摘要
OBJECTIVE. The purpose of this study is to evaluate whether the application of spectral pre-saturation inversion recovery (SPIR) fat suppression in standard multiecho gradient-echo sequences has a significant effect on hepatic R2* quantitation in patients with iron overload syndromes. MATERIALS AND METHODS. Eighty patients were scanned with a multiecho gradient-echo sequence without and with the application of SPIR. Six different postprocessing approaches were used to extract R2* values for maximum generality. RESULTS. SPIR fat suppression lowered R2* values by 3.9-7.0% (p < 0.0001 in all pair-wise comparisons), independently of the postprocessing algorithm. Coefficients of variation for R2* ranged from 4.5% to 10.0%. Regardless of the size of the ROI (area of homogeneous tissue or entire liver profile in the slice), pixelwise approaches combined with an exponential-plus-constant fitting model yielded the lowest coefficients of variation (4.5% and 5.1%), whereas truncated exponential fits of the averaged signals produced the highest coefficients of variation (7.8% and 10%). For R2* values exceeding 200 Hz, a Bland-Altman analysis showed a bias that grew linearly for all postprocessing methods. CONCLUSION. SPIR fat suppression resulted in systematically lower hepatic R2* estimates. Because calibration curves were derived using images without fat suppression, these biases should be corrected when reporting liver iron concentrations estimated from fat-suppressed multiecho T2*-weighted images.
引用
收藏
页码:625 / 629
页数:5
相关论文
共 50 条
  • [21] Estimation of cardiac left ventricular ejection fraction in transfusional cardiac iron overload by R2* magnetic resonance
    Sakuta, Juri
    Ito, Yoshikazu
    Kimura, Yukihiko
    Park, Jinho
    Tokuuye, Koichi
    Ohyashiki, Kazuma
    INTERNATIONAL JOURNAL OF HEMATOLOGY, 2010, 92 (05) : 708 - 712
  • [22] Inversion of the pair of weighted and classical circular Radon transforms in C(R2)
    Aramyan, Rafik
    JOURNAL OF MATHEMATICAL ANALYSIS AND APPLICATIONS, 2024, 530 (02)
  • [23] On the Inversion of Generalized V-Line Transform of a Vector Field in R2
    Bhardwaj, Rahul
    Mishra, Rohit Kumar
    Vashisth, Manmohan
    MATHEMATICAL METHODS IN THE APPLIED SCIENCES, 2025, 48 (06) : 6512 - 6520
  • [24] Determination of the R2* relaxation rate constant for estimating hepatic iron concentration: A customized approach that considers liver fat infiltration
    Gkotsis, D. E.
    Gotsis, E. D.
    Lymperopoulou, G.
    Karaiskos, P.
    Seimenis, I
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 76 : 150 - 158
  • [25] Impact of particle size on R2* and fat fraction estimation for accurate assessment of hepatic iron overload and steatosis using MRI
    Shrestha, Utsav
    Brasher, Sarah
    Abramson, Zachary
    Morin, Cara E.
    Tipirneni-Sajja, Aaryani
    MAGNETIC RESONANCE IN MEDICINE, 2025, 93 (05) : 2176 - 2185
  • [26] The intelligibility of r or r2 as an effect size statistic: dichotomous variables
    Trafimow, David
    FRONTIERS IN PSYCHOLOGY, 2015, 6
  • [27] Evaluation of optimized inversion-recovery fat-suppression techniques for T2-weighted abdominal MR imaging
    Lauenstein, Thomas C.
    Sharma, Puneet
    Hughes, Timothy
    Heberlein, Keith
    Tudorascu, Dana
    Martin, Diego R.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 27 (06) : 1448 - 1454
  • [28] Comparison of short inversion time inversion recovery (STIR) and fat-saturated (chemsat) techniques for background fat intensity suppression in cervical and thoracic MR imaging
    Nakatsu, M
    Hatabu, H
    Itoh, H
    Morikawa, K
    Miki, Y
    Kasagi, K
    Shimono, T
    Shoji, K
    Shimada, Y
    Imamura, M
    Konishi, J
    JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 2000, 11 (01): : 56 - 60
  • [29] Suppression of pain and the R2 component of the blink reflex during optokinetic stimulation
    Drummond, PD
    CEPHALALGIA, 2004, 24 (01) : 44 - 51
  • [30] Free-breathing MRI techniques for fat and R2* quantification in the liver
    Shih, Shu-Fu
    Wu, Holden H.
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2024, 37 (04): : 583 - 602