Impact of iterative reconstructions on image noise and low-contrast object detection in low kVp simulated abdominal CT: a phantom study

被引:18
|
作者
Holmquist, Fredrik [1 ]
Nyman, Ulf [2 ]
Siemund, Roger [1 ]
Geijer, Mats [1 ]
Soederberg, Marcus [3 ]
机构
[1] Lund Univ, Skane Univ Hosp, Dept Med Imaging & Physiol, Lund, Sweden
[2] Lund Univ, Skane Univ Hosp, Div Med Radiol, Dept Translat Med, Malmo, Sweden
[3] Lund Univ, Med Radiat Phys, Dept Translat Med, Skane Univ Hosp, Malmo, Sweden
关键词
Computed tomography (CT); filtered back-projection; image noise; iterative reconstruction; low tube voltage; noise reduction; radiation; LOW TUBE VOLTAGE; COMPUTED-TOMOGRAPHY; OBSERVER PERFORMANCE; DOSE REDUCTION; PATIENT SIZE; ROW CT; RADIATION; QUALITY; LIVER; PARAMETERS;
D O I
10.1177/0284185115617347
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background Low kilovoltage (kVp) computed tomography (CT) may be used to reduce contrast medium dose in patients at risk of contrast nephropathy, at the cost of increased image noise. Purpose To evaluate: (i) the impact of iterative reconstructions (Siemens SAFIRE) on low-contrast object detection to compensate for increased noise instead of increased tube loading when decreasing tube potential; and (ii) the change in iodine attenuation in simulated abdominal CT. Material and Methods A phantom was scanned at 70, 80, 100, and 120kVp at fixed effective tube loading (170mAs(EFF)) and fixed radiation dose (CTDIVOL 10mGy). Images were reconstructed with filtered back-projection (FBP) and SAFIRE strengths S1-S5. Iodine attenuation, objective image noise, contrast-to-noise ratio (CNR), noise power spectrum (NPS), spatial resolution, and subjective detectability of low-contrast objects were evaluated. Results Compared with 120kVp iodine attenuation increased by a factor 1.6 and 2.0, and image noise increased by a factor 1.9 and 2.5 at 80 and 70kVp, respectively. Compared with FBP, SAFIRE showed objective reduction in image noise and increased CNR without loss of spatial resolution or any significant NPS alteration, with general tendency to improve subjective detectability of low-contrast objects. At 170mAs(EFF) the number of discernible 1.0% contrast objects at 70kVp/S5 and 80kVp/S5 was similar to that at 120kVp/FBP. Conclusion With the SAFIRE algorithm image noise, CNR and detectability of low-contrast objects may be kept unchanged without increased tube loading when using lowkVp settings to reduce contrast medium dose in azotemic patients.
引用
收藏
页码:1079 / 1088
页数:10
相关论文
共 50 条
  • [1] Feasibility of low-radiation-dose CT for abdominal examinations with hybrid iterative reconstruction algorithm: Low-contrast phantom study
    Kondo M.
    Hatakenaka M.
    Higuchi K.
    Fujioka T.
    Shirasaka T.
    Nakamura Y.
    Nakamura K.
    Yoshiura T.
    Honda H.
    Radiological Physics and Technology, 2013, 6 (2) : 287 - 292
  • [2] Can image-domain filtering of FBP CT reconstructions match low-contrast performance of iterative reconstructions?
    Divel, Sarah E.
    Hsieh, Scott S.
    Wang, Jia
    Pelc, Norbert J.
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [3] Impact of Incremental Increase in CT Image Noise on Detection of Low-contrast Hypodense Liver Lesions
    Kanal, Kalpana M.
    Chung, Jonathan H.
    Wang, Jin
    Bhargava, Puneet
    Gunn, Martin L.
    Shuman, William P.
    Stewart, Brent K.
    ACADEMIC RADIOLOGY, 2014, 21 (10) : 1233 - 1239
  • [4] The effects of iterative reconstruction in CT on low-contrast liver lesion volumetry: a phantom study
    Li, Qin
    Berman, Benjamin P.
    Schumacher, Justin
    Liang, Yongguang
    Gavrielides, Marios A.
    Yang, Hao
    Zhao, Binsheng
    Petrick, Nicholas
    MEDICAL IMAGING 2017: COMPUTER-AIDED DIAGNOSIS, 2017, 10134
  • [5] A statistical-based automatic detection of a low-contrast object in the ACR CT phantom for measuring contrast-to-noise ratio of CT images
    Anam, Choirul
    Amilia, Riska
    Naufal, Ariij
    Fujibuchi, Toshioh
    Dougherty, Geoff
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2025, 11 (01):
  • [6] Effect of Low Tube Voltage on Image Quality, Radiation Dose, and Low-Contrast Detectability at Abdominal Multidetector CT: Phantom Study
    Tang, Kun
    Wang, Ling
    Li, Rui
    Lin, Jie
    Zheng, Xiangwu
    Cao, Guoquan
    JOURNAL OF BIOMEDICINE AND BIOTECHNOLOGY, 2012,
  • [7] Evaluation of low-contrast detectability for iterative reconstruction in pediatric abdominal computed tomography: a phantom study
    Nicholas Rubert
    Richard Southard
    Susan M. Hamman
    Ryan Robison
    Pediatric Radiology, 2020, 50 : 345 - 356
  • [8] Evaluation of low-contrast detectability for iterative reconstruction in pediatric abdominal computed tomography: a phantom study
    Rubert, Nicholas
    Southard, Richard
    Hamman, Susan M.
    Robison, Ryan
    PEDIATRIC RADIOLOGY, 2020, 50 (03) : 345 - 356
  • [9] The Quantitative Effect of Noise and Object Diameter on Low-Contrast Detectability of AAPM CT Performance Phantom Images
    Setiawati, E.
    Anam, C.
    Widyasari, W.
    Dougherty, G.
    ATOM INDONESIA, 2023, 49 (01) : 61 - 66
  • [10] Hybrid Iterative Reconstruction Technique for Abdominal CT Protocols in Obese Patients: Assessment of Image Quality, Radiation Dose, and Low-Contrast Detectability in a Phantom
    Schindera, Sebastian T.
    Odedra, Devang
    Mercer, Diego
    Thipphavong, Seng
    Chou, Paul
    Szucs-Farkas, Zsolt
    Rogalla, Patrik
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 202 (02) : W146 - W152