Continuous-Flow Cardiac Assistance: Effects on Aortic Valve Function in a Mock Loop

被引:29
|
作者
Tuzun, Egemen [1 ,2 ,3 ,4 ]
Rutten, Marcel [3 ]
Dat, Marco [3 ]
van de Vosse, Frans [3 ]
Kadipasaoglu, Cihan [1 ,2 ]
de Mol, Bas [3 ,4 ]
机构
[1] St Lukes Episcopal Hosp, Texas Heart Inst, Cardiovasc Res Labs, Houston, TX 77030 USA
[2] St Lukes Episcopal Hosp, Ctr Cardiac Support, Houston, TX 77030 USA
[3] Eindhoven Univ Technol, Dept Biomed Engn, NL-5600 MB Eindhoven, Netherlands
[4] Univ Amsterdam, Acad Med Ctr, Dept Cardiovasc Surg, NL-1105 AZ Amsterdam, Netherlands
关键词
left ventricular assist device; continuous flow; aortic valve; mock circulation; LEAFLETS;
D O I
10.1016/j.jss.2010.05.040
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background. As the use of left ventricular assist devices (LVADs) to treat end-stage heart failure has become more widespread, leaflet fusion-with resultant aortic regurgitation-has been observed more frequently. To quantitatively assess the effects of nonpulsatile flow on aortic valve function, we tested a continuous-flow LVAD in a mock circulatory system (MCS) with an interposed valve. Materials and Methods. To mimic the hemodynamic characteristics of LVAD patients, we utilized an MCS in which a Jarvik 2000 LVAD was positioned at the base of a servomotor-operated piston pump (left ventricular chamber). We operated the LVAD at 8000 to 12,000 rpm, changing the speed in 1000-rpm increments. At each speed, we first varied the outflow resistance at a constant stroke volume, then varied the stroke volume at a constant outflow resistance. We measured the left ventricular pressure, aortic pressure, pump flow, and total flow, and used these values to compute the change, if any, in the aortic duty cycle (aortic valve open time) and transvalvular aortic pressure loads. Results. Validation of the MCS was demonstrated by the simulation of physiologic pressure and flow waveforms. At increasing LVAD speeds, the mean aortic pressure load steadily increased, while the aortic duty cycle steadily decreased. Changes were consistent for each MCS experimental setting, despite variations in stroke volume and outflow resistance. Conclusions. Increased LVAD flow results in an impaired aortic valve-open time due to a pressure overload above the aortic valve. Such an overload may initiate structural changes, causing aortic leaflet fusion and/or regurgitation. (C) 2011 Elsevier Inc. All rights reserved.
引用
收藏
页码:443 / 447
页数:5
相关论文
共 50 条
  • [31] Concomitant Aortic Valve Procedures in Patients Undergoing Implantation of Continuous-Flow LVADs: An INTERMACS Database Analysis
    Robertson, J. O.
    Meyers, S. L.
    Kirklin, J. K.
    Mertz, G. D.
    Prasad, S.
    Itoh, A.
    Silvestry, S. C.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2014, 33 (04): : S9 - S9
  • [32] A Novel Approach to the In Vitro Hydrodynamic Study of the Aortic Valve: Mock Loop Development and Test
    Vismara, Riccardo
    Fiore, Gianfranco B.
    Mangini, Andrea
    Contino, Monica
    Lemma, Massimo
    Redaelli, Alberto
    Antona, Carlo
    ASAIO JOURNAL, 2010, 56 (04) : 279 - 284
  • [33] EFFECTS OF HYDROCORTISONE ON YIELD AND BACTERICIDAL FUNCTION OF GRANULOCYTES COLLECTED BY CONTINUOUS-FLOW CENTRIFUGATION
    SHOJI, M
    VOGLER, WR
    BLOOD, 1974, 44 (03) : 435 - 443
  • [34] THE EFFECTS OF CONTINUOUS-FLOW AND DEMAND FLOW VENTILATION IN NEONATES
    WAHLIG, T
    HOLLOMAN, KK
    BING, DR
    BOROS, SJ
    MAMMEL, MC
    CLINICAL RESEARCH, 1994, 42 (01): : A34 - A34
  • [35] The Effects of Continuous-flow Left Ventricular Assist Devices on Peripheral Vascular Function
    Witman, Melissa A. H.
    Groot, H. Jonathan
    Gifford, Jayson R.
    Trinity, Joel D.
    Stehlik, Josef
    Drakos, Stavros G.
    Richardson, Russell S.
    FASEB JOURNAL, 2013, 27
  • [36] The Effects of Preserving Mitral Valve Function on a Left Atrial Assist Device: An In Vitro Mock Circulation Loop Study
    Kado, Yuichiro
    Polakowski, Anthony R.
    Kuban, Barry D.
    Horvath, David J.
    Miyamoto, Takuma
    Karimov, Jamshid H.
    Starling, Randall C.
    Fukamachi, Kiyotaka
    ASAIO JOURNAL, 2021, 67 (05) : 567 - 572
  • [37] Concomitant or late aortic valve intervention and its efficacy for aortic insufficiency associated with continuous-flow left ventricular assist device implantation
    Ando, Masahiko
    Ono, Minoru
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2022, 9
  • [38] EFFECT OF WEANING ON OXYGEN-CONSUMPTION AND CARDIOVASCULAR FUNCTION - A COMPARISON OF CONTINUOUS-FLOW AND DEMAND VALVE SYSTEMS
    YAM, PCI
    APPADURAI, IR
    KOX, WJ
    ANAESTHESIA, 1994, 49 (05) : 391 - 393
  • [39] The Hemodynamic Effects of Aortic Insufficiency in Patients Supported With Continuous-Flow Left Ventricular Assist Devices
    Sayer, Gabriel
    Sarswat, Nitasha
    Kim, Gene H.
    Adatya, Sirtaz
    Medvedofsky, Diego
    Rodgers, Daniel
    Kruse, Eric
    Ota, Takeyoshi
    Jeevanandam, Valluvan
    Lang, Roberto
    Uriel, Nir
    JOURNAL OF CARDIAC FAILURE, 2017, 23 (07) : 545 - 551
  • [40] Isolated Permanent Right Ventricular Assistance Using the HVAD Continuous-Flow Pump
    Deuse, Tobias
    Schirmer, Johannes
    Kubik, Mathias
    Reichenspurner, Hermann
    ANNALS OF THORACIC SURGERY, 2013, 95 (04): : 1434 - 1436