Study of ion transmission for a linear mode Bradbury-Nielsen gate in ion mobility spectrometer

被引:5
|
作者
Ni, Kai [1 ]
Guo, Jingran [1 ]
Ou, Guangli [1 ]
Zhang, Xiaoguo [1 ]
Yu, Quan [1 ]
Qian, Xiang [1 ]
Wang, Xiaohao [1 ,2 ]
机构
[1] Tsinghua Univ, Grad Sch Shenzhen, Div Adv Mfg, Shenzhen 518055, Peoples R China
[2] Tsinghua Univ, State Key Lab Precis Measure Technol & Instrument, Beijing 100084, Peoples R China
基金
中国国家自然科学基金;
关键词
Ion mobility spectrometer; Bradbury-Nielsen gate; Ion transmission; Quantitatively control; Fourier transform; IONIZATION; RESOLUTION; SIMION;
D O I
10.1016/j.ijms.2014.12.012
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
Fourier transform ion mobility spectrometer (FT-IMS) has the advantage of high ion utilization efficiency. Its two ion gates work in binary switch mode with square wave modulation function, which causes interference from high order scaled versions of the original spectrum. This problem can be avoided by using cosine modulation function, which requires the ion gates working in linear mode by quantitatively controlling its transmission in real time. In this paper, the influence of the structure parameters of Bradbury-Nielsen gate (BNG) and the electrical parameters on the ion transmission is investigated by SIMION simulation and experiment. The results show that the transmission in the open state is determined by the mechanical aperture ratio, and in the transition state is determined by the electric field ratio. In practical, the ion transmission can be quantitatively controlled by simply adjusting the voltage difference between adjacent wires when in the transition state. The nonlinearity between the transmission and the control voltage can be corrected by a simple calibration. This job proves the possibility of implementing a linear mode BNG, which can be used in further to improve the precision of the FT-IMS. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:75 / 79
页数:5
相关论文
共 50 条
  • [21] A tandem ion trap/ion mobility spectrometer
    Creaser, CS
    Benyezzar, M
    Griffiths, JR
    Stygall, JW
    ANALYTICAL CHEMISTRY, 2000, 72 (13) : 2724 - 2729
  • [22] Expanded theory for the resolving power of a linear ion mobility spectrometer
    Spangler, GE
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2002, 220 (03) : 399 - 418
  • [23] Transmission mode Ion/Ion electron-transfer dissociation in a linear ion trap
    Liang, Xiaorong
    Hager, James W.
    McLuckey, Scott A.
    ANALYTICAL CHEMISTRY, 2007, 79 (09) : 3363 - 3370
  • [24] Ion mobility spectrometer-field asymmetric ion mobility spectrometer-mass spectrometry
    Pollard, Matthew J.
    Hilton, Christopher K.
    Li, Hongli
    Kaplan, Kimberly
    Yost, Richard A.
    Hill, Herbert H., Jr.
    INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2011, 14 (01) : 15 - 22
  • [25] Using a Cyclic Ion Mobility Spectrometer for Tandem Ion Mobility Experiments
    Ollivier, Simon
    Fanuel, Mathieu
    Rogniaux, Helene
    Ropartz, David
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2022, (179):
  • [26] Dual Mode Ion Mobility Spectrometer High Voltage Formation Circuit
    Gromov, E.
    Matusko, M.
    Shaltaeva, Y.
    Pershenkov, V.
    Belyakov, V.
    Golovin, A.
    Malkin, E.
    Ivanov, I.
    Vasilyev, V.
    2019 IEEE 31ST INTERNATIONAL CONFERENCE ON MICROELECTRONICS (MIEL 2019), 2019, : 329 - 332
  • [27] Increased ion transmission in IMS: A high resolution, periodic-focusing DC ion guide ion mobility spectrometer
    Blase, Ryan C.
    Silveira, Joshua A.
    Gillig, Kent J.
    Gamage, Chaminda M.
    Russell, David H.
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2011, 301 (1-3) : 166 - 173
  • [28] A laser ion-mobility spectrometer
    G. E. Kotkovskii
    I. L. Martynov
    V. V. Novikova
    A. A. Chistyakov
    Instruments and Experimental Techniques, 2009, 52 : 253 - 259
  • [29] MINIATURE ION MOBILITY SPECTROMETER CELL
    SPANGLER, GE
    VORA, KN
    CARRICO, JP
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1986, 19 (03): : 191 - 198
  • [30] A laser ion-mobility spectrometer
    Kotkovskii, G. E.
    Martynov, I. L.
    Novikova, V. V.
    Chistyakov, A. A.
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2009, 52 (02) : 253 - 259