Calibration update strategies for an array of potentiometric chemical sensors

被引:18
|
作者
Rudnitskaya, Alisa [1 ,2 ]
Costa, Ana Maria S. [2 ,3 ]
Delgadillo, Ivonne [2 ,3 ]
机构
[1] Univ Aveiro, CESAM, Campus Univ Santiago, P-3810193 Aveiro, Portugal
[2] Univ Aveiro, Dept Chem, Campus Univ Santiago, P-3810193 Aveiro, Portugal
[3] Univ Aveiro, QOPNA, Campus Univ Santiago, P-3810193 Aveiro, Portugal
来源
关键词
Calibration update; Electronic tongue; Multivariate calibration standardization; Weighting; Tikhonov regularization; Joint-Y partial least square regression; MULTIVARIATE STANDARDIZATION; REGRESSION; MAINTENANCE; SELECTIVITY; VARIANTS; DESIGN;
D O I
10.1016/j.snb.2016.06.075
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
One of the obstacles to the practical use of the multisensor systems - electronic tongues is a drift, i.e. gradual change of the sensor characteristics occurring in the process of their exploitation. Sensor drift leads to the deterioration of the performance of the calibration model that was calculated before changes in the sensor responses occurred. Two approaches can be employed to deal with this phenomenon. One consists of regular re-calibration of the sensor array, which is effective but time and labor consuming. Alternatively, multivariate statistical methods can be applied to the calibration model update using small subset of the standard samples. While significant efforts have been directed to the development of the calibration transfer and calibration update techniques, they were mostly applied to the near infrared spectroscopic instruments. Very few works addressed this issue for the potentiometric sensor arrays. In the present study, applicability of calibration update methods including slope and bias correction, direct standardization using PLS2 regression and auto-associative neural network, calibration update by weighting, Tikhonov regularization and Joint Y-PLS regression to the sensor array data was evaluated. Calibration models for copper and lead quantification in mixed solutions of transition metals using an array of 7 potentiometric sensors was used as an example. Of all studied methods, weighting and JYPLS regression were shown to be the most effective permitting to reduce prediction error for new data to the level of cross-validation error for the initial data. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:1181 / 1189
页数:9
相关论文
共 50 条
  • [21] Nanostructured array materials for chemical sensors.
    Han, L
    Zhong, CI
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2004, 228 : U159 - U159
  • [22] Mutual coupling calibration for conformal array antenna with instrumental sensors
    Wang, Bu-Hong
    Hou, Qing-Song
    Guo, Ying
    Wang, Yong-Liang
    Tien Tzu Hsueh Pao/Acta Electronica Sinica, 2009, 37 (06): : 1283 - 1288
  • [23] Array errors active calibration algorithm based on instrumental sensors
    Ding Wang
    Ying Wu
    Science China Information Sciences, 2011, 54 : 1500 - 1511
  • [24] Array errors active calibration algorithm based on instrumental sensors
    Wang Ding
    Wu Ying
    SCIENCE CHINA-INFORMATION SCIENCES, 2011, 54 (07) : 1500 - 1511
  • [25] Joint Calibration and Direct Position Determination for Moving Array Sensors
    Springer, Jannik
    Oispuu, Marc
    Koch, Wolfgang
    2021 IEEE 24TH INTERNATIONAL CONFERENCE ON INFORMATION FUSION (FUSION), 2021, : 969 - 976
  • [26] Array errors active calibration algorithm based on instrumental sensors
    WANG Ding & WU Ying Institute of Information Engineering
    Science China(Information Sciences), 2011, 54 (07) : 1500 - 1511
  • [27] Array shape calibration using carry-on instrumental sensors
    Wang, BH
    Wang, YL
    Chen, H
    2003 PROCEEDINGS OF THE INTERNATIONAL CONFERENCE ON RADAR, 2003, : 636 - 639
  • [28] Two Analyte Calibration From The Transient Response Of Potentiometric Sensors Employed With The SIA Technique
    Cartas, Raul
    Mimendia, Aitor
    Legin, Andrey
    del Valle, Manel
    OLFACTION AND ELECTRONIC NOSE, PROCEEDINGS, 2009, 1137 : 543 - +
  • [29] The computational methods in the development of a novel multianalyte calibration technique for potentiometric integrated sensors systems
    Urbanowicz, Marcin
    Pijanowska, Dorota G.
    Jasinski, Artur
    Bochenska, Maria
    JOURNAL OF SOLID STATE ELECTROCHEMISTRY, 2019, 23 (07) : 2251 - 2260
  • [30] The computational methods in the development of a novel multianalyte calibration technique for potentiometric integrated sensors systems
    Marcin Urbanowicz
    Dorota G. Pijanowska
    Artur Jasiński
    Maria Bocheńska
    Journal of Solid State Electrochemistry, 2019, 23 : 2251 - 2260