Analysis of metal elements by solution cathode glow discharge-atomic emission spectrometry with a modified pulsation damper

被引:29
|
作者
Wang, Jinmei [1 ]
Tang, Pengfei [1 ]
Zheng, Peichao [1 ]
Zhai, Xiang [1 ]
机构
[1] Chongqing Univ Posts & Telecommun, Coll Optoelect Engn, Chongqing Municipal Level Key Lab Photoelect Info, Chongqing 400065, Peoples R China
关键词
FLOWING ELECTROLYTE-SOLUTIONS; APGD IONIZATION SOURCE; ATMOSPHERIC-PRESSURE; LIQUID CATHODE; ONLINE SEPARATION; ANALYTICAL PERFORMANCE; MASS-SPECTROMETRY; CONTACT; SAMPLES; PRECONCENTRATION;
D O I
10.1039/c7ja00212b
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Solution-cathode glow discharge-atomic emission spectrometry (SCGD-AES) with a portable spectrometer was used to detect metal elements in solution. A modified pulsation damper was designed and optimized to improve discharge plasma stability. The effects of pulsation damper parameters as well as the discharge current and solution flow rate were investigated. The plasma stability of Na, K and Li elements under optimized parameters with the modified pulsation damper was discussed and they showed good stability over a long time. The standard deviation (SD) of background intensity was low, and the detection limits for Zn, Mg, Ag, Pb, Sr, Na, Li, K, Rb and Cs were 72, 2.1, 1.2, 35, 86, 0.04, 0.16, 0.14, 0.38 and 1.1 mu g L-1, respectively. Ag, Zn and Pb elements in a certified reference material were quantitatively analyzed, and the measurement results obtained by SCGD-AES agreed well with the reference values. Specifically, the precision (relative standard deviation of 10 measured values) of detection was lower than 1.85%. The results show that the SCGD-AES system with the modified pulsation damper can accurately and stably detect the content of metal elements in water.
引用
下载
收藏
页码:1925 / 1931
页数:7
相关论文
共 50 条
  • [21] Rapid determination of indium in water samples using a portable solution cathode glow discharge-atomic emission spectrometer
    Zu, Wenchuan
    Yang, Yin
    Wang, Yu
    Yang, Xiaotao
    Liu, Cong
    Ren, Min
    MICROCHEMICAL JOURNAL, 2018, 137 : 266 - 271
  • [22] Evaluation of solution-cathode glow discharge atomic emission spectrometry for the analysis of nanoparticle containing solutions
    Hazel, Nicholas
    Orejas, Jaime
    Ray, Steven J.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2021, 176 (176)
  • [23] Ultratrace Determination of Tin, Germanium, and Selenium by Hydride Generation Coupled with a Novel Solution-Cathode Glow Discharge-Atomic Emission Spectrometry Method
    Huang, Chuchu
    Li, Qing
    Mo, Jiamei
    Wang, Zheng
    ANALYTICAL CHEMISTRY, 2016, 88 (23) : 11559 - 11567
  • [24] Fundamental studies of electrolyte-as-cathode glow discharge-atomic emission spectrometry for the determination of trace metals in flowing water
    Park, YS
    Ku, SH
    Hong, SH
    Kim, HJ
    Piepmeier, EH
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1998, 53 (6-8) : 1167 - 1179
  • [26] Design modification of a solution-cathode glow discharge-atomic emission spectrometer for the determination of trace metals in titanium dioxide
    Wang, Zheng
    Gai, Rongyin
    Zhou, Lei
    Zhang, Zhen
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2014, 29 (11) : 2042 - 2049
  • [27] Solution Cathode Glow Discharge-Atomic Emission Spectroscopy Coupled With Hydride Generation for Detecting Trace Mercury and Tin in Water
    Zheng Pei-chao
    Liu Ran-ning
    Wang Jin-mei
    Feng Chu-hui
    He Yu-tong
    Wu Mei-ni
    He Yu-xin
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42 (04) : 1139 - 1143
  • [28] Determination of trace heavy metals in environmental and biological samples by solution cathode glow discharge-atomic emission spectrometry and addition of ionic surfactants for improved sensitivity
    Zhang, Zhen
    Wang, Zheng
    Li, Qing
    Zou, Huijun
    Shi, Ying
    TALANTA, 2014, 119 : 613 - 619
  • [29] Determination of gallium in water samples by atomic emission spectrometry based on solution cathode glow discharge
    Li, Hongwei
    Zu, Wenchuan
    Liu, Fengkui
    Wang, Yu
    Yang, Yin
    Yang, Xiaotao
    Liu, Cong
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2019, 152 : 25 - 29
  • [30] Determination of gallium and indium by solution cathode glow discharge as an excitation source for atomic emission spectrometry
    Yu, Jie
    Kang, Yuejing
    Lu, Quanfang
    Luo, Hui
    Cui, Lijuan
    Lu, Zhuanhong
    Li, Juanlong
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2020, 172