRONICS, 1999, 14 (02): : 211 - 220
  • [5] CHEMILUMINOMETRIC L-LYSINE DETERMINATION WITH IMMOBILIZED LYSINE OXIDASE BY FLOW-INJECTION ANALYSIS
    PREUSCHOFF, F
    SPOHN, U
    WEBER, E
    UNVERHAU, K
    MOHR, KH
    ANALYTICA CHIMICA ACTA, 1993, 280 (02) : 185 - 189
  • [6] AN OPTICAL BIOSENSOR FOR LYSINE BASED ON THE USE OF LYSINE DECARBOXYLASE AND A CADAVERINE-SENSITIVE MEMBRANE
    LI, H
    HE, HR
    WOLFBEIS, OS
    BIOSENSORS & BIOELECTRONICS, 1992, 7 (10): : 725 - 732
  • [7] Liquid chromatographic determination of lysine by potentiometric detection with a biosensor
    García-Villar, N
    Saurina, J
    Hernández-Cassou, S
    ANALYTICAL LETTERS, 2002, 35 (08) : 1313 - 1325
  • [8] Determination of lysine in pharmaceutical samples containing endogenous ammonium ions by using a lysine oxidase biosensor based on an all-solid-state potentiometric ammonium electrode
    Saurina, J
    Hernández-Cassou, S
    Alegret, S
    Fàbregas, E
    BIOSENSORS & BIOELECTRONICS, 1999, 14 (01): : 67 - 75
  • [9] POTENTIOMETRIC BIOSENSOR BASED ON IMMOBILIZED ENZYME MEMBRANE AND FLUORIDE DETECTION
    HO, MH
    SENSORS AND ACTUATORS, 1988, 15 (04): : 445 - 450
  • [10] Flow injection amperometric and chemiluminescence individual and simultaneous determination of lysine and glucose with immobilized lysine oxidase and glucose oxidase
    Almuaibed, AM
    Townshend, A
    ANALYTICA CHIMICA ACTA, 1997, 338 (1-2) : 149 - 154