Effect of Inversion Recovery Fat Suppression on Hepatic R2*Quantitation in Transfusional Siderosis

被引:19
|
作者
Meloni, Antonella [1 ,2 ]
Tyszka, J. Michael [3 ]
Pepe, Alessia [1 ]
Wood, John C. [2 ,4 ]
机构
[1] Fdn G Monasterio CNR Reg Toscana, CMR Unit, Pisa, Italy
[2] Childrens Hosp Los Angeles, Dept Pediat, Div Cardiol, Los Angeles, CA 90027 USA
[3] CALTECH, Div Biol, Pasadena, CA 91125 USA
[4] Childrens Hosp Los Angeles, Dept Radiol, Los Angeles, CA 90027 USA
基金
美国国家卫生研究院;
关键词
fat suppression; iron overload; liver; R2*; MAGNETIC-RESONANCE; IRON OVERLOAD; LIVER IRON; THALASSEMIA MAJOR; MRI; QUANTIFICATION; WATER; RELAXOMETRY; DIAGNOSIS; DISEASE;
D O I
10.2214/AJR.14.12786
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of this study is to evaluate whether the application of spectral pre-saturation inversion recovery (SPIR) fat suppression in standard multiecho gradient-echo sequences has a significant effect on hepatic R2* quantitation in patients with iron overload syndromes. MATERIALS AND METHODS. Eighty patients were scanned with a multiecho gradient-echo sequence without and with the application of SPIR. Six different postprocessing approaches were used to extract R2* values for maximum generality. RESULTS. SPIR fat suppression lowered R2* values by 3.9-7.0% (p < 0.0001 in all pair-wise comparisons), independently of the postprocessing algorithm. Coefficients of variation for R2* ranged from 4.5% to 10.0%. Regardless of the size of the ROI (area of homogeneous tissue or entire liver profile in the slice), pixelwise approaches combined with an exponential-plus-constant fitting model yielded the lowest coefficients of variation (4.5% and 5.1%), whereas truncated exponential fits of the averaged signals produced the highest coefficients of variation (7.8% and 10%). For R2* values exceeding 200 Hz, a Bland-Altman analysis showed a bias that grew linearly for all postprocessing methods. CONCLUSION. SPIR fat suppression resulted in systematically lower hepatic R2* estimates. Because calibration curves were derived using images without fat suppression, these biases should be corrected when reporting liver iron concentrations estimated from fat-suppressed multiecho T2*-weighted images.
引用
收藏
页码:625 / 629
页数:5
相关论文
共 50 条
  • [1] Effect of hepatocyte-specific gadolinium-based contrast agents on hepatic fat-fraction and R2
    Hernando, Diego
    Wells, Shane A.
    Vigen, Karl K.
    Reeder, Scott B.
    MAGNETIC RESONANCE IMAGING, 2015, 33 (01) : 43 - 50
  • [2] Does Fat Suppression via Chemically Selective Saturation Affect R2*-MRI for Transfusional Iron Overload Assessment? A Clinical Evaluation at 1.5T and 3T
    Krafft, Axel J.
    Loeffler, Ralf B.
    Song, Ruitian
    Bian, Xiao
    McCarville, M. Beth
    Hankins, Jane S.
    Hillenbrand, Claudia M.
    MAGNETIC RESONANCE IN MEDICINE, 2016, 76 (02) : 591 - 601
  • [3] APPROPRIATE EVALUATION OF TRANSFUSIONAL CARDIAC IRON OVERLOAD BY T2*AND R2*MAGNETIC RESONANCE
    Sakuta, J.
    Ito, Y.
    Kimura, Y.
    Park, J.
    Tokuuye, K.
    Ohyashiki, K.
    HAEMATOLOGICA-THE HEMATOLOGY JOURNAL, 2010, 95 : 710 - 711
  • [4] R2* Impact on Hepatic Fat Quantification With a Commercial Single Voxel Technique at 1.5 and 3.0 T
    Fortier, Veronique
    Mohamed, Ahmed
    McNabb, Evan
    Dana, Jeremy
    Zakarian, Rita
    Levesque, Ives
    Reinhold, Caroline
    CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 2024, 75 (04): : 838 - 846
  • [5] Cross-Sectional Correlation Between Hepatic R2☆ and Proton Density Fat Fraction (PDFF) in Children With Hepatic Steatosis
    Mamidipalli, Adrija
    Hamilton, Gavin
    Manning, Paul
    Hong, Cheng William
    Park, Charlie C.
    Wolfson, Tanya
    Hooker, Jonathan
    Heba, Elhamy
    Schlein, Alexandra
    Gamst, Anthony
    Durelle, Janis
    Paiz, Melissa
    Middleton, Michael S.
    Schwimmer, Jeffrey B.
    Sirlin, Claude B.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2018, 47 (02) : 418 - 424
  • [6] Double Inversion Recovery MRI with Fat Suppression at 7 Tesla: Initial Experience
    Madelin, Guillaume
    Oesingmann, Niels
    Inglese, Matilde
    JOURNAL OF NEUROIMAGING, 2010, 20 (01) : 87 - 92
  • [7] Nonspecificity of short inversion time inversion recovery (STIR) as a technique of fat suppression: Pitfalls in image interpretation
    Krinsky, G
    Rofsky, NM
    Weinreb, JC
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 166 (03) : 523 - 526
  • [8] Relationships between MR transverse relaxation parameters R2*, R2 and R2′ and hepatic iron content in thalassemic mice at 1.5 T and 3 T
    Song, Ruitian
    Lin, Wei
    Chen, Qiukan
    Asakura, Toshio
    Wehrli, Felix W.
    Song, Hee Kwon
    NMR IN BIOMEDICINE, 2008, 21 (06) : 574 - 580
  • [9] Optimal Fat Suppression in Head and Neck MRI: Comparison of Multipoint Dixon with 2 Different Fat-Suppression Techniques, Spectral Presaturation and Inversion Recovery, and STIR
    Gaddikeri, S.
    Mossa-Basha, M.
    Andre, J. B.
    Hippe, D. S.
    Anzai, Y.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2018, 39 (02) : 362 - 368
  • [10] R2*imaging of transfusional iron burden at 3T and comparison with 1.5T
    Storey, Pippa
    Thompson, Alexis A.
    Carqueville, Christine L.
    Wood, John C.
    de Freitas, R. Andrew
    Rigsby, Cynthia K.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 25 (03) : 540 - 547