Impact of iterative reconstructions on image noise and low-contrast object detection in low kVp simulated abdominal CT: a phantom study

被引:18
|
作者
Holmquist, Fredrik [1 ]
Nyman, Ulf [2 ]
Siemund, Roger [1 ]
Geijer, Mats [1 ]
Soederberg, Marcus [3 ]
机构
[1] Lund Univ, Skane Univ Hosp, Dept Med Imaging & Physiol, Lund, Sweden
[2] Lund Univ, Skane Univ Hosp, Div Med Radiol, Dept Translat Med, Malmo, Sweden
[3] Lund Univ, Med Radiat Phys, Dept Translat Med, Skane Univ Hosp, Malmo, Sweden
关键词
Computed tomography (CT); filtered back-projection; image noise; iterative reconstruction; low tube voltage; noise reduction; radiation; LOW TUBE VOLTAGE; COMPUTED-TOMOGRAPHY; OBSERVER PERFORMANCE; DOSE REDUCTION; PATIENT SIZE; ROW CT; RADIATION; QUALITY; LIVER; PARAMETERS;
D O I
10.1177/0284185115617347
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background Low kilovoltage (kVp) computed tomography (CT) may be used to reduce contrast medium dose in patients at risk of contrast nephropathy, at the cost of increased image noise. Purpose To evaluate: (i) the impact of iterative reconstructions (Siemens SAFIRE) on low-contrast object detection to compensate for increased noise instead of increased tube loading when decreasing tube potential; and (ii) the change in iodine attenuation in simulated abdominal CT. Material and Methods A phantom was scanned at 70, 80, 100, and 120kVp at fixed effective tube loading (170mAs(EFF)) and fixed radiation dose (CTDIVOL 10mGy). Images were reconstructed with filtered back-projection (FBP) and SAFIRE strengths S1-S5. Iodine attenuation, objective image noise, contrast-to-noise ratio (CNR), noise power spectrum (NPS), spatial resolution, and subjective detectability of low-contrast objects were evaluated. Results Compared with 120kVp iodine attenuation increased by a factor 1.6 and 2.0, and image noise increased by a factor 1.9 and 2.5 at 80 and 70kVp, respectively. Compared with FBP, SAFIRE showed objective reduction in image noise and increased CNR without loss of spatial resolution or any significant NPS alteration, with general tendency to improve subjective detectability of low-contrast objects. At 170mAs(EFF) the number of discernible 1.0% contrast objects at 70kVp/S5 and 80kVp/S5 was similar to that at 120kVp/FBP. Conclusion With the SAFIRE algorithm image noise, CNR and detectability of low-contrast objects may be kept unchanged without increased tube loading when using lowkVp settings to reduce contrast medium dose in azotemic patients.
引用
收藏
页码:1079 / 1088
页数:10
相关论文
共 50 条
  • [21] Physical and visual evaluations of CT image quality of large low-contrast objects with visual model-based iterative reconstruction technique: a phantom study
    Hideki Shibata
    Kosuke Matsubara
    Yasuki Asada
    Akihiro Takemura
    Isao Kozawa
    Physical and Engineering Sciences in Medicine, 2023, 46 : 141 - 150
  • [22] Iterative reconstruction with multifrequency signal recognition technology to improve low-contrast detectability: A phantom study
    Funama, Yoshinori
    Shirasaka, Takashi
    Goto, Taiga
    Aoki, Yuko
    Tanaka, Kana
    Yoshida, Ryo
    ACTA RADIOLOGICA OPEN, 2022, 11 (06)
  • [23] Radiation dose reduction without degradation of low-contrast detectability at abdominal multisection CT with a low-tube voltage technique: Phantom study
    Funama, Y
    Awai, K
    Nakayama, Y
    Kakei, K
    Nagasue, N
    Shimamura, M
    Sato, N
    Sultana, S
    Morishita, S
    Yamashita, Y
    RADIOLOGY, 2005, 237 (03) : 905 - 910
  • [24] Low-contrast lesion detection in neck CT: a multireader study comparing deep learning, iterative, and filtered back projection reconstructions using realistic phantoms
    Bellmann, Quirin
    Peng, Yang
    Genske, Ulrich
    Yan, Li
    Wagner, Moritz
    Jahnke, Paul
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2024, 8 (01)
  • [25] Effects of various generations of iterative CT reconstruction algorithms on low-contrast detectability as a function of the effective abdominal diameter: A quantitative task-based phantom study
    Viry, Anais
    Aberle, Christoph
    Racine, Damien
    Knebel, Jean-Francois
    Schindera, Sebastian T.
    Schmidt, Sabine
    Becce, Fabio
    Verdun, Francis R.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 48 : 111 - 118
  • [26] CT Detectability of Small Low-Contrast Hypoattenuating Focal Lesions: Iterative Reconstructions versus Filtered Back Projection
    Mileto, Achille
    Zamora, David A.
    Alessio, Adam M.
    Pereira, Carina
    Liu, Jin
    Bhargava, Puneet
    Carnell, Jonathan
    Cowan, Sophie M.
    Dighe, Manjiri K.
    Gunn, Martin L.
    Kim, Sooah
    Kolokythas, Orpheus
    Lee, Jean H.
    Maki, Jeffery H.
    Moshiri, Mariam
    Nasrullah, Ayesha
    O'Malley, Ryan B.
    Schmiedl, Udo P.
    Soloff, Erik V.
    Toia, Giuseppe V.
    Wang, Carolyn L.
    Kanal, Kalpana M.
    RADIOLOGY, 2018, 289 (02) : 443 - 454
  • [28] Impact of 80 kVp with iterative reconstruction algorithm and low-dose contrast medium on the image quality of craniocervical CT angiography
    Chen, Po-An
    Chen, Chih-Wei
    Chou, Chiung-Chen
    Fu, Jui-Hsun
    Wang, Po-Chin
    Hsu, Shuo-Hsiu
    Lai, Ping-Hong
    CLINICAL IMAGING, 2020, 68 (68) : 124 - 130
  • [29] Method for determining slice sensitivity profile of iterative reconstruction CT images using low-contrast sphere phantom
    Akihiro Narita
    Masaki Ohkubo
    Takahiro Fukaya
    Yoshiyuki Noto
    Radiological Physics and Technology, 2021, 14 : 366 - 372
  • [30] Method for determining slice sensitivity profile of iterative reconstruction CT images using low-contrast sphere phantom
    Narita, Akihiro
    Ohkubo, Masaki
    Fukaya, Takahiro
    Noto, Yoshiyuki
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2021, 14 (04) : 366 - 372