New Non-contact Tonometer HNT-1P Reliability: Comparing Intraocular Pressure, Central Corneal Thickness, and Corrected Intraocular Pressure

被引:2
|
作者
Eah, Kyu Sang [1 ]
Shin, Joong Won [1 ]
Sung, Kyung Rim [1 ]
机构
[1] Univ Ulsan, Coll Med, Dept Ophthalmol Asan, Asan Med Ctr, Seoul, South Korea
来源
关键词
Central corneal thickness; Corrected intraocular pressure; HNT-1P; Intraocular pressure; Tonometry; GOLDMANN APPLANATION TONOMETRY; DYNAMIC CONTOUR TONOMETRY; BIOMECHANICAL PROPERTIES; ULTRASOUND PACHYMETRY; AGREEMENT; PARAMETERS;
D O I
10.3341/jkos.2020.61.5.524
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the reliability of a new non-contact tonometer, HNT-1P, by comparing its intraocular pressure (IOP), central corneal thickness (CCT), and corrected IOP (cIOP) measurements with the IOP measured by Goldmann applanation tonometry (GAT), the CCT measured with ultrasound pachymetry (USP), and the cIOP derived from GAT IOP and USP CCT. Methods: A total of 120 eyes of 65 healthy or glaucoma suspects were included in this study. The IOP was measured with the HNT-1P and GAT. The CCT was determined using the HNT-1P and USP. The IOP measured with GAT was corrected according to the CCT measured by USP. The correlations of the measurements obtained with the various instruments were evaluated using Pearson's correlation coefficient, and the agreement was assessed using Bland-Altman plots. Results: The average IOP measured with the HNT-1P and GAT was 14.8 +/- 5.1 mmHg and 15.6 +/- 5.2 mmHg, respectively, and the correlation coefficient between the two IOP measurements was 0.964 (p < 0.001). The mean CCT was 553.5 +/- 27.4 mu m using the HNT-1P and 550.6 +/- 26.3 mu m using USP, and the correlation coefficient between the two measurements was 0.913 (p < 0.001). The average cIOP was 14.6 +/- 5.0 and 13.4 +/- 6.5 mmHg according to the HNT-1P and GAT, respectively, and the correlation coefficient between the two corrected measurements was 0.956 (p < 0.001). Bland-Altman plots showed a high degree of agreement between the HNT-1P measurements and those of the other devices. Conclusions: The new non-contact HNT-1P tonometer provided reliable IOP, CCT, and cIOP measurements when compared with GAT and USP. HNT-1P could therefore be used clinically for reliable and accurate measurements of the IOP, CCT, and cIOP.
引用
收藏
页码:524 / 531
页数:8
相关论文
共 50 条
  • [1] Comparability analysis of the HNT-1P Huvitz non contact tonometer for the measurement of intraocular pressure
    Montoya, Alicia Carceles
    Lucas, Carlos Enrique Monera
    Jaimes, Claudia Patricia Tarazona
    Valero, Daniel Romero
    Martinez, Alejandro Moya
    Toldos, Jose Juan Martinez
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2024,
  • [2] Assessment of intraocular pressure measurement between Goldman applanation tonometer, rebound tonometer, non-contact tonometer, and its correlation with central corneal thickness
    Porwal, Amit C.
    Shrishrimal, Meghna
    Punamia, Ravin P.
    Mathew, Bennet Chacko
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2023, 71 (05) : 1927 - 1931
  • [3] The applicability of correction factor for corneal thickness on non-contact tonometer measured intraocular pressure in LASIK treated eyes
    Jethani, Jitendra
    Dave, Paaraj
    Jethani, Monica
    Desai, Yogesh
    Patel, Purvi
    SAUDI JOURNAL OF OPHTHALMOLOGY, 2016, 30 (01) : 25 - 28
  • [4] On Imaging based Non-contact Tonometer for Intraocular Pressure Measurement
    Bhatt, Mahabaleswara Ram
    Rao, Shyam Vasudeva
    2013 IEEE POINT-OF-CARE HEALTHCARE TECHNOLOGIES (PHT), 2013, : 97 - 100
  • [5] Relationship of Central Corneal Thickness and Intraocular Pressure by iCare Rebound Tonometer
    Rao, Aparna
    Kumar, Mukesh
    Prakash, B.
    Varshney, Gopal
    JOURNAL OF GLAUCOMA, 2014, 23 (06) : 380 - 384
  • [6] Corrected Intraocular Pressure Variability with Central Corneal Thickness Measurement
    Ahmed, Mohamed Attia Ali
    Abdelhalim, Ahmed Shawkat
    CLINICAL OPHTHALMOLOGY, 2020, 14 : 4501 - 4506
  • [7] COMPARATIVE MEASUREMENTS OF INTRAOCULAR-PRESSURE WITH THE NON-CONTACT TONOMETER AND THE GOLDMANN APPLANATION TONOMETER
    LANGMANN, G
    SCHUHMANN, G
    SCHWAIGER, W
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1985, 187 (05) : 345 - 347
  • [8] Intraocular pressure and central corneal thickness
    Manni, Gianluca
    Oddone, Francesco
    Parisi, Vincenzo
    Tosto, Adriana
    Centofanti, Marco
    GLAUCOMA: AN OPEN WINDOW TO NEURODEGENERATION AND NEUROPROTECTION, 2008, 173 : 25 - 30
  • [9] Effect of central corneal thickness on intraocular pressure with the rebound tonometer and the applanation tonometer in normal dogs
    Park, Young-Woo
    Jeong, Man-Bok
    Kim, Tae-Hyun
    Ahn, Jae-Sang
    Ahn, Jeong-Taek
    Park, Shin-Ae
    Kim, Se-Eun
    Seo, Kangmoon
    VETERINARY OPHTHALMOLOGY, 2011, 14 (03) : 169 - 173
  • [10] Intraocular pressure and corneal thickness - A comparison between non-contact tonometry and applanation tonometry
    Domke, N.
    Hager, A.
    Wiegand, W.
    OPHTHALMOLOGE, 2006, 103 (07): : 583 - 587