Influence of laser parameters on isotope fractionation and optimisation of lithium and boron isotope ratio measurements using laser ablation-multiple Faraday collector-inductively coupled plasma mass spectrometry

被引:31
|
作者
Kimura, Jun-ichi [1 ]
Chang, Qing [1 ]
Ishikawa, Tsuyoshi [2 ]
Tsujimori, Tatsuki [3 ]
机构
[1] Japan Agcy Marine Earth Sci & Technol JAMSTEC, Dept Solid Earth Geochem, 2-15 Natsushima Cho, Yokosuka, Kanagawa 2370061, Japan
[2] Japan Agcy Marine Earth Sci & Technol JAMSTEC, Kochi Inst Core Sample Res, 200 Monobe Otsu, Nankoku, Kochi 7838502, Japan
[3] Tohoku Univ, Ctr Northeast Asian Studies, Aoba Ku, 41 Kawauchi, Sendai, Miyagi 9808576, Japan
关键词
UV-FEMTOSECOND LASER; ICP-MS; IN-SITU; ULTRAVIOLET FEMTOSECOND; GLASS; MINERALS; ROCKS; SYSTEMATICS; DELTA-B-11; SAMPLES;
D O I
10.1039/c6ja00283h
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We report the origin of isotope fractionation induced by the choice of laser parameters and a method for accurate in situ determination of lithium (delta Li-7) and boron (delta B-11) isotope ratios in glasses and minerals using laser ablation multiple Faraday collector inductively coupled plasma mass spectrometry (LA-MFC-ICPMS). Laser ablation parameters were examined using 266 nm femtosecond (266FsLA) and 193 nm nanosecond excimer (193ExLA) laser ablation systems for crater diameters of 30-200 mu m. We found that higher laser repetition rates and larger crater diameters have led to enhanced fractionation of lighter isotopes, as much as -8 parts per thousand for both delta Li-7 and delta B-11. Fractionation was primarily affected by the ICP aerosol loading and secondly by the thermal fractionation at the LA site. The former was accounted for by mass loading effects, which lowered the plasma temperature and led to insufficient aerosol vaporisation. The latter was related to the molten layer on the crater walls, which resulted in coarser and heavier delta Li-7 and delta B-11 aerosols that did not reach the ICP. Both processes can result in Rayleigh fractionation during aerosol formation and vaporisation. Controlled ablation using a constant crater size, repetition rate, and high laser fluence of 193ExLA enabled reproducible ablation for the standard NIST SRM 61X glasses and unknown basalt glasses. Based on the principles of isotopic fractionation deduced from our experiments, we propose a novel ablation volume correction (AVC) protocol for accurate isotopic analyses of various samples with different matrices. Both the repeatability and the laboratory bias of the delta Li-7 and delta B-11 measurements using the new AVC protocol were better than 1 parts per thousand for samples containing a few tens to a few tens of thousands ppm Li and B. We also report significant local heterogeneity of up to several parts per thousand found in some basalt glasses, but not in NIST SRM 612 and 610.
引用
下载
收藏
页码:2305 / 2320
页数:16
相关论文
共 50 条
  • [21] The effect of matrix interferences on in situ boron isotope analysis by laser ablation multi-collector inductively coupled plasma mass spectrometry
    Standish, Christopher D.
    Chalk, Thomas B.
    Babila, Tali L.
    Milton, J. Andy
    Palmer, Martin R.
    Foster, Gavin L.
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2019, 33 (10) : 959 - 968
  • [22] Isotope ratio measurements of dysprosium by multiple collection inductively coupled plasma mass spectrometry
    Segal, I
    Halicz, L
    Platzner, IT
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2002, 216 (02) : 177 - 184
  • [23] Quantification of transient signals in multiple collector inductively coupled plasma mass spectrometry: accurate lead isotope ratio determination by laser ablation of individual fluid inclusions
    Pettke, Thomas
    Oberli, Felix
    Audetat, Andreas
    Wiechert, Uwe
    Harris, Caroline R.
    Heinrich, Christoph A.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2011, 26 (03) : 475 - 492
  • [24] Elemental fractionation in laser ablation inductively coupled plasma mass spectrometry
    Longerich, HP
    Gunther, D
    Jackson, SE
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1996, 355 (5-6): : 538 - 542
  • [25] Separation and Isotopic Measurement of Mg Isotope Ratios in Tooth Samples Using Multiple Collector-Inductively Coupled Plasma-Mass Spectrometry
    Li Zi-Xia
    He Mao-Yong
    Lu Hai
    Ren Tong-Xiang
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2016, 44 (05) : 787 - 791
  • [26] Isotope ratio and isotope dilution measurements using axial inductively coupled plasma time of flight mass spectrometry
    Emteborg, H
    Tian, XD
    Ostermann, M
    Berglund, M
    Adams, FC
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (03) : 239 - 246
  • [27] Isotope ratio and isotope dilution measurements using axial inductively coupled plasma time of flight mass spectrometry
    Emteborg, Hakan, 1600, Royal Soc of Chemistry, Cambridge, United Kingdom (15):
  • [28] Elemental fractionation of glass using laser ablation inductively coupled plasma mass spectrometry
    Figg, D
    Kahr, MS
    APPLIED SPECTROSCOPY, 1997, 51 (08) : 1185 - 1192
  • [29] LEAD ISOTOPE RATIO MEASUREMENT OF NIST 610 GLASS BY LASER ABLATION INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY
    WALDER, AJ
    ABELL, ID
    PLATZNER, I
    FREEDMAN, PA
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1993, 48 (03) : 397 - 402
  • [30] Strontium isotope analysis in prehistoric cod otoliths by laser ablation multi-collector inductively coupled plasma mass spectrometry
    Glykou, Aikaterini
    Ritchie, Kenneth
    Hargrave, Matthew S.
    Visch, Wouter
    Liden, Kerstin
    JOURNAL OF ARCHAEOLOGICAL SCIENCE-REPORTS, 2021, 37