Study of ion transmission for a linear mode Bradbury-Nielsen gate in ion mobility spectrometer

被引:5
|
作者
Ni, Kai [1 ]
Guo, Jingran [1 ]
Ou, Guangli [1 ]
Zhang, Xiaoguo [1 ]
Yu, Quan [1 ]
Qian, Xiang [1 ]
Wang, Xiaohao [1 ,2 ]
机构
[1] Tsinghua Univ, Grad Sch Shenzhen, Div Adv Mfg, Shenzhen 518055, Peoples R China
[2] Tsinghua Univ, State Key Lab Precis Measure Technol & Instrument, Beijing 100084, Peoples R China
基金
中国国家自然科学基金;
关键词
Ion mobility spectrometer; Bradbury-Nielsen gate; Ion transmission; Quantitatively control; Fourier transform; IONIZATION; RESOLUTION; SIMION;
D O I
10.1016/j.ijms.2014.12.012
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
Fourier transform ion mobility spectrometer (FT-IMS) has the advantage of high ion utilization efficiency. Its two ion gates work in binary switch mode with square wave modulation function, which causes interference from high order scaled versions of the original spectrum. This problem can be avoided by using cosine modulation function, which requires the ion gates working in linear mode by quantitatively controlling its transmission in real time. In this paper, the influence of the structure parameters of Bradbury-Nielsen gate (BNG) and the electrical parameters on the ion transmission is investigated by SIMION simulation and experiment. The results show that the transmission in the open state is determined by the mechanical aperture ratio, and in the transition state is determined by the electric field ratio. In practical, the ion transmission can be quantitatively controlled by simply adjusting the voltage difference between adjacent wires when in the transition state. The nonlinearity between the transmission and the control voltage can be corrected by a simple calibration. This job proves the possibility of implementing a linear mode BNG, which can be used in further to improve the precision of the FT-IMS. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:75 / 79
页数:5
相关论文
共 50 条
  • [1] Modeling the modulation characteristics of the Bradbury-Nielsen gate in ion mobility spectrometers
    Guo, Kaitai
    Zhang, Chong
    Ni, Kai
    Wang, Xiaohao
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2022, 93 (08):
  • [2] Field Switching Combined with Bradbury-Nielsen Gate for Ion Mobility Spectrometry
    Chen, Chuang
    Tabrizchi, Mahmoud
    Wang, Weiguo
    Li, Haiyang
    ANALYTICAL CHEMISTRY, 2015, 87 (15) : 7925 - 7930
  • [3] Bradbury-Nielsen gate electrode potential switching modes optimizing the ion packet time width in an ion mobility spectrometer
    Kurnin, Igor V.
    Krasnov, Nikolai V.
    Semenov, Sergey Y.
    Smirnov, Valentin N.
    INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2014, 17 (02) : 79 - 85
  • [4] Improving Performance of Ion Mobility Spectrometry Using Bradbury-Nielsen Ion Gate Pulse Waveform Modulation
    Cang Huai-Wen
    Li Hang
    Huang Wei
    Zhang Yuan-Zhi
    Li Jing-Hua
    Wang Wei-Guo
    Li Hai-Yang
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2021, 49 (12) : 2067 - 2074
  • [5] Resolution Enhancement of Ion Mobility Spectrometry by Improving the Three-Zone Properties of the Bradbury-Nielsen Gate
    Du, Yongzhai
    Wang, Weiguo
    Li, Haiyang
    ANALYTICAL CHEMISTRY, 2012, 84 (03) : 1725 - 1731
  • [6] Note: Design and construction of a simple and reliable printed circuit board-substrate Bradbury-Nielsen gate for ion mobility spectrometry
    Du, Yongzhai
    Cang, Huaiwen
    Wang, Weiguo
    Han, Fenglei
    Chen, Chuang
    Li, Lin
    Hou, Keyong
    Li, Haiyang
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2011, 82 (08):
  • [7] Enhancing ion mobility spectrometry performance through a programmable ion swarm shaping method based on Bradbury-Nielsen gates
    Wang, Jiyao
    Chen, Hai
    Ling, Yiyuan
    Zhang, Chong
    Zhou, Haobo
    Wang, Xiaohao
    Ni, Kai
    TALANTA, 2024, 269
  • [8] Simple high-resolution Bradbury-Nielsen mass gate
    Jia, Zezhao
    Liu, Zhaojun
    Wang, Xiaohan
    Wang, Runyu
    Ma, Tianjie
    Zhou, Ziwen
    Moro, Ramiro
    von Issendorff, Bernd
    Ma, Lei
    AIP ADVANCES, 2024, 14 (03)
  • [9] Bradbury-Nielsen vs. Field switching shutters for high resolution drift tube ion mobility spectrometers
    Kirk, Ansgar T.
    Zimmermann, Stefan
    INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2014, 17 (3-4) : 131 - 137
  • [10] A simple template-based transfer method to fabricate Bradbury-Nielsen gates with uniform tension for ion mobility spectrometry
    Ni Kai
    Guo Jingran
    Ou Guangli
    Lei Yu
    Yu Quan
    Qian Xiang
    Wang Xiaohao
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2014, 85 (08):