Laboratory and field testing of sampling methods for inhalable and respirable dust

被引:19
|
作者
Linnainmaa, Markku [1 ]
Laitinen, Juha [1 ]
Leskinen, Ari [2 ]
Sippula, Olli [3 ]
Kalliokoski, Pentti [3 ]
机构
[1] Finnish Inst Occupat Hlth, FI-70701 Kuopio, Finland
[2] Finnish Meteorol Inst, Kuopio Unit, Kuopio, Finland
[3] Univ Kuopio, FIN-70211 Kuopio, Finland
关键词
aerosol monitoring; cascade impactor; inhalable dust; IOM samplers; particle size-selective sampling; respirable dust;
D O I
10.1080/15459620701763723
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The performance of four sampling devices for inhalable dust and three devices for respirable dust was tested with different kinds of dusts in the laboratory and in the field. The IOM sampler was chosen as the reference method for inhalable dust, and the IOM sampler provided with the porous plastic foam media was used as the reference method for respirable dust. The other tested instruments were the Button sampler, the optical Grimm aerosol monitor, and the Dekati two-stage cascade impactor with cutoff sizes of 10 and 4 m. The study confirmed the applicability of the IOM and Button samplers. The new foam product followed the respirable criteria well. However, the foam sampler was unstable for measuring inhalable dust, probably due to its moisture absorption. In addition, high dust loads should be avoided with the foam sampler due to increase in filtering efficiency. The concentrations of inhalable dust measured with the Button sampler, the Grimm monitor, and the impactor sampler were usually close to those measured with the reference sampler. On the other hand, impactor sampling yielded higher respirable dust concentrations than the reference method in the field, which may have been caused by particle bounce; high dust loads should be avoided while using the impactor. The results also showed that the Grimm monitor enables real-time dust concentration determinations that are accurate enough for routine monitoring of occupational exposure and for testing efficiency of control measures in workplaces.
引用
收藏
页码:28 / 35
页数:8
相关论文
共 50 条
  • [31] A field evaluation of a single sampler for respirable and inhalable indium and dust measurements at an indium-tin oxide manufacturing facility
    Blackley, Brie Hawley
    Gibbs, Jenna L.
    Cummings, Kristin J.
    Stefaniak, Aleksandr B.
    Park, Ji Young
    Stanton, Marcia
    Virji, M. Abbas
    JOURNAL OF OCCUPATIONAL AND ENVIRONMENTAL HYGIENE, 2019, 16 (01) : 66 - 77
  • [32] A Comparison of the Performance of Samplers for Respirable Dust in Workplaces and Laboratory Analysis for Respirable Quartz
    Verpaele, Steven
    Jouret, Jonathan
    ANNALS OF OCCUPATIONAL HYGIENE, 2013, 57 (01): : 54 - 62
  • [33] RESPIRABLE DUST SAMPLING IN CZECHOSLOVAK COAL-MINES
    VITEK, J
    AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1977, 38 (06): : 247 - 252
  • [34] Laboratory and field testing of particle size-selective sampling methods for mineral dusts
    Teikari, M
    Linnainmaa, M
    Laitinen, J
    Kalliokoski, P
    Vincent, J
    Tiita, P
    Raunemaa, T
    AIHA JOURNAL, 2003, 64 (03): : 312 - 318
  • [35] COMPARISON OF RECOMMENDED RESPIRABLE MASS DUST SAMPLING DEVICES
    THOMPSON, EM
    TREAFTIS, HN
    TOMB, TF
    BECKERT, AJ
    AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1977, 38 (10): : 523 - 535
  • [36] INSTRUMENT FOR SAMPLING OF RESPIRABLE DUST FOR SUBSEQUENT GRAVIMETRIC ASSESSMENT
    DUNMORE, JH
    HAMILTON, RJ
    SMITH, DSG
    JOURNAL OF SCIENTIFIC INSTRUMENTS, 1964, 41 (11): : 669 - &
  • [37] Laboratory testing of a shuttle car canopy air curtain for respirable coal mine dust control
    Reed W.R.
    Zheng Y.
    Yekich M.
    Ross G.
    Salem A.
    International Journal of Coal Science and Technology, 2018, 5 (03): : 305 - 314
  • [38] IMPROVED SAMPLING AND RECOVERY OF WOOD DUST USING MSA RESPIRABLE DUST CASSETTES
    MOORE, LL
    DUBE, DJ
    BURK, T
    AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1990, 51 (07): : A475 - A476
  • [39] FIELD COMPARISON OF PERSONAL SAMPLERS FOR INHALABLE DUST
    VAUGHAN, NP
    CHALMERS, CP
    BOTHAM, RA
    ANNALS OF OCCUPATIONAL HYGIENE, 1990, 34 (06): : 553 - 573
  • [40] The need for an international sampling convention for inhalable dust in calm air
    Liden, Goran
    Harper, Martin
    JOURNAL OF OCCUPATIONAL AND ENVIRONMENTAL HYGIENE, 2006, 3 (10) : D94 - D101