Impact of iterative reconstructions on image noise and low-contrast object detection in low kVp simulated abdominal CT: a phantom study

被引:18
|
作者
Holmquist, Fredrik [1 ]
Nyman, Ulf [2 ]
Siemund, Roger [1 ]
Geijer, Mats [1 ]
Soederberg, Marcus [3 ]
机构
[1] Lund Univ, Skane Univ Hosp, Dept Med Imaging & Physiol, Lund, Sweden
[2] Lund Univ, Skane Univ Hosp, Div Med Radiol, Dept Translat Med, Malmo, Sweden
[3] Lund Univ, Med Radiat Phys, Dept Translat Med, Skane Univ Hosp, Malmo, Sweden
关键词
Computed tomography (CT); filtered back-projection; image noise; iterative reconstruction; low tube voltage; noise reduction; radiation; LOW TUBE VOLTAGE; COMPUTED-TOMOGRAPHY; OBSERVER PERFORMANCE; DOSE REDUCTION; PATIENT SIZE; ROW CT; RADIATION; QUALITY; LIVER; PARAMETERS;
D O I
10.1177/0284185115617347
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background Low kilovoltage (kVp) computed tomography (CT) may be used to reduce contrast medium dose in patients at risk of contrast nephropathy, at the cost of increased image noise. Purpose To evaluate: (i) the impact of iterative reconstructions (Siemens SAFIRE) on low-contrast object detection to compensate for increased noise instead of increased tube loading when decreasing tube potential; and (ii) the change in iodine attenuation in simulated abdominal CT. Material and Methods A phantom was scanned at 70, 80, 100, and 120kVp at fixed effective tube loading (170mAs(EFF)) and fixed radiation dose (CTDIVOL 10mGy). Images were reconstructed with filtered back-projection (FBP) and SAFIRE strengths S1-S5. Iodine attenuation, objective image noise, contrast-to-noise ratio (CNR), noise power spectrum (NPS), spatial resolution, and subjective detectability of low-contrast objects were evaluated. Results Compared with 120kVp iodine attenuation increased by a factor 1.6 and 2.0, and image noise increased by a factor 1.9 and 2.5 at 80 and 70kVp, respectively. Compared with FBP, SAFIRE showed objective reduction in image noise and increased CNR without loss of spatial resolution or any significant NPS alteration, with general tendency to improve subjective detectability of low-contrast objects. At 170mAs(EFF) the number of discernible 1.0% contrast objects at 70kVp/S5 and 80kVp/S5 was similar to that at 120kVp/FBP. Conclusion With the SAFIRE algorithm image noise, CNR and detectability of low-contrast objects may be kept unchanged without increased tube loading when using lowkVp settings to reduce contrast medium dose in azotemic patients.
引用
收藏
页码:1079 / 1088
页数:10
相关论文
共 50 条
  • [31] Low radiation and low-contrast dose pulmonary CT angiography: Comparison of 80 kVp/60 ml and 100 kVp/80 ml protocols
    Viteri-Ramirez, G.
    Garcia-Lallana, A.
    Simon-Yarza, I.
    Broncano, J.
    Ferreira, M.
    Pueyo, J. C.
    Villanueva, A.
    Bastarrika, G.
    CLINICAL RADIOLOGY, 2012, 67 (09) : 833 - 839
  • [32] Low-contrast detectability in volume rendering: A phantom study on multidetector-spiral-CT data
    Shin, H
    Von Falck, C
    Galanski, MW
    RADIOLOGY, 2002, 225 : 451 - 452
  • [33] Effects of different levels of CT iterative reconstruction on low-contrast detectability and radiation dose in patients of different sizes: an anthropomorphic phantom study
    Rampado, Osvaldo
    Depaoli, Alessandro
    Marchisio, Filippo
    Gatti, Marco
    Racine, Damien
    Ruggeri, Valeria
    Ruggirello, Irene
    Darvizeh, Fatemeh
    Fonio, Paolo
    Ropolo, Roberto
    RADIOLOGIA MEDICA, 2021, 126 (01): : 55 - 62
  • [34] Effects of different levels of CT iterative reconstruction on low-contrast detectability and radiation dose in patients of different sizes: an anthropomorphic phantom study
    Osvaldo Rampado
    Alessandro Depaoli
    Filippo Marchisio
    Marco Gatti
    Damien Racine
    Valeria Ruggeri
    Irene Ruggirello
    Fatemeh Darvizeh
    Paolo Fonio
    Roberto Ropolo
    La radiologia medica, 2021, 126 : 55 - 62
  • [35] DETECTION OF LOW-CONTRAST IMAGES IN FILM-GRAIN NOISE
    NADERI, F
    SAWCHUK, AA
    APPLIED OPTICS, 1978, 17 (18): : 2883 - 2891
  • [36] Low-contrast detectability in volume rendering: a phantom study on multidector-row spiral CT data
    Shin, HO
    Falck, CV
    Galanski, M
    EUROPEAN RADIOLOGY, 2004, 14 (02) : 341 - 349
  • [37] Low-contrast detectability in volume rendering: a phantom study on multidector-row spiral CT data
    Hoen-Oh Shin
    Christian V. Falck
    Michael Galanski
    European Radiology, 2004, 14 : 341 - 349
  • [38] Helical CT of an upper abdomen phantom: Signal-to-noise ratio and detection rate of low-contrast liver lesions with different scan parameters
    Kopka, L
    Hermann, K
    Funke, M
    Grabbe, EH
    RADIOLOGY, 1996, 201 : 340 - 340
  • [39] Detection of low-contrast objects: Experimental comparison of single- and multi-detector row CT with a phantom
    Verdun, FR
    Denys, A
    Valley, JF
    Schnyder, P
    Meuli, RA
    RADIOLOGY, 2002, 223 (02) : 426 - 431
  • [40] Low contrast detectability performance of model observers based on CT phantom images: kVp influence
    Hernandez-Giron, I.
    Calzado, A.
    Geleijns, J.
    Joemai, R. M. S.
    Veldkamp, W. J. H.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (07): : 798 - 807