Assessment of skyshine photon dose rates from 9 and 18 MV medical linear accelerators

被引:3
|
作者
Rostampour, N. [1 ]
Jafari, S. [2 ]
Saeb, M. [3 ]
Keshtkar, M. [4 ]
Shokrani, P. [5 ]
Ahnasi, T. [1 ]
机构
[1] Kermanshah Univ Med Sci, Dept Med Phys, Kermanshah, Iran
[2] Hamadan Univ Med Sci, Dept Radiol, Hamadan, Iran
[3] Isfahan Univ Med Sci, Seyed Al Shohada Hosp, Dept Radiat Oncol, Esfahan, Iran
[4] Gonabad Univ Med Sci, Fac Med, Dept Med Phys & Radiol, Gonabad, Iran
[5] Isfahan Univ Med Sci, Dept Med Phys, Esfahan, Iran
来源
关键词
Skyshine; shielding; radiation measurement; radiation protection; radiotherapy; scattering; medical accelerator; ROOF;
D O I
10.18869/acadpub.ijrr.16.4.499
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Skyshine describes the radiation scattered by the atmosphere above a LINAC facility to a point on the ground. The aim of this study was to measure the skyshine photon dose rates from two different (9 MV and 18 MV) medical linear accelerators. Materials and Methods: The photon beam was directed upward (180 degrees gantry position), with a maximum photon field size (40 x 40 cm2) at the isocenter. Measurements were obtained around the external points selected outside the room facilities at a horizontal distance from the target by the calibrated RDS-110 survey meter at four points around the isocenter. Results: The measured values of the skyshine photon exposure rates at four points for 9 MV and 18 MV were 0.6, 0.5, 0.5, and 0.4 mu Sv/h, and 0.6, 0.4, 0.4, and 0.5 mu Sv/h, respectively. All the measured skyshine photon exposure rates were lower than the values recommended by NCRP 147. Conclusion: There is a poor agreement between the measured and the calculated values; therefore it seems that caution is needed while using the equations available in NCRP 147 or 151.
引用
收藏
页码:499 / 503
页数:5
相关论文
共 50 条
  • [41] Measurements of the neutron dose near a 15 Mv medical linear accelerator
    Golnik, N.
    Zielczynski, M.
    Bulski, W.
    Tulik, P.
    Palko, T.
    RADIATION PROTECTION DOSIMETRY, 2007, 126 (1-4) : 619 - 622
  • [42] SURFACE DOSE FROM 6 MV PHOTON INTERACTIONS IN AIR
    LARIVIERE, PD
    PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (03): : 285 - 287
  • [43] Photoneutrons and Gamma Capture Dose Rates at the Maze Entrance of Varian TrueBeam and Elekta Versa HD Medical Linear Accelerators
    Suliman, Ibrahim I. I.
    Khouqeer, Ghada A. A.
    Mayhoub, Fareed H. H.
    TOXICS, 2023, 11 (01)
  • [44] Ambient neutron dose equivalent outside concrete vault rooms for 15 and 18 MV radiotherapy accelerators
    Martinez-Ovalle, S. A.
    Barquero, R.
    Gomez-Ros, J. M.
    Lallena, A. M.
    RADIATION PROTECTION DOSIMETRY, 2012, 148 (04) : 457 - 464
  • [45] Optimization of Target Design and Dose Rates for FLASH-RT of An 18 MV Medical Linear Accelerator Model Via GEANT4 Monte Carlo Simulation
    Li, A.
    Kutsaev, S.
    Yu, V.
    Boucher, S.
    Sheng, K.
    MEDICAL PHYSICS, 2022, 49 (06) : E548 - E548
  • [46] Performance of Radiation Dosimeters in Gradient Regions at Different Dose Rates of Linear Accelerators
    Ahmed E.M.B.
    Abouelenein H.
    Khalil M.M.
    El-Nagdy M.S.
    Ammar H.
    Iranian Journal of Medical Physics, 2022, 19 (02) : 66 - 73
  • [47] Surface dose measurements with GafChromic EBT film for 6 and 18 MV photon beams
    Bilge, Hatice
    Cakir, Aydin
    Okutan, Murat
    Acar, Hilal
    PHYSICA MEDICA, 2009, 25 (02) : 101 - 104
  • [48] ASSESSMENT OF THE STABILITY OF REFERENCE DOSES OF A 18-MV PHOTON-BEAM
    MARTI, MG
    NUNEZ, L
    ISPIZUA, MC
    PAREDES, MC
    RADIOLOGIA, 1982, 24 (03): : 265 - 268
  • [49] Measurement of the leakage from linear accelerators in the backward direction for 4, 6, 10, 15 and 18 MV x-rays
    Jaradat, A
    Biggs, P
    MEDICAL PHYSICS, 2005, 32 (06) : 2023 - 2023
  • [50] MCNPX Estimation of Photoneutron Dose to Eye Voxel Anthropomorphic Phantom From 18 MV Linear Accelerator
    Alghamdi, Ali A. A.
    DOSE-RESPONSE, 2023, 21 (02):